Effect of Autoclaving Cycles on the Cyclic Fatigue Resistance of Race and Race Evo Nickel-Titanium Endodontic Rotary Files: An In Vitro Study

被引:13
作者
Almohareb, Rahaf A. [1 ]
Barakat, Reem [1 ]
Albakri, Aroob [2 ]
Altamimi, Manal [2 ]
机构
[1] Princess Nourah Bint Abdulrahman Univ, Coll Dent, Clin Dent Sci Dept, Riyadh 11671, Saudi Arabia
[2] Princess Nourah Bint Abdulrahman Univ, Coll Dent, Riyadh 11671, Saudi Arabia
关键词
autoclave; cyclic fatigue; heat treatment; NiTi; race; scanning electron microscope; SHAPE-MEMORY; ROTATIONAL SPEED; ROOT CANALS; INSTRUMENTS; STERILIZATION; PROTAPER; IMMERSION; BEHAVIOR;
D O I
10.3390/met11121947
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Objective: The aim of this study was to evaluate the influence of autoclave sterilization on the resistance to cyclic fatigue of two nickel-titanium (NiTi) endodontic files of identical design and taper, but with different NiTi alloy treatments: the newly introduced heat-treated Race Evo and the electropolished Race files. Materials and methods: Fifteen Race (25/0.06) files and fifteen Race Evo (25/0.06) files (n = 30 in total) were randomly assigned to five sub-groups each consisting of three files of the same NiTi alloy treatment. One group served as a control with files unautoclaved. The four remaining groups were sterilized in a steam sterilizer for 1, 3, 5, and 10 autoclave cycles, respectively. Files then underwent cyclic fatigue testing in a simulated metal canal block. A scanning electron microscope was used to inspect the surface of the fractured instruments. Statistical analysis was conducted using independent t-test and multi-factorial analysis of variance with significance set at a p value of <= 0.05. Results: Both Race Evo and Race files showed no significant difference between the different autoclaving cycles in terms of the number of cycles to fracture (p = 0.232 and p = 0.359). Despite rotating at a higher speed, the number of cycles to fracture of heat-treated Race Evo files was significantly higher than that of Race files (p <= 0.0001). Conclusion: Autoclave sterilization has no significant effect on the resistance to cyclic fatigue of heat-treated Race Evo or electropolished Race files. However, Race Evo files showed superior resistance to cyclic fatigue irrespective of autoclaving cycles.
引用
收藏
页数:8
相关论文
共 29 条
[1]  
Azimi Shahram, 2011, Iran Endod J, V6, P80
[2]   Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization [J].
Bulem, Ureyen Kaya ;
Kececi, Ayse Diljin ;
Guldas, Hilmi Egemen .
JOURNAL OF APPLIED ORAL SCIENCE, 2013, 21 (06) :505-510
[3]  
Champa C, 2017, J Conserv Dent, V20, P194, DOI 10.4103/0972-0707.218307
[4]   Investigation on the fracture Behavior of shape memory alloy NiTi [J].
Chen, JH ;
Sun, W ;
Wang, GZ .
METALLURGICAL AND MATERIALS TRANSACTIONS A-PHYSICAL METALLURGY AND MATERIALS SCIENCE, 2005, 36A (04) :941-955
[5]   Comparison of fracture rate, deformation rate, and efficiency between rotary endodontic instruments driven at 150 rpm and 350 rpm [J].
Daugherty, DW ;
Gound, TG ;
Comer, TL .
JOURNAL OF ENDODONTICS, 2001, 27 (02) :93-95
[6]  
Lopes NID, 2017, MATER RES-IBERO-AM J, V20, P572, DOI [10.1590/1980-5373-mr-2016-0933, 10.1590/1980-5373-MR-2016-0933]
[7]   Phase Transformation Behavior and Resistance to Bending and Cyclic Fatigue of ProTaper Gold and ProTaper Universal Instruments [J].
Hieawy, Ahmed ;
Haapasalo, Markus ;
Zhou, Huimin ;
Wang, Zhe-jun ;
Shen, Ya .
JOURNAL OF ENDODONTICS, 2015, 41 (07) :1134-1138
[8]   Multiple Autoclave Cycle Effects on Cyclic Fatigue of Nickel-Titanium Rotary Files Produced by New Manufacturing Methods [J].
Hilfer, Paul B. ;
Bergeron, Brian E. ;
Mayerchak, Michael J. ;
Roberts, Howard W. ;
Jeansonne, Billie G. .
JOURNAL OF ENDODONTICS, 2011, 37 (01) :72-74
[9]   Evaluation of the surface characteristics of used and new ProTaper instruments: An atomic force microscopy study [J].
Inan, Ugur ;
Aydin, Cumhur ;
Uzun, Ozgur ;
Topuz, Ozgur ;
Alacam, Tayfun .
JOURNAL OF ENDODONTICS, 2007, 33 (11) :1334-1337
[10]   Failure analysis of NiTi wires used in medical applications [J].
James B. ;
Foulds J. ;
Eiselstein L. .
Journal of Failure Analysis and Prevention, 2005, 5 (5) :82-87