Mechanisms of rupture of knitted polyester vascular prostheses:: An in vitro analysis of virgin prostheses

被引:23
作者
Dieval, F
Chakfé, N
Wang, L
Riepe, G
Thaveau, F
Heintz, C
Mathieu, D
Le Magnen, JF
Kretz, JG
Durand, B
机构
[1] Les Hop Univ Strasbourg, Dept Vasc Surg, F-67091 Strasbourg, France
[2] Lab Phys & Mecan Text, Mulhouse, France
[3] Allgemeines Krankenhaus Harburg, Dept Gen Thorac & Vasc Surg, Hamburg, Germany
关键词
blood vessel prosthesis; degradation; polyester; rupture; vascular grafts;
D O I
10.1016/S1078-5884(03)00257-0
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: Previous explant retrieval studies have shown ruptures occurring on the remeshing line and the guide line of two types of warp-knitted grafts. The aim of our study was to characterize the mechanisms these ruptures. Materials and methods: We performed an in vitro study of the mechanical and chemical characteristics of virgin prostheses. We studied 2 virgin polyester warp-knitted grafts models: the Cooley Double Velour and the Microvel Double Velour constructed by Meadox (USA), using the following techniques: characterization and de-knitting of the textile structure, circumferential tensile strength, filament dynamometry, critical dissolution time of the filaments and scanning electron microscopy. Results: Both prostheses were constructed in the same way but the texturized yarns of the Cooley graft included twice as many filaments (54) than the Microvel (27). There was more adsorbed tension in the Cooley structure than in the Microvel. The circumferential tensile strength test demonstrated that the Cooley graft always ruptured on the remeshing line and the Microvel graft always ruptured at the interface between the remeshing line and the standard line. Filament dynamometry demonstrated a heterogeneous behavior of the filaments inside the yarns, mainly at the remeshing line of the Cooley graft (27.1 cN/tex +/- 11.5% versus 26.1 cN/tex +/- 2.2% for the guide line and 28 cN/tex +/- 6.7% for the standard knit). Critical dissolution time of the filaments was significantly lower for the Microvel grafts (2.5 sec versus 17.2 sec for the Cooley). Conclusions: Rupture of knitted polyester prostheses are probably an underestimated phenomenon. They may occur at specific areas of the graft. Further studies are required to determine whether all grafts of this type are at risk.
引用
收藏
页码:429 / 436
页数:8
相关论文
共 17 条
[1]  
BERGER K, 1981, ANN SURG, V193, P477
[2]  
BLUMENBERG RM, 1977, SURGERY, V81, P493
[3]   Longitudinal ruptures of polyester knitted vascular prostheses [J].
Chakfe, N ;
Riepe, G ;
Dieval, F ;
Le Magnen, JF ;
Wang, L ;
Urban, E ;
Beaufigeau, M ;
Durand, B ;
Imig, H ;
Kretz, JG .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (05) :1015-1021
[4]  
COOKE PA, 1974, ARCH SURG-CHICAGO, V103, P101
[5]   Polyester fiber and microfiber structure by critical dissolution time [J].
Diéval, F ;
Mathieu, D ;
Viallier, P ;
Durand, B .
TEXTILE RESEARCH JOURNAL, 2001, 71 (03) :239-246
[6]  
Guidoin R, 1990, CURRENT THERAPY VASC, P324
[7]  
HUSSEY HH, 1976, JAMA-J AM MED ASSOC, V235, P848
[8]  
NUCHO RC, 1984, ARCH SURG-CHICAGO, V119, P1182
[9]   POST-OPERATIVE ALTERATIONS IN SIZE OF DACRON AORTIC GRAFTS - ULTRASONIC EVALUATION [J].
NUNN, DB ;
FREEMAN, MH ;
HUDGINS, PC .
ANNALS OF SURGERY, 1979, 189 (06) :741-745
[10]  
OTTINGER LW, 1976, ARCH SURG-CHICAGO, V111, P146