Quality assurance and long-term stability of a novel 3-in-1 X-ray system for brachytherapy

被引:10
|
作者
Karius, Andre [1 ,2 ]
Szkitsak, Juliane [1 ,2 ]
Boronikolas, Vasilios [3 ,4 ,5 ,6 ]
Fietkau, Rainer [1 ,2 ]
Bert, Christoph [1 ,2 ]
机构
[1] Friedrich Alexander Univ Erlangen Nurnberg, Univ Klinikum Erlangen, Dept Radiat Oncol, Univ Str 27, D-91054 Erlangen, Germany
[2] Comprehens Canc Ctr Erlangen EMN CCC ER EMN, Erlangen, Germany
[3] Univ Klinikum Freiburg, Klin Strahlenheilkunde, Abt Med Phys, Freiburg, Germany
[4] Albert Ludwigs Univ Freiburg, Med Fak, Freiburg, Germany
[5] Deutsch Konsortium Translat Krebsforsch DKTK, Partnerstandort Freiburg, Freiburg, Germany
[6] Deutsch Krebsforschungszentrum DKFZ, Partnerstandort Freiburg, Heidelberg, Germany
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2022年 / 23卷 / 09期
关键词
3-in-1 X-ray system; cone-beam computed tomography; planar imaging; quality assurance; CONE-BEAM CT; SURGERY;
D O I
10.1002/acm2.13727
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose A novel, mobile 3-in-1 X-ray system featuring radiography, fluoroscopy, and cone-beam computed tomography (CBCT) has been launched for brachytherapy recently. Currently, there is no quality assurance (QA) procedure explicitly applicable to this system equipped with innovative technologies such as dynamic jaws and motorized lasers. We developed a dedicated QA procedure and, based on its performance for a duration of 6 months, provide an assessment of the device's stability over time. Methods With the developed QA procedure, we assessed the system's planar and CBCT-imaging performance by investigating geometric accuracy, CT-number stability, contrast-noise-ratio, uniformity, spatial resolution, low-contrast detectability, dynamic range, and X-ray exposure using dedicated phantoms. Furthermore, we evaluated geometric stability by using the flexmap-approach and investigated the device's laser- and jaw-positioning accuracy with an in-house test phantom. CBCT- and planar-imaging protocols for pelvis, breast, and abdomen imaging were examined. Results Planar- and CBCT-imaging performances were widely stable with a geometric accuracy <= 1 mm, CT-number stability of up to 46 HU, and uniformity variations of up to 48 HU over time. For planar imaging, low-contrast detectability and dynamic range exceeded current recommendations. Although geometric stability was considered tolerable, partly substantial positioning inaccuracies of up to more than 120 mm and -13 mm were obtained for lasers and jaws, respectively. X-ray exposure showed small variations of <= 0.56 mu Gy and <= 0.76 mGy for planar- and CBCT-imaging, respectively. The conductance of the QA procedure allowed a smooth evaluation of the system's overall performance. Conclusion We developed a QA workflow for a novel 3-in-1 X-ray system allowing to assess the device's imaging and hardware performance. The system showed in general a reasonable imaging performance and stability over time, whereas improvements regarding laser and jaw accuracy are strictly required.
引用
收藏
页数:11
相关论文
共 12 条
  • [1] Low-cost x-ray quality assurance system based on optical sensors and phosphor screen
    Fleming, Peter
    Sikstrom, Lee
    Cunningham, Ian A.
    Wasswa, William
    Holdsworth, David W.
    OPTICS AND BIOPHOTONICS IN LOW-RESOURCE SETTINGS X, 2024, 12832
  • [2] Could X-ray Predict Long-term Complications in Patients with Diabetic Foot Osteomyelitis?
    Tardaguila-Garcia, Aroa
    Garcia-Alvarez, Yolanda
    Sanz-Corbalan, Irene
    Lopez-Moral, Mateo
    Molines-Barroso, Raul J.
    Luis Lazaro-Martinez, Jose
    ADVANCES IN SKIN & WOUND CARE, 2022, 35 (08) : 1 - 5
  • [3] Long-term two-dimensional pixel stability of EPIDs used for regular linear accelerator quality assurance
    B. W. King
    L. Clews
    P. B. Greer
    Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 459 - 466
  • [4] Long-term two-dimensional pixel stability of EPIDs used for regular linear accelerator quality assurance
    King, B. W.
    Clews, L.
    Greer, P. B.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (04) : 459 - 466
  • [5] Towards a Quality Assurance Protocol for 2D and 3D Breast X-Ray Imaging Techniques
    Bosmans, H.
    Vollmar, S.
    Lemmens, K.
    Kalender, W.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 647 - 649
  • [6] Proficiency testing of instruments - a key element for quality assurance in X-ray powder diffraction - part 1: Matching the effort to the actual needs
    Peplinski, B
    Wenzel, J
    EUROPEAN POWDER DIFFRACTION, PTS 1 AND 2, 2000, 321-3 : 144 - 149
  • [7] 3D cone-beam CT of the ankle using a novel twin robotic X-ray system: Assessment of image quality and radiation dose
    Grunz, Jan-Peter
    Kunz, Andreas Steven
    Gietzen, Carsten Herbert
    Weng, Andreas Max
    Veyhl-Wichmann, Maike
    Ergun, Suleyman
    Schmitt, Rainer
    Bley, Thorsten Alexander
    Gassenmaier, Tobias
    EUROPEAN JOURNAL OF RADIOLOGY, 2019, 119
  • [8] 1H-MRS internal thermometry in test-objects (phantoms) to within 0.1 K for quality assurance in long-term quantitative MR studies
    Samson, R. S.
    Thornton, J. S.
    McLean, M. A.
    Williams, S. C. R.
    Tofts, P. S.
    NMR IN BIOMEDICINE, 2006, 19 (05) : 560 - 565
  • [9] Twin Robotic X-Ray System for 3D Cone-Beam CT of the Wrist: An Evaluation of Image Quality and Radiation Dose
    Grunz, Jan-Peter
    Gietzen, Carsten H.
    Kunz, Andreas S.
    Weng, Andreas M.
    Veyhl-Wichmann, Maike
    Erguen, Sueleyman
    Bley, Thorsten A.
    Schmitt, Rainer
    Gassenmaier, Tobias
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 214 (02) : 422 - 427
  • [10] Phase 1 clinical study to assess the safety of a novel drug delivery system providing long-term topical steroid therapy for chronic rhinosinusitis
    Douglas, Richard G.
    Psaltis, Alkis J.
    Rimmer, Joanne
    Kuruvilla, Tom
    Cervin, Anders
    Kuang, Yina
    INTERNATIONAL FORUM OF ALLERGY & RHINOLOGY, 2019, 9 (04) : 378 - 387