INFLUENCE OF AXIAL MOVEMENT ON FATIGUE OF PROFILE® NI-TI ROTARY INSTRUMENTS: AN IN VITRO EVALUATION

被引:2
作者
Avoaka, Marie-Chantal [1 ,2 ]
Haikel, Youssef [1 ,2 ]
机构
[1] INSERM, U977, F-67085 Strasbourg, France
[2] Univ Strasbourg, Fac Dent, Dept Restorat Dent & Endodont, F-67000 Strasbourg, France
关键词
axial movement; fatigue; instrument separation; life-span; Ni-Ti instruments; simulated canals; TITANIUM ENDODONTIC INSTRUMENTS; CYCLIC FATIGUE; TORSIONAL PROPERTIES; STAINLESS-STEEL; ENGINE-DRIVEN; FILES; FRACTURE; HAND;
D O I
10.17305/bjbms.2010.2703
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The aim of this study was to evaluate the influence of the axial movement and the angle of curve (in degrees) on fatigue of nickel-titanium (Ni-TI) ProFile (R) rotary endodontic instruments. Ni-Ti ProFile (R) rotary instruments (Maillefer SA, Ballaigues, Switzerland), 25 mm long in the range of ISO size is to 40 with two tapers (0.4 and 0.6) were evaluated. They are divided in two groups: the instruments with axial movement and those without axial movement. The system used to test the fatigue is maintained in mechanical conditions as close as possible to the clinical situation. The axial movement is in the order of 2 mm in corono-apical direction with a frequency of 1Hz. The concave radii incorporating a notched V-form for guiding the instruments were: 5; 7,5 and 10 mm. The rotary system is mounted on an electric handpiece and rotated at 350 rpm speed as recommended by the manufacturers. The instruments are rotated until their separation, and the time, in seconds, is recorded. Statistical evaluation is undertaken using a two-way t-test to identify significant differences between variables in the study (p<0,05). We found significant statistical difference (p<0,05) between Ni-Ti engine drive ProFile (R) instruments incorporating an axial movement and the instruments without axial movement with the same radius of curvature, size and taper. The incorporation of the axial movement increases significantly the life-span of the ProFile (R) rotary instruments. This should reduce the risk of the instrument separation during the endodontic treatment.
引用
收藏
页码:107 / 111
页数:5
相关论文
共 20 条
[1]  
*AM NAT STAND I, 1981, AM DENT ASS SPEC, V28
[2]   TORSIONAL PROPERTIES OF STAINLESS-STEEL CANAL MASTER-U AND FLEXOGATES [J].
CAMPS, JJ ;
PERTOT, WJ .
INTERNATIONAL ENDODONTIC JOURNAL, 1994, 27 (06) :334-338
[3]   THE EFFECTS OF CYCLICAL AXIAL MOTION ON ROTARY ENDODONTIC INSTRUMENT FATIGUE [J].
DEDERICH, DN ;
ZAKARIASEN, KL .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1986, 61 (02) :192-196
[4]   Effect of rotational speed on the breakage of nickel-titanium rotary files [J].
Dietz, DB ;
Di Fiore, PM ;
Bahcall, JK ;
Lautenschlager, EP .
JOURNAL OF ENDODONTICS, 2000, 26 (02) :68-71
[5]   A COMPARISON OF ROOT-CANAL PREPARATIONS USING NI-TI HAND, NI-TI ENGINE-DRIVEN, AND K-FLEX ENDODONTIC INSTRUMENTS [J].
GLOSSON, CR ;
HALLER, RH ;
DOVE, SB ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1995, 21 (03) :146-151
[6]   Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems [J].
Grande, N. M. ;
Plotino, G. ;
Pecci, R. ;
Bedini, R. ;
Malagnino, V. A. ;
Somma, F. .
INTERNATIONAL ENDODONTIC JOURNAL, 2006, 39 (10) :755-763
[7]   GUIDELINES FOR PREVENTION OF FRACTURE OF ROOT CANAL INSTRUMENTS [J].
GROSSMAN, LI .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1969, 28 (05) :746-&
[8]   DYNAMIC FRACTURE OF HYBRID ENDODONTIC HAND INSTRUMENTS COMPARED WITH TRADITIONAL FILES [J].
HAIKEL, Y ;
GASSER, P ;
ALLEMANN, C .
JOURNAL OF ENDODONTICS, 1991, 17 (05) :217-220
[9]   Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments [J].
Haïkel, Y ;
Serfaty, R ;
Bateman, G ;
Senger, B ;
Allemann, C .
JOURNAL OF ENDODONTICS, 1999, 25 (06) :434-440
[10]  
HAIKEL Y, 1998, J ENDODONT, V24, P735