Allergen-specific IgE measured by a continuous random-access immunoanalyzer: Interassay comparison and agreement with skin testing

被引:53
作者
Ollert, M
Weissenbacher, S
Rakoski, J
Ring, J
机构
[1] Tech Univ Munich, Dept Dermatol & Allergy, D-80802 Munich, Germany
[2] Tech Univ Munich, Clin Res Div Mol & Clin Allergotoxicol, D-80802 Munich, Germany
[3] Tech Univ Munich, GSF, Div Environm Dermatol & Allergy, D-80802 Munich, Germany
关键词
D O I
10.1373/clinchem.2004.046565
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Our aims were to evaluate the performance of a fully automated system for measuring circulating allergen-specific IgE (sIgE) against an established in vitro assay and to assess the system's diagnostic accuracy against objective clinical criteria for identifying sensitization to specific allergens. Methods: Using both the IMMULITE (R) 2000 Allergy system (IML) and an assay based on the widely used ImmunoCAP (R) technology (CAP), we measured sIgE in serum samples from 169 persons with suspected allergies to airborne or insect venom allergens. Skin testing outcome served as the clinical comparison method. Results: Interassay classification agreement between the IML and CAP, relative to the usual allergen-specific IgE cutoff of 0.35 kIU/L, ranged from 76% (yellow jacket venom) to 95% (orchard grass); agreement was 88.3% for all 9 allergens combined (766 results). The 90 discordant results, when resolved by skin testing, showed better agreement with the IML (72%) than with the CAP (28%). Compared with skin testing, for each of the 9 allergens studied, the area under the ROC curve was at least as large for the IML as for the CAP, reflecting in part the more extensive working range of the IML (0.10-100 kIU/L vs 0.35-100 kIU/L for CAP). Conclusion: Laboratory testing for sIgE can be performed on a fully automated, random-access system with an extended working range and with diagnostic accuracy for representative allergens equivalent to or better than that of the semiautomated CAP technology. (c) 2005 American Association for Clinical Chemistry.
引用
收藏
页码:1241 / 1249
页数:9
相关论文
共 30 条