Evaluation of a treatment planning system developed for clinical boron neutron capture therapy and validation against an independent Monte Carlo dose calculation system

被引:43
作者
Hu, Naonori [1 ,2 ]
Tanaka, Hiroki [2 ]
Kakino, Ryo [1 ]
Yoshikawa, Syuushi [1 ]
Miyao, Mamoru [3 ]
Akita, Kazuhiko [1 ]
Isohashi, Kayako [1 ]
Aihara, Teruhito [1 ]
Nihei, Keiji [1 ,4 ]
Ono, Koji [1 ]
机构
[1] Osaka Med & Pharmaceut Univ, Kansai BNCT Med Ctr, Osaka Fu Takatsuki Shi Daigakumachi 2-7, Takatsuki, Osaka, Japan
[2] Kyoto Univ, Inst Integrated Radiat & Nucl Sci, Kyoto, Japan
[3] Osaka Med & Pharmaceut Univ Hosp, Cent Dept Radiol, Takatsuki, Osaka, Japan
[4] Osaka Med & Pharmaceut Univ Hosp, Dept Radiat Oncol, Takatsuki, Osaka, Japan
关键词
Boron neutron capture therapy; Treatment planning system; Commissioning; Monte Carlo simulation; COMPUTATIONAL DOSIMETRY; VERIFICATION;
D O I
10.1186/s13014-021-01968-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Boron neutron capture therapy (BNCT) for the treatment of unresectable, locally advanced, and recurrent carcinoma of the head and neck cancer has been approved by the Japanese government for reimbursement under the national health insurance as of June 2020. A new treatment planning system for clinical BNCT has been developed by Sumitomo Heavy Industries, Ltd. (Sumitomo), NeuCure (R) Dose Engine(.) To safely implement this system for clinical use, the simulated neutron flux and gamma ray dose rate inside a water phantom was compared against experimental measurements. Furthermore, to validate and verify the new planning system, the dose distribution inside an anthropomorphic head phantom was compared against a BNCT treatment planning system SERA and an in-house developed Monte Carlo dose calculation program. The simulated results closely matched the experimental results, within 5% for the thermal neutron flux and 10% for the gamma ray dose rate. The dose distribution inside the head phantom closely matched with SERA and the in-house developed dose calculation program, within 3% for the tumour and a difference of 0.3 Gy(w) for the brain.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged Treatment Fields
    Raghavi, Sajjad
    Sadoughi, Hamid-Reza
    Ravari, Mohammad Ehsan
    Behmadi, Marziyeh
    JOURNAL OF MEDICAL SIGNALS & SENSORS, 2024, 14 (10):
  • [42] Evaluation of dose calculation accuracy of a commercial radiotherapy treatment planning system for adjacent radiation fields
    Rasouli, Ali
    Arani, Mahmud Naraqi
    Aliasgharzadeh, Akbar
    Farhood, Bagher
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2023, 22
  • [43] Comparison of PRIMO Monte Carlo code and Eclipse treatment planning system in calculation of dosimetric parameters in brain cancer radiotherapy
    Altuwayrish, Ali
    Ghorbani, Mahdi
    Bakhshandeh, Mohsen
    Roozmand, Zahra
    Hoseini-Ghahfarokhi, Mojtaba
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (05) : 863 - 874
  • [44] The accuracy of out of field dose calculations in commercial treatment planning system using GATE/GEANT4 Monte Carlo simulation
    Attalla, Ehab Marouf
    Sinousy, Dina Mohamed
    Ibrahim, Hassan Fathy
    Elhussiny, Fathi Ahmed
    Elmekawy, Ahmed Farouk
    RADIATION PHYSICS AND CHEMISTRY, 2023, 206
  • [46] Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer
    Komori, Shinya
    Hirose, Katsumi
    Sato, Mariko
    Yamazaki, Yuhei
    Takeuchi, Akihiko
    Kato, Ryohei
    Motoyanagi, Tomoaki
    Narita, Yuki
    Kato, Takahiro
    Takai, Yoshihiro
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 126
  • [47] Development of a dose distribution shifter to fit inside the collimator of a Boron Neutron Capture Therapy irradiation system to treat superficial tumours
    Hu, N.
    Tanaka, H.
    Yoshikawa, S.
    Miyao, M.
    Akita, K.
    Aihara, T.
    Ono, K.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 82 : 17 - 24
  • [48] Evaluation of monitor unit calculation based on measurement and calculation with a simplified Monte Carlo method for passive beam delivery system in proton beam therapy
    Hotta, Kenji
    Kohno, Ryosuke
    Nagafuchi, Kohsuke
    Yamaguchi, Hidenori
    Tansho, Ryohei
    Takada, Yoshihisa
    Akimoto, Tetsuo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05): : 228 - 238
  • [49] Independent dose verification system with Monte Carlo simulations using TOPAS for passive scattering proton therapy at the National Cancer Center in Korea
    Shin, Wook-Geun
    Testa, Mauro
    Kim, Hak Soo
    Jeong, Jong Hwi
    Lee, Se Byeong
    Kim, Yeon-Joo
    Min, Chul Hee
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (19) : 7598 - 7616
  • [50] Gel dosimetry measurements and Monte Carlo modeling for external radiotherapy photon beams comparison with a treatment planning system dose distribution
    Valente, M.
    Aon, E.
    Brunetto, A.
    Castellano, G.
    Gallivanone, F.
    Gambarini, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01) : 497 - 501