High frequency circular translation pin-on-disk method for accelerated wear testing of ultrahigh molecular weight polyethylene as a bearing material in total hip arthroplasty

被引:9
作者
Saikko, Vesa [1 ]
机构
[1] Aalto Univ, Sch Engn, Dept Engn Design & Prod, FI-00076 Espoo, Finland
关键词
Accelerated wear testing; Circular translation; High frequency; Pin-on-disk; UHMWPE; ORTHOPEDIC BIOMATERIALS; MULTIDIRECTIONAL MOTION; TRACK ANALYSIS; SIMULATION; DEVICE; DEBRIS;
D O I
10.1016/j.jbiomech.2014.11.035
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The temporal change of the direction of sliding relative to the ultrahigh molecular weight polyethylene (UHMWPE) component of prosthetic joints is known to be of crucial importance with respect to wear. One complete revolution of the resultant friction vector is commonly called a wear cycle. It was hypothesized that in order to accelerate the wear test, the cycle frequency may be substantially increased if the circumference of the slide track is reduced in proportion, and still the wear mechanisms remain realistic and no overheating takes place. This requires an additional slow motion mechanism with which the lubrication of the contact is maintained and wear particles are conveyed away from the contact. A three-station, dual motion high frequency circular translation pin-on-disk (HF-CTPOD) device with a relative cycle frequency of 25.3 Hz and an average sliding velocity of 27.4 mm/s was designed. The pins circularly translated at high frequency (1.0 mm per cycle, 24.8 Hz, clockwise), and the disks at low frequency (31.4 mm per cycle, 0.5 Hz, counter-clockwise). In a 22 million cycle (10 day) test, the wear rate of conventional gamma-sterilized UHMWPE pins against polished CoCr disks in diluted serum was 1.8 mg per 24 h, which was six times higher than that in the established 1 Hz CTPOD device. The wear mechanisms were similar. Burnishing of the pin was the predominant feature. No overheating took place. With the dual motion HF-CTPOD method, the wear testing of UHMWPE as a bearing material in total hip arthroplasty can be substantially accelerated without concerns of the validity of the wear simulation. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:401 / 404
页数:4
相关论文
共 28 条
[1]   Tribology and total hip joint replacement: Current concepts in mechanical simulation [J].
Affatato, S. ;
Spinelli, A. ;
Zavalloni, M. ;
Mazzega-Fabbro, C. ;
Viceconti, A. .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (10) :1305-1317
[2]  
[Anonymous], 2009, 142423 ISO
[3]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[4]   A new pin-on-disk wear testing method for simulating wear of polyethylene on cobalt-chrome alloy in total hip arthroplasty [J].
Bragdon, CR ;
O'Connor, DO ;
Lowenstein, JD ;
Jasty, M ;
Biggs, SA ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2001, 16 (05) :658-665
[5]   Force track analysis of contemporary hip simulators [J].
Calonius, O ;
Saikko, V .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1719-1726
[6]   The quantification of physiologically relevant cross-shear wear phenomena on orthopedic bearing materials using the MAX-shear wear testing system [J].
Gevaert, MR ;
LaBerge, M ;
Gordon, JM ;
DesJardins, JD .
JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2005, 127 (04) :740-749
[7]  
Hall R M, 1996, Proc Inst Mech Eng H, V210, P197, DOI 10.1243/PIME_PROC_1996_210_413_02
[8]   Development of a BiotriboPOD testing methodology for the wear evaluation of orthopaedic biomaterials [J].
Hua, Zikai ;
Zhang, Huanhuan ;
Fan, Yongwei ;
Jin, Zhongmin .
RSC ADVANCES, 2014, 4 (38) :19987-19991
[9]  
Joyce TJ, 2000, BIO-MED MATER ENG, V10, P241
[10]  
Kurtz SM, 2009, UHMWPE BIOMATERIALS HANDBOOK: ULTRA-HIGH MOLECULAR WEIGHT POLYETHYLENE IN TOTAL JOINT REPLACEMENT AND MEDICAL DEVICES, 2ND EDITION, P1