Comparison of dynamic contour tonometry and Goldmann applanation tonometry following penetrating keratoplasty

被引:6
|
作者
Kandarakis, Artemios [1 ]
Soumplis, Vasileios [1 ]
Pitsas, Christos [1 ]
Kandarakis, Stylianos [2 ]
Halikias, Jiannis [3 ]
Karagiannis, Dimitrios [1 ]
机构
[1] Athens Eye Hosp, A Eye Clin, Athens 10672, Greece
[2] Mt Sinai Sch Med, Dept Ophthalmol, New York, NY USA
[3] Athens Univ Econ, Dept Mkt & Commun, Athens, Greece
关键词
dynamic contour tonometry; Goldmann applanation tonometry; intraocular pressure; penetrating keratoplasty; MEASURING INTRAOCULAR-PRESSURE; CENTRAL CORNEAL THICKNESS; TONO-PEN; POSTKERATOPLASTY EYES; GLAUCOMA;
D O I
10.3129/i10-035
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To evaluate dynamic contour tonometry (DCT) versus Goldmann applanation tonometry (GAT) intraocular pressure (IOP) measurements in eyes that underwent penetrating keratoplasty (PKP). Design: Prospective, cross-sectional, observational study. Participants: Thirty-one eyes of 28 patients were examined after PKP. Methods: All eyes had undergone PKP with interrupted sutures. The postoperative period was more than 1 year for 25 eyes and less than 1 year for 6. Sutures were removed based on corneal topography and refraction. IOP was measured by both DCT and GAT methods and was correlated to the number of remaining sutures. Results: IOP readings were successfully obtained in 25/31 (80.6%) with DTC and in 21/31(67.7%) with GAT (p = 0.25). In eyes with fewer than 4 remaining sutures, both methods were successful. In eyes with more than 4 sutures, the success rates of DCT and GAT were 66.7% and 44.4%, respectively (p = 0.18). In PKPs with a postoperative period of more than I year, the success rates of DCT and GAT were 96% and 84%, respectively (p = 0.16). In 20 eyes, both methods measured the IOP. The mean IOP obtained by DCT (16.6 [SD 2.8] mm Hg) was higher than the mean IOP obtained from GAT (15.1 [SD 3.6] mm Hg). The IOPs from the 2 instruments correlated significantly (p < 0.05) and the mean difference was 1.5 mm Hg. Conclusions: The success rate in measuring IOP with DCT and GAT did not show any statistically significant difference. Both methods were less effective measuring the IOP after recent PKPs and regrafts. However, DCT seemed to be superior to GAT in corneas with more than 4 remaining sutures and in PKPs performed more than 1 year earlier. The absolute values of IOP were higher with DCT than with GAT.
引用
收藏
页码:489 / 493
页数:5
相关论文
共 50 条
  • [1] Comparison of dynamic contour tonometry with Goldmann applanation tonometry
    Kaufmann, C
    Bachmann, LM
    Thiel, MA
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (09) : 3118 - 3121
  • [2] Comparison of dynamic contour tonometry and Goldmann applanation tonometry in deep lamellar and penetrating keratoplasties
    Ceruti, Piero
    Morbio, Roberta
    Marraffa, Michele
    Marchini, Giorgio
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (02) : 215 - 221
  • [3] Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Keratoconus
    Unterlauft, Jan Darius
    Schaedle, Nina
    Kasper, Karsten
    Klink, Thomas
    Geerling, Gerd
    CORNEA, 2011, 30 (10) : 1078 - 1082
  • [4] A Clinical Comparison of Dynamic Contour Tonometry Versus Goldmann Applanation Tonometry
    Yalcinbayir, Ozgur
    Baykara, Mehmet
    Atasoy, Aydin
    Ozcetin, Hikmet
    OPHTHALMIC SURGERY LASERS & IMAGING, 2010, 41 (04) : 437 - 442
  • [5] Comparison of Dynamic Contour Tonometry with Goldmann Applanation Tonometry following Descemet's Stripping Automated Endothelial Keratoplasty (DSAEK)
    Bochmann, F.
    Kaufmann, C.
    Becht, C.
    Bachmann, L. M.
    Thiel, M. A.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2009, 226 (04) : 241 - 244
  • [6] Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice
    Halkiadakis, Ioannis
    Patsea, Eleni
    Chatzimichali, Katerina
    Skouriotis, Sotiris
    Chalkidou, Sonia
    Amariotakis, Georgios
    Papakonstadinou, Dimitrios
    Theodossiadis, George
    Amariotakis, Apostolos
    Georgopoulos, Gerasimos
    ACTA OPHTHALMOLOGICA, 2009, 87 (03) : 323 - 328
  • [7] Differences between Goldmann Applanation Tonometry and Dynamic Contour Tonometry following Trabeculectomy
    Detorakis, Efstathios T.
    Grammenandi, Emilia
    Pallikaris, Ioannis G.
    Tsilimbaris, Miltiadis K.
    JOURNAL OF OPHTHALMOLOGY, 2010, 2010
  • [8] Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes
    Ceruti, P.
    Morbio, R.
    Marraffa, M.
    Marchini, G.
    EYE, 2009, 23 (02) : 262 - 269
  • [9] Intraoular pressure measurement - Comparison of dynamic contour tonometry and Goldmann applanation tonometry
    Schneider, ED
    Grehn, F
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [10] Comparison between dynamic contour tonometry and Goldmann applanation tonometry correcting equations
    Maddalena De Bernardo
    Claudia Casaburi
    Ilaria De Pascale
    Luigi Capasso
    Ferdinando Cione
    Nicola Rosa
    Scientific Reports, 12