Evaluation of a flash glucose monitoring system in nondiabetic dogs with rapidly changing blood glucose concentrations

被引:14
|
作者
Howard, Leigh A. [1 ]
Lidbury, Jonathan A. [1 ]
Jeffe, Nicholas [1 ]
Washburn, Shannon E. [2 ]
Patterson, Carly A. [2 ]
机构
[1] Texas A&M Univ, Coll Vet Med & Biomed Sci, Dept Small Anim Clin Sci, College Stn, TX 77843 USA
[2] Texas A&M Univ, Coll Vet Med & Biomed Sci, Dept Vet Physiol & Pharmacol, College Stn, TX 77843 USA
关键词
dog; FreeStyle Libre; hypoglycemia; interstitial glucose; OF-CARE GLUCOMETER; ACCURACY; METERS; PLASMA; CANINE; CATS;
D O I
10.1111/jvim.16273
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background A flash glucose monitoring system (FGMS; FreeStyle Libre) is useful for monitoring hypoglycemic dogs with diabetes. Objective To assess the utility of this FGMS in dogs with induced hypoglycemia and rapid fluctuations in blood glucose (BG) concentrations. Animals Twenty-four apparently healthy research (n = 10) and teaching (n = 14) dogs. Methods Prospective, observational study performed in tandem with a teaching laboratory. Regular insulin was administered to dogs and resulting hypoglycemia was corrected. Before insulin administration and every 10 minutes over a 90-minute period, serial measurements of interstitial glucose (IG) with FGMS and BG with a portable blood glucose meter (PBGM) and clinical chemistry analyzer concentrations were made. Portable blood glucose meter and FGMS readings were compared to that of the clinical chemistry analyzer. Analytical and clinical accuracy were assessed using ISO 15197:2013 criteria, including Parkes error grid analysis. Results The proportions of readings in the low BG range (BG <100 mg/dL) for which the test method measurement was within +/- 15 mg/dL of the reference BG for the PBGM and FGMS were 81.7% (161/197) and 39.1% (72/184), respectively. The proportions of readings for the PBGM and FGMS, which were not likely to affect clinical outcome according to Parkes error grid analysis, were 97.9% (233/238) and 80.1% (177/221), respectively. Conclusions and Clinical Importance In this model, there was limited agreement between the FGMS and reference standard BG measurements. The FGMS (measuring IG concentrations) was compared to peripheral BG concentrations, not brain-tissue glucose concentrations, and failed to reliably detect hypoglycemia.
引用
收藏
页码:2628 / 2635
页数:8
相关论文
共 50 条
  • [21] Evaluation of artificial pancreas technology for continuous blood glucose monitoring in dogs
    Akihiro Mori
    Peter Lee
    Takeshi Yokoyama
    Hitomi Oda
    Kaori Saeki
    Yohei Miki
    Satoshi Nozawa
    Daigo Azakami
    Yutaka Momota
    Yuki Makino
    Takako Matsubara
    Motohisa Osaka
    Katsumi Ishioka
    Toshiro Arai
    Toshinori Sako
    Journal of Artificial Organs, 2011, 14
  • [22] Flash glucose monitoring system in special situations
    Rigon, Fernanda Augustini
    Ronsoni, Marcelo Fernando
    Vianna, Andre Gustavo Daher
    Schiavon, Leonardo de Lucca
    Hohl, Alexandre
    van de Sande-Lee, Simone
    ARCHIVES OF ENDOCRINOLOGY METABOLISM, 2022, 66 (06): : 883 - 894
  • [23] Effect of site of sample collection on blood glucose concentrations measured with a portable blood glucose meter in healthy and diabetic dogs
    Jahan, Sina
    Jahan, Samin
    Jamshidi, Shahram
    Linari, Guido
    Fracassi, Federico
    Akbarein, Hesameddin
    OPEN VETERINARY JOURNAL, 2023, 13 (03) : 322 - 326
  • [24] Evaluation of the FreeStyle® Libre Flash Glucose Monitoring System in Children and Adolescents with Type 1 Diabetes
    Massa, Guy G.
    Gys, Inge
    Op't Eyndt, Anniek
    Bevilacqua, Esmiralda
    Wijnands, Anne
    Declercq, Peter
    Zeevaert, Renate
    HORMONE RESEARCH IN PAEDIATRICS, 2018, 89 (03): : 189 - 199
  • [25] Ambulatory glucose profile: Flash glucose monitoring
    Kalra, Sanjay
    Gupta, Yashdeep
    JOURNAL OF THE PAKISTAN MEDICAL ASSOCIATION, 2015, 65 (12) : 1360 - 1362
  • [26] Evaluation of six portable blood glucose meters for measuring blood glucose concentration in dogs
    Cohen, Todd A.
    Nelson, Richard W.
    Kass, Philip H.
    Christopher, Mary M.
    Feldman, Edward C.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2009, 235 (03): : 276 - 280
  • [27] Challenges of Inpatient Blood Glucose Monitoring: Standards, Methods, and Devices to Measure Blood Glucose
    Hermayer, Kathie L.
    Loftley, Aundrea S.
    Reddy, Sumana
    Narla, Satya Nandana
    Epps, Nina A.
    Zhu, Yusheng
    CURRENT DIABETES REPORTS, 2015, 15 (03)
  • [28] Clinical evaluation and alternative site blood glucose testing of the EasyPlus mini R2N blood glucose monitoring system
    Wei, Chien-Hua
    Peng, Ting-Chun
    Wang, Peng-Yuan
    Lin, Jui-Jane
    Chuang, Mei-Lin
    Chen, Chien-Chih
    CLINICA CHIMICA ACTA, 2009, 403 (1-2) : 167 - 172
  • [29] The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
    Bailey, Timothy
    Bode, Bruce W.
    Christiansen, Mark P.
    Klaff, Leslie J.
    Alva, Shridhara
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) : 787 - 794
  • [30] Etiology and prognosis for dogs with abnormal blood glucose concentrations evaluated in an emergency room
    Hagley, Simon P.
    Hopper, Kate
    Epstein, Steven E.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2020, 30 (05) : 567 - 573