Three-dimensional dose verification for intensity-modulated radiation therapy in the radiological physics centre head-and-neck phantom using optical computed tomography scans of ferrous xylenol-orange gel dosimeters

被引:48
作者
Babic, Steven [1 ,2 ]
Battista, Jerry [1 ,2 ]
Jordan, Kevin [1 ,2 ]
机构
[1] London Hlth Sci Ctr, London Reg Canc Program, Dept Phys & Engn, London, ON N6A 4L6, Canada
[2] Univ Western Ontario, Dept Med Biophys, London, ON, Canada
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2008年 / 70卷 / 04期
关键词
ferrous xylenol-orange gel; optical computed tomography; three-dimensional dosimetry; anthropomorphic; RPC head phantom; intensity-modulated radiation therapy;
D O I
10.1016/j.ijrobp.2007.11.032
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To extend the Radiological Physics Centre (RPC) intensity-modulated radiation therapy dose verification protocol to three dimensions using optical computed tomography (CT) scans of ferrous xylenol-orange (FX) gels. Methods and Materials: The dosimetry insert in the RPC head-and-neck phantom was replaced with an FX cylindrical gel dosimeter. Two gels were calibrated, independently irradiated with 6-MV X-ray beams and scanned using laser and cone-beam (Vista(TM)) optical CT, respectively. For matching dose slices, measured dose distributions were compared with Pinnacle 3 computed distributions. Results: Within high-dose regions and low gradients, doses measured using laser CT were 2% to 3% less than the computed dose, whereas with cone-beam CT they were 4% to 5% less. Inside the central 90% of the gel cylinder diameter, the fraction of voxels satisfying the two-dimensional gamma analysis (5% dose difference, 3-mm distance to agreement) with laser-beam- and cone-beam-measured dose distributions were 98.4% and 99.0%, respectively. A three-dimensional gamma analysis with cone-beam data revealed that 96.7% of voxels within the central 90% gel volume satisfied the above criteria. Within the axial and sagittal planes through the primary planning target volume (PTV), computed and measured doses using GAFChromic(R)EBT film (RPC measured) and cone-beam scanned FX gel generally agreed. At equivalent points in the planning target volumes, computed, thermoluminescent dosimeter (RPC-measured), and gel point doses agreed to within 5.1% in absolute dose. Conclusions: Laser and cone-beam CT yield comparable dose distributions in high-dose regions. The RPC head phantom and optical CT-scanned FX gels can be used for accurate intensity-modulated radiation therapy dose verification in three dimensions. (C) 2008 Elsevier Inc.
引用
收藏
页码:1281 / 1291
页数:11
相关论文
共 21 条
[1]  
Babic S., 2006, Journal of Physics: Conference Series, V56, P304, DOI 10.1088/1742-6596/56/1/056
[2]   Experimental determination of the diffusion coefficient in two-dimensions in ferrous sulphate gels using the finite element method [J].
Baldock C. ;
Harris P.J. ;
Piercy A.R. ;
Healy B. .
Australasian Physics & Engineering Sciences in Medicine, 2001, 24 (1) :19-30
[3]   Radiochromic gel dosemeter for three-dimensional dosimetry [J].
Bero, MA ;
Gilboy, WB ;
Glover, PM .
RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) :433-435
[4]  
De Deene Y, 2002, PHYS MED BIOL, V47, P2459, DOI 10.1088/0031-9155/47/14/307
[5]   Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels .1. Development of an optical scanner [J].
Gore, JC ;
Ranade, M ;
Maryanski, MJ ;
Schulz, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12) :2695-2704
[6]  
IBBOTT G, 2003, P INT S STAND COD PR, P361
[7]   Independent evaluations of IMRT through the use of an anthropomorphic phantom [J].
Ibbott, Geoffrey S. ;
Molineu, Andrea ;
Followill, David S. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2006, 5 (05) :481-487
[8]  
IBBOTT GS, 2006, J PHYS C SER, V56, P237
[9]   Initial evaluation of commercial optical CT-based 3D gel dosimeter [J].
Islam, KTS ;
Dempsey, JF ;
Ranade, MK ;
Maryanski, MJ ;
Low, DA .
MEDICAL PHYSICS, 2003, 30 (08) :2159-2168
[10]  
JORDAN K, 2006, J PHYS C SER, V56, P141