Evaluation of a treatment planning system developed for clinical boron neutron capture therapy and validation against an independent Monte Carlo dose calculation system

被引:43
|
作者
Hu, Naonori [1 ,2 ]
Tanaka, Hiroki [2 ]
Kakino, Ryo [1 ]
Yoshikawa, Syuushi [1 ]
Miyao, Mamoru [3 ]
Akita, Kazuhiko [1 ]
Isohashi, Kayako [1 ]
Aihara, Teruhito [1 ]
Nihei, Keiji [1 ,4 ]
Ono, Koji [1 ]
机构
[1] Osaka Med & Pharmaceut Univ, Kansai BNCT Med Ctr, Osaka Fu Takatsuki Shi Daigakumachi 2-7, Takatsuki, Osaka, Japan
[2] Kyoto Univ, Inst Integrated Radiat & Nucl Sci, Kyoto, Japan
[3] Osaka Med & Pharmaceut Univ Hosp, Cent Dept Radiol, Takatsuki, Osaka, Japan
[4] Osaka Med & Pharmaceut Univ Hosp, Dept Radiat Oncol, Takatsuki, Osaka, Japan
关键词
Boron neutron capture therapy; Treatment planning system; Commissioning; Monte Carlo simulation; COMPUTATIONAL DOSIMETRY; VERIFICATION;
D O I
10.1186/s13014-021-01968-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Boron neutron capture therapy (BNCT) for the treatment of unresectable, locally advanced, and recurrent carcinoma of the head and neck cancer has been approved by the Japanese government for reimbursement under the national health insurance as of June 2020. A new treatment planning system for clinical BNCT has been developed by Sumitomo Heavy Industries, Ltd. (Sumitomo), NeuCure (R) Dose Engine(.) To safely implement this system for clinical use, the simulated neutron flux and gamma ray dose rate inside a water phantom was compared against experimental measurements. Furthermore, to validate and verify the new planning system, the dose distribution inside an anthropomorphic head phantom was compared against a BNCT treatment planning system SERA and an in-house developed Monte Carlo dose calculation program. The simulated results closely matched the experimental results, within 5% for the thermal neutron flux and 10% for the gamma ray dose rate. The dose distribution inside the head phantom closely matched with SERA and the in-house developed dose calculation program, within 3% for the tumour and a difference of 0.3 Gy(w) for the brain.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Point dose comparison of Monte Carlo and pencil beam algorithms in treatment planning system for proton therapy
    Monkongsubsin, Wiroon
    Keawsamur, Mintra
    Sanghangthum, Taweap
    Ayuthaya, Isra Israngkul Na
    HEALTH AND TECHNOLOGY, 2024,
  • [22] Comparison of CCC and ETAR dose calculation algorithms in pituitary adenoma radiation treatment planning; Monte Carlo evaluation
    Tanha, K.
    Mahdavi, S. R.
    Geraily, G.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2014, 13 (04) : 447 - 455
  • [23] Experimental and Monte Carlo evaluation of Eclipse treatment planning system for effects on dose distribution of the hip prostheses
    Catli, Serap
    Tanir, Gunes
    MEDICAL DOSIMETRY, 2013, 38 (03) : 332 - 336
  • [24] Characterization and validation of a Monte Carlo code for independent dose calculation in proton therapy treatments with pencil beam scanning
    Fracchiolla, F.
    Lorentini, S.
    Widesott, L.
    Schwarz, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (21) : 8601 - 8619
  • [25] Boron neutron capture therapy: Dosimetric evaluation of a brain tumor and surrounding healthy tissues using Monte Carlo Simulation
    Tedjani, A.
    Abdi, R.
    Boumala, D.
    Belafrites, A.
    Groetz, J-E
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1040
  • [26] Clinical evaluation of performance, stability, and longevity of an accelerator system designed for boron neutron capture therapy utilising a beryllium target
    Hu, Naonori
    Kakino, Ryo
    Sasaki, Akinori
    Yoshikawa, Syuushi
    Akita, Kazuhiko
    Takeno, Satoshi
    Yoshino, Yuki
    Aihara, Teruhito
    Nihei, Keiji
    Nojiri, Mai
    Matsubayashi, Nishiki
    Takata, Takushi
    Tanaka, Hiroki
    Suzuki, Minoru
    Ono, Koji
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2025, 1072
  • [27] Experimental validation of a Monte Carlo-based treatment-planning system for electron beams
    Mika, Sabine
    Christ, Gunter
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 (03) : 150 - 156
  • [28] Experimental validation of a Monte Carlo-based treatment-planning system for electron beams
    Mika S.
    Christ G.
    Strahlentherapie und Onkologie, 2007, 183 (3) : 150 - 156
  • [29] Experimental validation of Monte Carlo based treatment planning system in bone density equivalent media
    Radojcic, Djeni Smilovic
    Casar, Bozidar
    Rajlic, David
    Kolacio, Manda Svabic
    Mendez, Ignasi
    Obajdin, Nevena
    Debeljuh, Dea Dundara
    Jurkovic, Slaven
    RADIOLOGY AND ONCOLOGY, 2020, 54 (04) : 495 - 504
  • [30] Evaluation of clinical IMRT treatment planning using the GATE Monte Carlo simulation platform for absolute and relative dose calculations
    Benhalouche, S.
    Visvikis, D.
    Le Maitre, A.
    Pradier, O.
    Boussion, N.
    MEDICAL PHYSICS, 2013, 40 (02)