Customisation and validation of a low-volume plasma renin activity immunoassay: Enabling of regulatory compliant determination in paediatric trials

被引:2
作者
Suessenbach, F. K. [1 ]
Tins, J. [1 ]
Burckhardt, B. B. [1 ]
机构
[1] Heinrich Heine Univ, Inst Clin Pharm & Pharmacotherapy, Univ Str 1, D-40225 Dusseldorf, Germany
关键词
Plasma renin activity; Paediatric; Immunoassay; Pharmacodynamic; Low-volume; Validation; HEART-FAILURE; ALDOSTERONE; BLOOD; ASSAY; CARE;
D O I
10.1016/j.plabm.2019.e00144
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Investigations of plasma renin activity (PRA) in children are urgently required. Small-volume, regulatory guideline compliant, bioanalytical assays tailored for paediatric application could facilitate to overcome this hurdle. Ethical constraints given e.g. by the European Medicines Agency need to be addressed and reliable data generation in line with Good Clinical Laboratory Practice must be ensured. Methods: A PRA enzyme-linked immunosorbent assay (ELISA) was tailored for paediatric application and validated in the context of the U.S. Food and Drug Administration bioanalytical guideline. Performance verification of the assay was conducted by participation in an interlaboratory ring test, evaluation of incurred sample reanalysis and an application-orientated approach in children. Results: A five-fold reduction of required plasma volume to 100 mu L was achieved without limiting the calibration range. Between-run accuracy and precision varied no more than 5.0% and 6.3%, respectively. No substantial matrix effect was detected and the inter-run precision for parallelism was 11.1%. Stability experiments approved the freeze-thaw stability, short-term stability as well as 37 weeks of long-term stability. The assay successfully participated in the interlaboratory ring test, showing non-inferiority regarding radioimmunoassay (RIA). Moreover, PRA in plasma samples of neonates was successfully determined. Conducted incurred sample reanalysis confirmed the comparability and reliability of the assay with regard to international regulatory bioanalytical guidelines. Conclusion: A fit-for-purpose PRA ELISA characterised by low-volume application was successfully established, indicating non-inferiority regarding commonly applied RIAs. Reliability of the regulatory-compliant PRA assay was proven by participation in an interlaboratory ring test and its application in a paediatric population.
引用
收藏
页数:12
相关论文
共 30 条
[1]  
Abd-Allah NM, 2004, BIOL RES, V37, P95, DOI 10.4067/S0716-97602004000100010
[2]  
[Anonymous], 2011, Science Medicines Health, Guideline on bioanalytical method validation.
[3]  
[Anonymous], 2018, GUID IND BIOAN METH
[4]   Roles of Angiotensin Peptides and Recombinant Human ACE2 in Heart Failure [J].
Basu, Ratnadeep ;
Poglitsch, Marko ;
Yogasundaram, Haran ;
Thomas, Jissy ;
Rowe, Brian H. ;
Oudit, Gavin Y. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (07) :805-819
[5]   COMPARISON OF PLASMA-RENIN AND PRORENIN IN HEALTHY INFANTS AND CHILDREN AS DETERMINED WITH AN ENZYMATIC METHOD AND A NEW DIRECT IMMUNORADIOMETRIC ASSAY [J].
BLAZY, I ;
GUILLOT, F ;
LABORDE, K ;
DECHAUX, M .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1989, 49 (05) :413-418
[6]   A liquid chromatography tandem mass spectrometry assay for plasma renin activity using online solid-phase extraction [J].
Carter, S. ;
Owen, L. J. ;
Kerstens, M. N. ;
Dullaart, R. P. F. ;
Keevil, B. G. .
ANNALS OF CLINICAL BIOCHEMISTRY, 2012, 49 :570-579
[7]  
Chappell DL, 2012, BIOANALYSIS, V4, P2843, DOI [10.4155/BIO.12.268, 10.4155/bio.12.268]
[8]   A feasibility study prior to an international multicentre paediatric study to assess pharmacokinetic/pharmacodynamic sampling and sample preparation procedures, logistics and bioanalysis [J].
Ciplea, Agnes Maria ;
Laeer, Stephanie ;
Burckhardt, Bjoern Bengt .
CONTEMPORARY CLINICAL TRIALS COMMUNICATIONS, 2018, 12 :32-39
[9]  
DEMEDITEC Diagnostics GmbH, PLASM REN ACT PRA EL
[10]  
DiaSorin Inc, GAMMACOAT PLASM REN