Fast photon counting CdTe detectors for diagnostic clinical CT: dynamic range, stability, and temporal response

被引:13
作者
Barber, William C. [1 ]
Nygard, Einar [2 ]
Wessel, Jan C. [2 ]
Malakhov, Nail [2 ]
Wawrzyniak, Gregor [2 ]
Hartsough, Neal E. [1 ]
Gandhi, Thulasi [1 ]
Iwanczyk, Jan S. [1 ]
机构
[1] DxRay Inc, 19355 Business Ctr Dr,Suite 10, Northridge, CA 91324 USA
[2] Interon AS, N-1395 Hvalstad, Norway
来源
MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING | 2010年 / 7622卷
基金
美国国家卫生研究院;
关键词
CdTe; ASIC; CT; photon counting; X-RAY; COMPUTED-TOMOGRAPHY; SYSTEM;
D O I
10.1117/12.844669
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report results from the development of a second-generation CdTe direct-conversion compound-semiconductor x-ray detector for photon-counting clinical CT. The first-generation detector has 512 pixels with a 1 mm pitch and is vertically integrated with the readout. A 32-row multi-slice CT system using first-generation detectors has been used for clinical low-dose CT applications. To provide adequate performance for whole-body diagnostic CT we have designed and fabricated new 0.25 mm(2) pixels to increase the maximum output to greater than 20 Mcps per mm(2) while preserving sufficient energy resolution for photon-counting CT. In addition to the need for dynamic range, CT places stringent uniformity and temporal response requirements on the detector. We have measured detector parameters including the dynamic range, energy resolution, noise floor, stability, and temporal response. Temporal response is determined by rapid cycling of the input flux with shutter driven attenuators. Cycling between high and low flux generates reproducible counts, within counting statistics, with a response time less than 1 ms. Stability is determined by measuring uniformity corrected flood images repeatedly over a time interval exceeding whole-body diagnostic CT scan times. Long exposure to uniform flux generates a number of counts which drift in some pixels slightly in excess of counting statistics. These results demonstrate the potential for these detectors to achieve whole-body CT.
引用
收藏
页数:8
相关论文
共 19 条
[1]   A System for the Characterization and Testing of CdZnTe/CdTe Pixel Detectors for X-ray and Gamma-ray Imaging [J].
Arodzero, Anatoli ;
Barber, William C. ;
Damron, Matthew Q. ;
Hartsough, Neal E. ;
Iwanczyk, Jan S. ;
Malakhov, Nail ;
Nygard, Einar ;
Moraes, Danielle ;
Weilhammer, Peter ;
Jarron, Pierre .
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, :3638-3642
[2]  
Barber WC, 2006, IEEE NUCL SCI CONF R, P2414
[3]   Current mode operation of a Cd0.9Zn0.1Te detector for CT imaging [J].
Barber, WC ;
Iwata, K ;
Hasegawa, BH ;
Bennett, PR ;
Cirignano, LJ ;
Shah, KS .
PENETRATING RADIATION SYSTEMS AND APPLICATIONS IV, 2002, 4786 :144-150
[4]   A digital system based on a bi-chromatic X ray source and a single photon counting device: a single exposure dual energy mammography approach [J].
Bisogni, MG ;
Fabbri, S ;
Gambaccini, M ;
Marziani, M ;
Novelli, A ;
Quattrocchi, M ;
Rosso, V ;
Russo, P ;
Sarnelli, A ;
Stefanini, A ;
Zucca, S .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :620-632
[5]   Detective quantum efficiency dependence on x-ray energy weighting in mammography [J].
Cahn, RN ;
Cederström, B ;
Danielsson, M ;
Hall, A ;
Lundqvist, M ;
Nygren, D .
MEDICAL PHYSICS, 1999, 26 (12) :2680-2683
[6]   First experimental tests with a CdTe photon counting pixel detector hybridized with a Medipix2 readout chip [J].
Chmeissani, M ;
Frojdh, C ;
Gal, O ;
Llopart, X ;
Ludwig, J ;
Maiorino, M ;
Manach, E ;
Mettivier, G ;
Montesi, MC ;
Ponchut, C ;
Russo, P ;
Tlustos, L ;
Zwerger, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2379-2385
[7]  
IWANCZYK JS, 2009, T NUCL SCI, V56, P1820
[8]  
JOHNS PC, 2002, P SOC PHOTO-OPT INS, P367
[9]   Dose reduction in pelvimetry using a digital technique [J].
Kheddache, S ;
Kullenberg, R ;
Kivilo-Carlsson, E .
RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) :275-278
[10]   Evaluation of a photon-counting breast tomosynthesis imaging system [J].
Maidment, A ;
Albert, M ;
Thunberg, S ;
Adelöw, L ;
Blom, O ;
Egerström, J ;
Eklund, M ;
Francke, T ;
Jordung, U ;
Kristoffersson, T ;
Lindman, K ;
Lindqvist, L ;
Marchal, D ;
Olla, H ;
Penton, E ;
Rantanen, J ;
Solokov, S ;
Ullberg, C ;
Weber, N .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :572-582