Single crystal artificial diamond detectors for VUV and soft X-rays measurements on JET thermonuclear fusion plasma

被引:32
作者
Angelone, M. [1 ]
Pillon, M. [1 ]
Marinelli, Marco [2 ]
Milani, E. [2 ]
Prestopino, G. [2 ]
Verona, C. [2 ]
Verona-Rinati, G. [2 ]
Coffey, I. [3 ]
Murari, A. [4 ]
Tartoni, N. [5 ]
机构
[1] Assoc EURATOM ENEA Fus, Rome, Italy
[2] Univ Roma Tor Vergata, Dip Ing Meccan, Rome, Italy
[3] Queens Univ Belfast, Belfast BT7 1NN, Antrim, North Ireland
[4] Assoc EURATOM ENEA Fus, Consorzio RFX, Padua, Italy
[5] Diamond Light Source, Chilton OX11 0DE, Oxon, England
[6] Culham Sci Ctr, JET EFDA, Abingdon OX14 3DB, Oxon, England
关键词
CVD diamond; Diamond detectors; UVU and X-ray monitors; Plasma diagnostics;
D O I
10.1016/j.nima.2010.04.021
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Diamond appears to be a promising material for VUV and soft X-ray radiation detection. Its wide band-gap (5.5 eV) results in a very low leakage current (it can operate above room temperature) and its electronic properties (high carrier mobility) allow a fast time response. More importantly, it is optimally suited for harsh environment applications, like those in the JET Tokamak located at the Culham laboratory (UK). Its extreme radiation hardness is well known and another interesting feature, again related to the wide band-gap, is its selective sensitivity to radiation with wavelengths shorter than 225 nm (visible-blind detectors).We report on the performances of two photodetectors based on Chemical Vapor Deposition (CVD) single crystal diamonds, one optimized for extreme UV detection, the other for soft X-ray radiation detection in the 0.8-8 keV range. These detectors have been fabricated at Roma "Tor Vergata" University using a p-type/intrinsic/metal configuration and they behave like photodiodes allowing operation with no external applied voltage. They have been installed on JET inside a vacuum chamber with a direct horizontal view of JET plasma without any wavelength selection. Their low thickness, low sensitivity to gamma ray and the unbiased operation mode make both detectors ideal for a Tokamak environment. The measurements routinely performed at JET show a low intrinsic dark current (similar to 0.01 pA) and very high signal to noise ratio (50 dB). Both detectors show a fast response and their signals are acquired using an electronic chain and ADC able to operate at 200 kHz, providing very interesting results for MHD and Edge Localized Modes (ELMS) instability studies on fusion plasmas. (C) 2010 EURATOM. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:726 / 730
页数:5
相关论文
共 6 条
[1]  
*23X 5 MONT CARL T, 2003, LAUR031987 23X 5 MON
[2]   Extreme UV photodetectors based on CVD single crystal diamond in a p-type/intrinsic/metal configuration [J].
Almaviva, S. ;
Marinelli, Marco ;
Milani, E. ;
Prestopino, G. ;
Tucciarone, A. ;
Verona, C. ;
Verona-Rinati, G. ;
Angelone, M. ;
Pillon, M. .
DIAMOND AND RELATED MATERIALS, 2009, 18 (01) :101-105
[3]  
ALMAVIVA S, J APPL PHYS, V107
[4]  
GIROUD, 2009, 12 INT WORKSH H MOD
[5]  
Gonçalves B, 2003, PLASMA PHYS CONTR F, V45, P1627, DOI 10.1088/0741-3335/45/9/305
[6]   Diamond metallization for device applications [J].
Werner, M .