Comparison of dynamic contour tonometry and Goldmann applanation tonometry following penetrating keratoplasty

被引:6
作者
Kandarakis, Artemios [1 ]
Soumplis, Vasileios [1 ]
Pitsas, Christos [1 ]
Kandarakis, Stylianos [2 ]
Halikias, Jiannis [3 ]
Karagiannis, Dimitrios [1 ]
机构
[1] Athens Eye Hosp, A Eye Clin, Athens 10672, Greece
[2] Mt Sinai Sch Med, Dept Ophthalmol, New York, NY USA
[3] Athens Univ Econ, Dept Mkt & Commun, Athens, Greece
来源
CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE | 2010年 / 45卷 / 05期
关键词
dynamic contour tonometry; Goldmann applanation tonometry; intraocular pressure; penetrating keratoplasty; MEASURING INTRAOCULAR-PRESSURE; CENTRAL CORNEAL THICKNESS; TONO-PEN; POSTKERATOPLASTY EYES; GLAUCOMA;
D O I
10.3129/i10-035
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To evaluate dynamic contour tonometry (DCT) versus Goldmann applanation tonometry (GAT) intraocular pressure (IOP) measurements in eyes that underwent penetrating keratoplasty (PKP). Design: Prospective, cross-sectional, observational study. Participants: Thirty-one eyes of 28 patients were examined after PKP. Methods: All eyes had undergone PKP with interrupted sutures. The postoperative period was more than 1 year for 25 eyes and less than 1 year for 6. Sutures were removed based on corneal topography and refraction. IOP was measured by both DCT and GAT methods and was correlated to the number of remaining sutures. Results: IOP readings were successfully obtained in 25/31 (80.6%) with DTC and in 21/31(67.7%) with GAT (p = 0.25). In eyes with fewer than 4 remaining sutures, both methods were successful. In eyes with more than 4 sutures, the success rates of DCT and GAT were 66.7% and 44.4%, respectively (p = 0.18). In PKPs with a postoperative period of more than I year, the success rates of DCT and GAT were 96% and 84%, respectively (p = 0.16). In 20 eyes, both methods measured the IOP. The mean IOP obtained by DCT (16.6 [SD 2.8] mm Hg) was higher than the mean IOP obtained from GAT (15.1 [SD 3.6] mm Hg). The IOPs from the 2 instruments correlated significantly (p < 0.05) and the mean difference was 1.5 mm Hg. Conclusions: The success rate in measuring IOP with DCT and GAT did not show any statistically significant difference. Both methods were less effective measuring the IOP after recent PKPs and regrafts. However, DCT seemed to be superior to GAT in corneas with more than 4 remaining sutures and in PKPs performed more than 1 year earlier. The absolute values of IOP were higher with DCT than with GAT.
引用
收藏
页码:489 / 493
页数:5
相关论文
共 38 条
  • [1] EVALUATION OF THE TONO-PEN AND THE PULSAIR TONOMETERS
    ARMSTRONG, TA
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1990, 109 (06) : 716 - 720
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Comparison of dynamic contour tonometry and Goldmann applanation tonometry in deep lamellar and penetrating keratoplasties
    Ceruti, Piero
    Morbio, Roberta
    Marraffa, Michele
    Marchini, Giorgio
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (02) : 215 - 221
  • [4] GLAUCOMA IN THE IMMEDIATE POSTOPERATIVE PERIOD AFTER PENETRATING KERATOPLASTY
    CHIEN, AM
    SCHMIDT, CM
    COHEN, EJ
    RAJPAL, RK
    SPERBER, LTD
    RAPUANO, CJ
    MOSTER, M
    SMITH, M
    LAIBSON, PR
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1993, 115 (06) : 711 - 714
  • [5] EVALUATION OF THE TONO-PEN-2 AND THE X-PERT(TM) NONCONTACT TONOMETERS IN CATARACT-SURGERY
    DENIS, P
    NORDMANN, JP
    BERTIN, V
    GAYRAUD, JM
    LAROCHE, L
    SARAUX, H
    [J]. OPHTHALMOLOGICA, 1993, 207 (03) : 155 - 161
  • [6] Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness
    Doyle, A
    Lachkar, Y
    [J]. JOURNAL OF GLAUCOMA, 2005, 14 (04) : 288 - 292
  • [7] EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
  • [8] FOULKS GN, 1987, OPHTHALMOLOGY, V94, P871
  • [9] TONO-PEN TONOMETRY IN NORMAL AND IN POSTKERATOPLASTY EYES
    GEYER, O
    MAYRON, Y
    LOEWENSTEIN, A
    NEUDORFER, M
    ROTHKOFF, L
    LAZAR, M
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 1992, 76 (09) : 538 - 540
  • [10] INCIDENCE OF INCREASED INTRAOCULAR-PRESSURE AFTER KERATOPLASTY
    GOLDBERG, DB
    SCHANZLIN, DJ
    BROWN, SI
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1981, 92 (03) : 372 - 377