Determination of iron and nickel by flame atomic absorption spectrophotometry after preconcentration on Saccharomyces cerevisiae immobilized sepiolite

被引:124
作者
Bag, H
Lale, M
Turker, AR [1 ]
机构
[1] Gazi Univ, Fen Edebiyat Fak, TR-06500 Ankara, Turkey
[2] Kirikkale Univ, Fen Edebiyat Fak, TR-71450 Kirikkale, Turkey
关键词
trace metal determination; preconcentration; Saccharomyces cerevisiae; immobilization; sepiolite; flame atomic absorption spectrometry;
D O I
10.1016/S0039-9140(98)00104-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Iron and nickel have been preconcentrated on Saccharomyces cerevisiae immobilized sepiolite and determined by flame atomic absorption spectrophotometry (FAAS). Preconcentration studies were conducted by the column method. Effect of pH, amount of adsorbent, elution solution, flow rate and interfering ions on the recovery of the analytes have been investigated. Recoveries of Fe and Ni were 95 +/- 1 and 99.5 +/- 0.1%, respectively, at 95% confidence level. The breakthrough capacities of analytes were also investigated and found to be 0.042 mmol g(-1) for Fe and 0.055 mmol g(-1) for Ni. The proposed method was applied to the determination of iron and nickel in brass (NBS SRM 37e). The detection limit of iron and nickel were found as 0.065 and 0.087 mu g ml(-1), respectively. The direct determination of trace metals by flame atomic absorption spectrometry (FAAS) is limited and difficult because of low concentration and/or matrix interferences. The proposed method is excellent for the determination of trace metal in matrixes, such as metal alloys. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:689 / 696
页数:8
相关论文
共 21 条