Comparison of dynamic contour tonometry and Goldmann applanation tonometry following penetrating keratoplasty

被引:6
|
作者
Kandarakis, Artemios [1 ]
Soumplis, Vasileios [1 ]
Pitsas, Christos [1 ]
Kandarakis, Stylianos [2 ]
Halikias, Jiannis [3 ]
Karagiannis, Dimitrios [1 ]
机构
[1] Athens Eye Hosp, A Eye Clin, Athens 10672, Greece
[2] Mt Sinai Sch Med, Dept Ophthalmol, New York, NY USA
[3] Athens Univ Econ, Dept Mkt & Commun, Athens, Greece
关键词
dynamic contour tonometry; Goldmann applanation tonometry; intraocular pressure; penetrating keratoplasty; MEASURING INTRAOCULAR-PRESSURE; CENTRAL CORNEAL THICKNESS; TONO-PEN; POSTKERATOPLASTY EYES; GLAUCOMA;
D O I
10.3129/i10-035
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To evaluate dynamic contour tonometry (DCT) versus Goldmann applanation tonometry (GAT) intraocular pressure (IOP) measurements in eyes that underwent penetrating keratoplasty (PKP). Design: Prospective, cross-sectional, observational study. Participants: Thirty-one eyes of 28 patients were examined after PKP. Methods: All eyes had undergone PKP with interrupted sutures. The postoperative period was more than 1 year for 25 eyes and less than 1 year for 6. Sutures were removed based on corneal topography and refraction. IOP was measured by both DCT and GAT methods and was correlated to the number of remaining sutures. Results: IOP readings were successfully obtained in 25/31 (80.6%) with DTC and in 21/31(67.7%) with GAT (p = 0.25). In eyes with fewer than 4 remaining sutures, both methods were successful. In eyes with more than 4 sutures, the success rates of DCT and GAT were 66.7% and 44.4%, respectively (p = 0.18). In PKPs with a postoperative period of more than I year, the success rates of DCT and GAT were 96% and 84%, respectively (p = 0.16). In 20 eyes, both methods measured the IOP. The mean IOP obtained by DCT (16.6 [SD 2.8] mm Hg) was higher than the mean IOP obtained from GAT (15.1 [SD 3.6] mm Hg). The IOPs from the 2 instruments correlated significantly (p < 0.05) and the mean difference was 1.5 mm Hg. Conclusions: The success rate in measuring IOP with DCT and GAT did not show any statistically significant difference. Both methods were less effective measuring the IOP after recent PKPs and regrafts. However, DCT seemed to be superior to GAT in corneas with more than 4 remaining sutures and in PKPs performed more than 1 year earlier. The absolute values of IOP were higher with DCT than with GAT.
引用
收藏
页码:489 / 493
页数:5
相关论文
共 50 条
  • [21] Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry
    J Y F Ku
    H V Danesh-Meyer
    J P Craig
    G D Gamble
    C N J McGhee
    Eye, 2006, 20 : 191 - 198
  • [22] Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry
    Ku, JYF
    Danesh-Meyer, HV
    Craig, JP
    Gamble, GD
    McGhee, CNJ
    EYE, 2006, 20 (02) : 191 - 198
  • [23] Comparison of dynamic contour, rebound and standard Goldmann applanation tonometry
    Martinez-de-la-Casa, J
    Garcia-Feijoo, J
    Vico, E
    Fernandez-Vidal, A
    Castillo, A
    Mendez, C
    Pablo, L
    Garcia-Sanchez, J
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [24] Comparison of dynamic contour tonometry and Goldmann applanation tonometry in glaucoma patients and healthy subjects
    Barleon, Lorenz
    Hoffmann, Esther M.
    Berres, Manfred
    Pfeiffer, Norbert
    Grus, Franz H.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 142 (04) : 583 - 590
  • [25] A comparison between Goldmann applanation tonometry and dynamic contour tonometry after photorefractive keratectomy
    Sadigh, Afshin Lotfi
    Fouladi, Rohollah F.
    Hashemi, Hassan
    Beheshtnejad, Amir Houshang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 251 (02) : 603 - 608
  • [26] Dynamic Contour Tonometry versus Goldmann Applanation Tonometry after vitrectomy
    Mamas, Nikolaos
    Fuest, Matthias
    Roeler, Gernot
    Mazinani, Babac
    Plange, Niklas
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [27] Goldmann Applanation Tonometry Versus Dynamic Contour Tonometry After Vitrectomy
    Mamas, Nikolaos
    Fuest, Matthias
    Koutsonas, Antonios
    Roessler, Gernot
    Mazinani, Babac E.
    Walter, Peter
    Plange, Niklas
    JOURNAL OF GLAUCOMA, 2016, 25 (08) : 663 - 668
  • [28] Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study
    Mona Pache
    Sonja Wilmsmeyer
    Sonja Lautebach
    Jens Funk
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2005, 243 : 763 - 767
  • [29] Dynamic contour tonometry (DCT) versus Goldmann applanation tonometry (GAT) - a comparison of agreement and reproducibility
    Sonja Herdener
    Mona Pache
    Sonja Lautebach
    Jens Funk
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2006, 245 : 1027 - 1030
  • [30] Dynamic contour tonometry (DCT) versus Goldmann applanation tonometry (GAT) - a comparison of agreement and reproducibility
    Herdener, Sonja
    Pache, Mona
    Lautebach, Sonja
    Funk, Jens
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2007, 245 (07) : 1027 - 1030