Placing a bone graft more posteriorly may reduce the risk of pedicle screw breakage: analysis of an unexpected case of pedicle screw breakage

被引:14
作者
Rohlmann, A
Bergmann, G
Graichen, F
Mayer, HM
机构
[1] Oskar Helene Heim, Biomech Lab, D-14195 Berlin, Germany
[2] Free Univ Berlin, Dept Orthopaed Surg, Oskar Helene Heim, Orthopaed Hosp, D-14195 Berlin, Germany
关键词
spine; internal fixation device; pedicle screw breakage; load measurement; bone graft;
D O I
10.1016/S0021-9290(98)00073-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Telemeterized internal spinal fixation devices were implanted in a patient with degenerative instability and a narrow spinal canal in order to measure the fixator loads during daily activities. Anterior interbody fusion was performed three weeks later. During walking, the typical maximum flexion bending moments were 10 N m in the left and 5 N m in the right fixator. On removal of the implants three months later, a fatigue fracture was found not on the high loaded left side but in the upper right pedicle screw. The crack started on the caudal side of the cross-sectional area and progressed cranially. Upper vertebral tilting in the sagittal plane must have caused the screw breakage. This would probably have been prevented by a more posteriorly placed bone graft. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:763 / 767
页数:5
相关论文
共 11 条
  • [1] BLUMENAUER H, 1987, WERKSTOFFPRUFUNG VEB, P60
  • [2] DICK W, 1989, INTERNAL FIXATION TH
  • [3] Patient monitoring system for load measurement with spinal fixation devices
    Graichen, F
    Bergmann, G
    Rohlmann, A
    [J]. MEDICAL ENGINEERING & PHYSICS, 1996, 18 (02) : 167 - 174
  • [4] A new microsurgical technique for minimally invasive anterior lumbar interbody fusion
    Mayer, HM
    [J]. SPINE, 1997, 22 (06) : 691 - 699
  • [5] Loads on an internal spinal fixation device during walking
    Rohlmann, A
    Bergmann, G
    Graichen, F
    [J]. JOURNAL OF BIOMECHANICS, 1997, 30 (01) : 41 - 47
  • [6] Rohlmann A, 1995, Eur Spine J, V4, P347, DOI 10.1007/BF00300295
  • [7] Telemeterized load measurement using instrumented spinal internal fixators in a patient with degenerative instability
    Rohlmann, A
    Bergmann, G
    Graichen, F
    Mayer, HM
    [J]. SPINE, 1995, 20 (24) : 2683 - 2689
  • [8] A SPINAL FIXATION DEVICE FOR IN-VIVO LOAD MEASUREMENT
    ROHLMANN, A
    BERGMANN, G
    GRAICHEN, F
    [J]. JOURNAL OF BIOMECHANICS, 1994, 27 (07) : 961 - &
  • [9] Influence of muscle forces on loads in internal spinal fixation devices
    Rohlmann, A
    Bergmann, G
    Graichen, F
    Mayer, HM
    [J]. SPINE, 1998, 23 (05) : 537 - 542
  • [10] YAHIRO MA, 1994, SPINE, V19, pS2297