Implantable cardioverter-defibrillators and the pathologist: Comment and cautionary notes

被引:0
作者
Walley, VM
Bourke, ME
Green, M
Stinson, WA
Veinot, JP
机构
[1] Ottawa Civic Hosp, Dept Lab Med, Ottawa, ON K1Y 4E9, Canada
[2] Univ Ottawa, Inst Heart, Dept Pathol & Lab Med, Ottawa, ON, Canada
[3] Univ Ottawa, Inst Heart, Dept Anaesthesia, Ottawa, ON, Canada
[4] Univ Ottawa, Inst Heart, Dept Internal Med, Ottawa, ON, Canada
关键词
forensic science; forensic pathology; sudden death; heart; arrhythmias; pathology; defibrillators; pacemakers; prevention; treatment;
D O I
暂无
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
This paper briefly reviews the components of, the clinical uses of, the techniques to place, and the complications related to implantable cardioverter-defibrillators (ICDs). Information useful in the specific identification of ICDs is presented. A series of recommendations for the autopsy examination or postmortem explantation of ICDs by the pathologist is given. Because of the serious risk of injury to the pathologist possible with postmortem discharges of ICDs which have not ben deactivated, and because of the risk of device explosion if the ICD is incinerated, a number of cautionary notes are provided. A brief case with occurrence of accidental postmortem discharge of an active ICD is also presented.
引用
收藏
页码:969 / 973
页数:5
相关论文
共 11 条
[1]  
AHERN TS, 1994, J AM COLL CARDIOL, V23, P1521
[2]  
ASHTAR M, 1993, PACE, V16, P511
[3]   IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS [J].
BARDY, GH ;
HOFER, B ;
JOHNSON, G ;
KUDENCHUK, PJ ;
POOLE, JE ;
DOLACK, GL ;
GLEVA, M ;
MITCHELL, R ;
KELSO, D .
CIRCULATION, 1993, 87 (04) :1152-1168
[4]   THE NASPE BPEG DEFIBRILLATOR CODE [J].
BERNSTEIN, AD ;
CAMM, AJ ;
FISHER, JD ;
FLETCHER, RD ;
MEAD, RH ;
NATHAN, AW ;
PARSONNET, V ;
RICKARDS, AF ;
SMYTH, NPD ;
SUTTON, R ;
TARJAN, PP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (09) :1776-1780
[5]  
Bourke Michael E., 1996, Canadian Journal of Anaesthesia, V43, pR24
[6]  
DAVIDSON T, 1994, HEART LUNG, V23, P205
[7]   SINGLE-INCISION IMPLANTATION OF CARDIOVERTER-DEFIBRILLATORS USING NONTHORACOTOMY LEAD SYSTEMS [J].
HAMMEL, D ;
BLOCK, M ;
GEIGER, A ;
BOCKER, D ;
STADLBAUER, T ;
BREITHARDT, G ;
SCHELD, HH .
ANNALS OF THORACIC SURGERY, 1994, 58 (06) :1614-1616
[8]   CLINICAL-EXPERIENCE WITH NONTHORACOTOMY CARDIOVERTER-DEFIBRILLATORS [J].
LAWTON, JS ;
ELLENBOGEN, KA ;
WOOD, MA ;
STAMBLER, BS ;
SPRATT, JA ;
WECHSLER, AS ;
DAMIANO, RJ .
ANNALS OF THORACIC SURGERY, 1995, 59 (05) :1092-1099
[9]   LIMITATIONS AND LATE COMPLICATIONS OF 3RD-GENERATION AUTOMATIC CARDIOVERTER-DEFIBRILLATORS [J].
NUNAIN, SO ;
ROELKE, M ;
TROUTON, T ;
OSSWALD, S ;
KIM, YH ;
SOSASUAREZ, G ;
BROOKS, DR ;
MCGOVERN, B ;
GUY, M ;
TORCHIANA, DF ;
VLAHAKES, GJ ;
GARAN, H ;
RUSKIN, JN .
CIRCULATION, 1995, 91 (08) :2204-2213
[10]   EXPERIENCE WITH 3 DIFFERENT 3RD-GENERATION CARDIOVERTER-DEFIBRILLATORS IN PATIENTS WITH CORONARY-ARTERY DISEASE OR CARDIOMYOPATHY [J].
PORTERFIELD, JG ;
PORTERFIELD, LM ;
SMITH, BA ;
BRAY, L .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (03) :301-304