Entrapment of subvalvular mitral tissue causing intermittent failure of a St Jude mitral prosthesis

被引:14
作者
Agostini, F
Click, RL
Mulvagh, SL
Abel, MD
Dearani, JA
机构
[1] Mayo Clin & Mayo Fdn, Div Cardiovasc Dis & Internal Med, Rochester, MN 55905 USA
[2] Mayo Clin & Mayo Fdn, Div Cardiovasc Anesthesia, Rochester, MN 55905 USA
[3] Mayo Clin & Mayo Fdn, Div Cardiovasc Surg, Rochester, MN 55905 USA
关键词
D O I
10.1067/mje.2000.107251
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Frequently portions of the mitral valve and subvalvular apparatus are left intact during mitral valve replacement to help preserve left ventricular function. We describe a patient with paroxysmal congestive heart failure caused by intermittent entrapment of the subvalvular apparatus in the prosthesis, preventing complete valve closure.