Verification procedure for isocentric alignment of proton beams

被引:23
作者
Ciangaru, George [1 ]
Yang, James N. [1 ]
Oliver, Patrick J. [1 ]
Bues, Martin [1 ]
Zhu, Mengping [1 ]
Nakagawa, Fumio [2 ]
Chiba, Hitoshi [2 ]
Nakamura, Shin [2 ]
Yoshino, Hirofumi [2 ]
Umezawa, Mosumi [2 ]
Smith, Alfred R. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Photon Therapy Ctr, Dept Radiat Phys, Houston, TX USA
[2] Hitachi Amer, Power Syst DIv, Houston, TX USA
关键词
proton beam therapy; radiation cancer therapy; gantry; isocenter; alignment; film dosimetry; Gafchromic EBT film;
D O I
10.1120/jacmp.v8i4.2671
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present a technique-based on the Lutz, Winston, and Maleki test used in stereotactic linear accelerator radiosurgery-for verifying whether proton beams are being delivered within the required spatial coincidence with the gantry mechanical isocenter. Our procedure uses a proton beam that is collimated by a circular aperture at its central axis and is then intercepted by a small steel sphere rigidly supported by the patient couch. A laser tracker measurement system and a correction algorithm for couch position assures precise positioning of the steel sphere at the mechanical isocenter of the gantry. A film-based radiation dosimetry technique, chosen for the good spatial resolution it achieves, records the proton dose distribution for optical image analysis. The optical image obtained presents a circular high-dose region surrounding a lower-dose area corresponding to the proton beam absorption by the steel sphere, thereby providing a measure of the beam alignment with the mechanical isocenter. We found the self-developing Gafchromic EBT film (International Specialty Products, Wayne, NJ) and commercial Epson 10000 XL flatbed scanner (Epson America, Long Beach, CA) to be accurate and efficient tools. The positions of the gantry mechanical and proton beam isocenters, as recorded on film, were clearly identifiable within the scanning resolution used for routine alignment testing (0.17 mm per pixel). The mean displacement of the collimated proton beam from the gantry mechanical isocenter was 0.22 +/- 0.1 mm for the gantry positions tested, which was well within the maximum deviation of 0.50 mm accepted at the Proton Therapy Center in Houston.
引用
收藏
页码:65 / 75
页数:11
相关论文
共 10 条
[1]   Verification of the alignment of a therapeutic radiation beam relative to its patient positioner [J].
Barkhof, J ;
Schut, G ;
Flanz, JB ;
Goitein, M ;
Schippers, JM .
MEDICAL PHYSICS, 1999, 26 (11) :2429-2437
[2]  
CHILDRESS NL, 2004, DOSELAB V 3 09 DOSE
[3]   Benchmarking analytical calculations of proton doses in heterogeneous matter [J].
Ciangaru, G ;
Polf, JC ;
Bues, M ;
Smith, AR .
MEDICAL PHYSICS, 2005, 32 (12) :3511-3523
[4]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[5]  
FLANDRA C, 2006, MED PHYS, V33, P4314
[6]   A SYSTEM FOR STEREOTACTIC RADIOSURGERY WITH A LINEAR-ACCELERATOR [J].
LUTZ, W ;
WINSTON, KR ;
MALEKI, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :373-381
[7]   Characterization and real-time optical measurements of the ionizing radiation dose response for a new radiochromic medium [J].
Rink, A ;
Vitkin, IA ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (08) :2510-2516
[8]   Evaluation of GafChromic EBT prototype B for external beam dose verification [J].
Todorovic, M. ;
Fischer, M. ;
Cremers, F. ;
Thom, E. ;
Schmidt, R. .
MEDICAL PHYSICS, 2006, 33 (05) :1321-1328
[9]   Comparison of the Epson Expression 1680 flatbed and the Vidar VXR-16 dosimetry PRO™ film scanners for use in IMRT dosimetry using gafchromic and radiographic film [J].
Wilcox, Ellen ;
Daskalov, George ;
Nedialkova, Lucy .
MEDICAL PHYSICS, 2007, 34 (01) :41-48
[10]   Characterization and use of EBT radiochromic film for IMRT dose verification [J].
Zeidan, Omar A. ;
Stephenson, Stacy Ann L. ;
Meeks, Sanford L. ;
Wagner, Thomas H. ;
Willoughby, Twyla R. ;
Kupelian, Patrick A. ;
Langen, Katja M. .
MEDICAL PHYSICS, 2006, 33 (11) :4064-4072