The influence of misalignment on the uncertainty of vertical torque calibration machine

被引:7
作者
Khaled, K. M. [1 ]
Aggag, G. [1 ]
Abuelezz, A. E. [1 ]
Elsherbiny, M. G. [1 ]
机构
[1] Cairo Univ, Fac Engn, Cairo, Egypt
来源
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA | 2011年 / 26卷 / 02期
关键词
Torque; Optimum Alignment; Relative Expanded Uncertainty; Flexible Coupling; Radial Shift;
D O I
10.1007/s12647-011-0015-4
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Misalignment of torque calibration machines can cause parasitic bending loads inside the torque transducers in use. These parasitic loads may affect the machine uncertainty, specially for secondary torque calibration machines. Therefore, this study discusses the contribution of misalignment to the measurement uncertainty for a newly designed 3 kNm secondary vertical torque calibration machine. Radial axes shift up to 4 mm was applied which introduce small increased uncertainty.