Evaluation and comparison of two stuart's liquid swab transport systems tested by the CLSI M40 method

被引:7
作者
Van Horn, K. G.
Rankin, I.
机构
[1] Focus Diagnost Inc, Cypress, CA 90630 USA
[2] Mt Sinai Med Ctr, Dept Pathol, New York, NY 10029 USA
关键词
D O I
10.1007/s10096-007-0328-y
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The Copan Diagnostics HealthLink Transporter swab and the Remel Bacti-Swab were evaluated by the Clinical and Laboratory Standards Institute M40-A quantitative swab elution method for maintenance of bacterial viability of six ATCC and six clinical strains. Acceptable levels of viability according to Clinical and Laboratory Standards Institute criteria were maintained for all organisms tested with the HealthLink Transporter swab and for 8 of 12 organisms tested with the Bacti-Swab. In this study, the HealthLink Transporter swab maintained organism viability better than the Bacti-Swab.
引用
收藏
页码:583 / 586
页数:4
相关论文
共 14 条
[1]   Evaluation of four commercial transport media for the survival of Neisseria Gonorrhoeae [J].
Arbique, JC ;
Forward, KR ;
LeBlanc, J .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2000, 36 (03) :163-168
[2]  
Bamberg R, 2002, WOUNDS, V14, P314
[3]   Evaluation of Copan swabs with liquid transport media for use in the Gen-Probe Group A Strep Direct Test [J].
Bourbeau, PP ;
Heiter, BJ .
JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (06) :2686-2689
[4]   Survival of aerobic and anaerobic bacteria in purulent clinical specimens maintained in the Copan Venturi Transystem and Becton Dickinson Port-a-Cul transport systems [J].
Citron, DM ;
Warren, YA ;
Hudspeth, MK ;
Goldstein, EJC .
JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (02) :892-894
[5]  
*CLIN LAB STAND I, 2003, QUAL CONTR MICR TRAN
[6]   Comparison of easy-flow Copan Liquid Stuart's and Starplex swab transport systems for recovery of fastidious aerobic bacteria [J].
Drake, C ;
Barenfanger, J ;
Lawhorn, J ;
Verhulst, S .
JOURNAL OF CLINICAL MICROBIOLOGY, 2005, 43 (03) :1301-1303
[7]   Efficacy of a swab transport system in maintaining viability of Neisseria gonorrhoeae and Streptococcus pneumoniae [J].
Farhat, SE ;
Thibault, M ;
Devlin, R .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (08) :2958-2960
[8]   Survival of Nisseria gonorrhoeae isolates of different auxotypes in six commercial transport systems [J].
Graver, MA ;
Wade, JJ .
JOURNAL OF CLINICAL MICROBIOLOGY, 2004, 42 (10) :4803-4804
[9]   Comparison of three transport systems (Starplex StarSwab II, the new Copan Vi-Pak Amies Agar Gel collection and transport swabs, and BBL Port-A-Cul) for maintenance of anaerobic and fastidious aerobic organisms [J].
Hindiyeh, M ;
Acevedo, V ;
Carroll, KC .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (01) :377-380
[10]   Evaluation of 4 swab transport systems for the recovery of ATCC and clinical strains with characterized resistance mechanisms [J].
Morosini, Maria-Isabel ;
Loza, Elena ;
Gutierrez, Olga ;
Almaraz, Felisa ;
Baquero, Fernando ;
Canton, Rafael .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2006, 56 (01) :19-24