In vivo measurement of the frame-based application accuracy of the Neuromate neurosurgical robot

被引:93
作者
von Langsdorff, Daniel [1 ]
Paquis, Philippe [1 ]
Fontaine, Denys [1 ,2 ]
机构
[1] CHU Nice, Dept Neurosurg, Clermont Ferrand, France
[2] Univ Auvergne, IGCN EA 7282, UMR ISIT 6284, Clermont Ferrand, France
关键词
accuracy; deep brain stimulation; Neuromate robot; stereotactic frame; functional neurosurgery; DEEP BRAIN-STIMULATION; SUBTHALAMIC NUCLEUS; ELECTRODE CONTACTS; PARKINSONS-DISEASE; GUIDANCE;
D O I
10.3171/2014.9.JNS14256
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECT The application accuracy of the Neuromate neurosurgical robot has been validated in vitro but has not been evaluated in vivo for deep brain stimulation (DBS) electrode implantations. The authors conducted a study to evaluate this application accuracy in routine frame-based DBS procedures, with an independent system of measurement. METHODS The Euclidian distance was measured between the point theoretically targeted by the robot and the point actually reached, based on their respective stereotactic coordinates. The coordinates of the theoretical target were given by the robot's dedicated targeting software. The coordinates of the point actually reached were recalculated using the Stereoplan localizer system. This experiment was performed in vitro, with the frame fixed in the robot space without a patient, for 21 points spatially distributed. The in vivo accuracy was then measured in 30 basal ganglia targets in 17 consecutive patients undergoing DBS for movement disorders. RESULTS The mean in vitro application accuracy was 0.44 +/- 0.23 mm. The maximal localization error was 1.0 mm. The mean (+/- SD) in vivo application accuracy was 0.86 +/- 0.32 mm (Delta x = 0.37 +/- 0.34 mm, Delta y = 0.32 +/- 0.24 mm, Delta z = 0.58 +/- 0.31 mm). The maximal error was 1.55 mm. CONCLUSIONS The in vivo application accuracy of the Neuromate neurosurgical robot, measured with a system independent from the robot, in frame-based DBS procedures was better than 1 mm. This accuracy is at least similar to the accuracy of stereotactic frame arms and is compatible with the accuracy required in DBS procedures.
引用
收藏
页码:191 / 194
页数:4
相关论文
共 14 条
[1]   Stereoelectroencephalography: Surgical Methodology, Safety, and Stereotactic Application Accuracy in 500 Procedures [J].
Cardinale, Francesco ;
Cossu, Massimo ;
Castana, Laura ;
Casaceli, Giuseppe ;
Schiariti, Marco Paolo ;
Miserocchi, Anna ;
Fuschillo, Dalila ;
Moscato, Alessio ;
Caborni, Chiara ;
Arnulfo, Gabriele ;
Lo Russo, Giorgio .
NEUROSURGERY, 2013, 72 (03) :353-366
[2]   Image-guided robotic neurosurgery-an in vitro and in vivo point accuracy evaluation experimental study [J].
Chan, Frank ;
Kassim, Irwan ;
Lo, Charles ;
Ho, Chi Long ;
Low, David ;
Ang, Beng Ti ;
Ng, Ivan .
SURGICAL NEUROLOGY, 2009, 71 (06) :640-648
[3]   Deep Brain Stimulation of the Subthalamic Nucleus in Parkinson's Disease: Usefulness of Intraoperative Radiological Guidance The Stereoplan [J].
Derrey, Stephane ;
Maltete, David ;
Chastan, Nathalie ;
Debono, Bertrand ;
Proust, Francois ;
Gerardin, Emmanuel ;
Weber, Jacques ;
Mihout, Bruno ;
Freger, Pierre .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2008, 86 (06) :351-358
[4]   Validation of the PathFinder™ neurosurgical robot using a phantom [J].
Eljamel, M. S. .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2007, 3 (04) :372-377
[5]   Subthalamic nucleus stimulation using a Fisher ZD stereotactic frame, MR-CT fusion guidance and peroperative orthogonal radiographs, in Parkinson disease [J].
Fontaine, D. ;
Borg, M. ;
Bayreuther, C. ;
von Langsdorff, D. ;
Magnie-Mauro, M. -N. ;
Chanalet, S. ;
Duval, M. ;
Levivier, M. ;
Paquis, P. .
NEUROCHIRURGIE, 2007, 53 (06) :463-469
[6]   Deep brain stimulation of the subthalamic nucleus in Parkinson's disease:: evaluation of active electrode contacts [J].
Hamel, W ;
Fietzek, U ;
Morsnowski, A ;
Schrader, B ;
Herzog, J ;
Weinert, D ;
Pfister, G ;
Müller, D ;
Volkmann, J ;
Deuschl, G ;
Mehdorn, HM .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2003, 74 (08) :1036-1046
[7]   MRI- and skull x-ray-based approaches to evaluate the position of deep brain stimulation electrode contacts -: a technical note [J].
Hamel, W ;
Schrader, B ;
Weinert, D ;
Herzog, J ;
Volkmann, J ;
Deuschl, G ;
Müller, D ;
Mehdorn, HM .
ZENTRALBLATT FUR NEUROCHIRURGIE, 2002, 63 (02) :65-69
[8]  
Li Qing Hang, 2002, Comput Aided Surg, V7, P90, DOI 10.3109/10929080209146020
[9]   THE APPLICATION ACCURACY OF STEREOTAXIC FRAMES [J].
MACIUNAS, RJ ;
GALLOWAY, RL ;
LATIMER, JW .
NEUROSURGERY, 1994, 35 (04) :682-694
[10]   Robotics in neurosurgery [J].
McBeth, PB ;
Louw, DF ;
Rizun, PR ;
Sutherland, GR .
AMERICAN JOURNAL OF SURGERY, 2004, 188 (4A) :68S-75S