Object size effect on the contact potential difference measured by scanning Kelvin probe method

被引:4
作者
Polyakov, B. [1 ]
Krutokhvostov, R. [1 ]
Kuzmin, A. [1 ]
Tamanis, E. [2 ]
Muzikante, I. [1 ]
Tale, I. [1 ]
机构
[1] Univ Latvia, Inst Solid State Phys, LV-1063 Riga, Latvia
[2] Daugavpils Univ, G Liberta Innovat Microscopy Ctr, LV-5401 Daugavpils, Latvia
关键词
FORCE MICROSCOPY; WORK FUNCTION; RESOLUTION; OXIDE; SCIENCE;
D O I
10.1051/epjap/2010088
中图分类号
O59 [应用物理学];
学科分类号
摘要
Contact potential difference (CPD) was measured by macroscopic Kelvin probe instrument and scanning Kelvin probe microscope on Al, Ni and Pt on ITO substrates at ambient conditions. CPD values measured by scanning Kelvin probe microscope and macroscopic Kelvin probe are close within the error of about 10-30% for large studied objects, whereas scanning Kelvin probe microscope signal decreases, when the object size becomes smaller than 1.4 mu m. CPD and electric field signals measured using many-pass technique allowed us to estimate the influence of electrostatic field disturbance, especially, in the case of small objects.
引用
收藏
页数:5
相关论文
共 26 条
[1]   SURFACE OXIDATION OF NICKEL METAL AS STUDIED BY X-RAY PHOTOELECTRON-SPECTROSCOPY [J].
ALLEN, GC ;
TUCKER, PM ;
WILD, RK .
OXIDATION OF METALS, 1979, 13 (03) :223-236
[2]   NOISE AND THE KELVIN METHOD [J].
BAIKIE, ID ;
MACKENZIE, S ;
ESTRUP, PJZ ;
MEYER, JA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (05) :1326-1332
[3]   Atomic force microscope based Kelvin measurements: Application to an electrochemical reaction [J].
Bohmisch, M ;
Burmeister, F ;
Rettenberger, A ;
Zimmermann, J ;
Boneberg, J ;
Leiderer, P .
JOURNAL OF PHYSICAL CHEMISTRY B, 1997, 101 (49) :10162-10165
[4]   Potential distribution of Cu(In,Ga)(S,Se)2-solar cell cross-sections measured by Kelvin probe force microscopy [J].
Glatzel, T ;
Steigert, H ;
Sadewasser, S ;
Klenk, R ;
Lux-Steiner, MC .
THIN SOLID FILMS, 2005, 480 :177-182
[5]   Amplitude or frequency modulation-detection in Kelvin probe force microscopy [J].
Glatzel, T ;
Sadewasser, S ;
Lux-Steiner, MC .
APPLIED SURFACE SCIENCE, 2003, 210 (1-2) :84-89
[6]  
Gupta Arjun K, 2004, HDB BETA DISTRIBUTIO
[7]   Nanoscale oxide growth on Al single crystals at low temperatures: Variable charge molecular dynamics simulations [J].
Hasnaoui, A ;
Politano, O ;
Salazar, JM ;
Aral, G .
PHYSICAL REVIEW B, 2006, 73 (03)
[8]  
Holzl J., 1979, Work Functions of Metals, in Solid Surface Physics
[9]   Practical aspects of Kelvin probe force microscopy [J].
Jacobs, HO ;
Knapp, HF ;
Stemmer, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (03) :1756-1760
[10]   Resolution and contrast in Kelvin probe force microscopy [J].
Jacobs, HO ;
Leuchtmann, P ;
Homan, OJ ;
Stemmer, A .
JOURNAL OF APPLIED PHYSICS, 1998, 84 (03) :1168-1173