Plasma-enhanced chemical vapor deposited silicon carbide as an implantable dielectric coating

被引:122
作者
Cogan, SF [1 ]
Edell, DJ
Guzelian, AA
Liu, YP
Edell, R
机构
[1] EIC Labs Inc, Norwood, MA 02062 USA
[2] InnerSea Technol, Lexington, MA 02421 USA
关键词
passivation; dielectrics; silicon carbide; stability; biocompatibility;
D O I
10.1002/jbm.a.10152
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Amorphous silicon carbide (a-SiC) films, deposited by plasma-enhanced chemical vapor deposition (PECVD), have been evaluated as insulating coatings for implantable microelectrodes. The a-SiC was deposited on platinum or iridium wire for measurement of electrical leakage through the coating in phosphate-buffered saline (PBS, pH 7.4). Low leakage currents of <10(-11) A were observed over a +/-5-V bias. The electronic resistivity of a-SiC was 3 x 10(13) Omega-cm. Dissolution rates of a-SiC in PBS at 37 and 90degreesC were determined from changes in infrared absorption band intensities and compared with those of silicon nitride formed by low-pressure chemical vapor deposition (LPCVD). Dissolution rates of LPCVD silicon nitride were 2 nm/h and 0.4 nm/day at 90 and 37degreesC, respectively, while a-SiC had a dissolution rate of 0.1 nm/h at 90degreesC and no measurable dissolution at 37degreesC. Biocompatibility was assessed by implanting a-SiC-coated quartz discs in the subcutaneous space of the New Zealand White rabbit. Histological evaluation showed no chronic inflammatory response and capsule thickness was comparable to silicone or uncoated quartz controls. Amorphous SiC-coated microelectrodes were implanted in the parietal cortex for periods up to 150 days and the cortical response evaluated by histological evaluation of neuronal viability at the implant site. The a-SiC was more stable in physiological saline than LPCVD Si,N, and well tolerated in the cortex. (C) 2003 Wiley Periodicals, Inc.
引用
收藏
页码:856 / 867
页数:12
相关论文
共 29 条
[1]  
ALLEN M, 1995, BIO-MED MATER ENG, V5, P151
[2]   BATCH-FABRICATED THIN-FILM ELECTRODES FOR STIMULATION OF THE CENTRAL AUDITORY-SYSTEM [J].
ANDERSON, DJ ;
NAJAFI, K ;
TANGHE, SJ ;
EVANS, DA ;
LEVY, KL ;
HETKE, JF ;
XUE, XL ;
ZAPPIA, JJ ;
WISE, KD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (07) :693-704
[3]   VISUAL SENSATIONS PRODUCED BY INTRACORTICAL MICROSTIMULATION OF THE HUMAN OCCIPITAL CORTEX [J].
BAK, M ;
GIRVIN, JP ;
HAMBRECHT, FT ;
KUFTA, CV ;
LOEB, GE ;
SCHMIDT, EM .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1990, 28 (03) :257-259
[4]   CHARGE INJECTION LIMITS OF ACTIVATED IRIDIUM OXIDE ELECTRODES WITH 0.2MS PULSES IN BICARBONATE BUFFERED SALINE [J].
BEEBE, X ;
ROSE, TL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (06) :494-495
[5]   Vagus nerve stimulation for epilepsy: a review [J].
Binnie, CD .
SEIZURE-EUROPEAN JOURNAL OF EPILEPSY, 2000, 9 (03) :161-169
[6]   THE FIRST 500 PATIENTS WITH SACRAL ANTERIOR ROOT STIMULATOR IMPLANTS - GENERAL DESCRIPTION [J].
BRINDLEY, GS .
PARAPLEGIA, 1994, 32 (12) :795-805
[7]   NOVEL PASSIVATION DIELECTRICS - THE BORON-DOPED OR PHOSPHORUS-DOPED HYDROGENATED AMORPHOUS-SILICON CARBIDE FILMS [J].
CHANG, CY ;
FANG, YK ;
HUANG, CF ;
WU, BS .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1985, 132 (02) :418-422
[8]  
Clark Graeme M., 1998, Auris Nasus Larynx, V25, P73, DOI 10.1016/S0385-8146(97)10030-X
[9]  
HAMBRECHT FT, 1995, BAILLIERE CLIN NEUR, V4, P147
[10]   Biostability of micro-photodiode arrays for subretinal implantation [J].
Hämmerle, H ;
Kobuch, K ;
Kohler, K ;
Nisch, W ;
Sachs, H ;
Stelzle, M .
BIOMATERIALS, 2002, 23 (03) :797-804