In situ generation of Co(II) by use of a solid-phase reactor in an FIA assembly for the spectrophotometric determination of penicillamine

被引:34
|
作者
Corominas, BGT
Pferzschner, J
Catalá-Icardo, M
Zamora, LL
Calatayud, JM [1 ]
机构
[1] Univ Cardenal Herrear CEU, Dept Quim Bioquim & Biol Mol, Valencia, Spain
[2] Univ Valencia, Dept Quim Analit, E-46003 Valencia, Spain
关键词
penicillamine; flow injection analysis; FIA; solid-phase reactor; pharmaceuticals; cobalt;
D O I
10.1016/j.jpba.2005.02.044
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection analysis (FIA) manifold for the determination of penicillamine in pharmaceutical preparations is proposed. The manifold includes a solid-phase reactor for the in situ production of the derivatizing reagent, Co(II) ion, which forms a coloured complex with penicillamine in an alkaline medium. The reactor is prepared by natural immobilization of cobalt carbonate on a polymer matrix, which endows it with a high mechanical and microbiological stability. The cobalt released by passage of a 5 x 10(-4) Mol l(-1) sulphuric acid stream at a flow-rate of 2.3 ml min(-1) is merged with a volume of 314 mu l of sample containing penicillamine in ammonium-ammonia buffer at pH 9.5 to measure the absorbance at 360 nm. Beer's law is obeyed over the penicillamine concentration range 5-60 mg l(-1). The limit of detection (LOD) of the method is 1 mg l(-1) and its throughput 70 samples h(-1). (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:281 / 284
页数:4
相关论文
共 50 条