Purpose To compare the visual outcomes of digital and slit-beam manual marking for toric intraocular lenses (IOL) in cataract surgery. Setting Single-center, Beijing Tongren Hospital, China. Design Retrospective study. Methods All patients with cataracts and regular corneal astigmatism greater than 0.75 diopters (D) underwent cataract surgery and astigmatism correction between June 2019 and June 2020. To mark the target axis of the toric IOL and the location of the incision, intraoperative digital marking was used by Callisto eye image-guided system in one group, while preoperative manual slit-beam marking was used in the other group. Uncorrected and best-corrected spectacle visual acuity, refraction, toric IOL axis, total higher order aberrations, coma, spherical aberration, and trefoil were evaluated at 1, 4, and 12 weeks postoperatively. Results Seventy-two eyes of 58 patients were included. At 3 months after surgery, the mean residual refractive cylinder was 0.42 +/- 0.45D in the digital group and 0.39 +/- 0.40D in the manual group (P = 0.844). There were no significant differences between groups in spherical equivalent refraction, uncorrected and best-corrected spectacle visual acuity, or the parameters of vector analysis. All toric IOL alignment errors were within 10 degrees of the intended axis, and among them, about 42% of eyes in the digital group and 61% of eyes in the manual group had a rotation of 0-2 degrees (P = 0.038). Trefoil in the manual group decreased postoperatively compared with the digital group (P = 0.012). Other aberration analyses did not reveal any statistical differences between groups. Conclusions Accurate slit-beam manual marking and digital image-guided marking are equally effective for toric IOL alignment.
引用
收藏
页数:8
相关论文
共 16 条
[1]   Astigmatism analysis by the Alpins method [J].
Alpins, N .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (01) :31-49
[2]   Anterior stromal puncture with staining: A modified technique for preoperative reference corneal marking for toric lenses and its retrospective analyses [J].
Bhandari, Sahil ;
Nath, Manas .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2016, 64 (08) :559-562
[3]   Accuracy of toric intraocular lens axis alignment using a 3-dimensional computer-guided visualization system [J].
de Oca, Ildamaris Montes ;
Kim, Eric J. ;
Wang, Li ;
Weikert, Mitchell P. ;
Khandelwal, Sumitra S. ;
Al-Mohtaseb, Zaina ;
Koch, Douglas D. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04) :550-555
[4]   Comparison Between Digital and Manual Marking for Toric Intraocular Lenses A Randomized Trial [J].
Elhofi, Abdel Hamid ;
Helaly, Hany Ahmed .
MEDICINE, 2015, 94 (38)
[5]   Comparison of two different methods of preoperative marking for toric intraocular lens implantation: bubble marker versus pendulum marker [J].
Farooqui, Javed Hussian ;
Koul, Archana ;
Dutta, Ranjan ;
Shroff, Noshir Minoo .
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (05) :703-706
[6]   Residual astigmatism produced by toric intraocular lens rotation [J].
Felipe, Adelina ;
Artigas, Jose M. ;
Diez-Ajenjo, Amparo ;
Garcia-Domene, Carmen ;
Alcocer, Pablo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (10) :1895-1901
[7]   Sources of Error in Toric Intraocular Lens Power Calculation [J].
Hirnschall, Nino ;
Findl, Oliver ;
Bayer, Natascha ;
Leisser, Christoph ;
Norrby, Sverker ;
Zimper, Eva ;
Hoffmann, Peter .
JOURNAL OF REFRACTIVE SURGERY, 2020, 36 (10) :646-652
[8]   Axis Misalignment of Toric Intraocular Lens: Placement Error and Postoperative Rotation [J].
Inoue, Yasushi ;
Takehara, Hiroyasu ;
Oshika, Tetsuro .
OPHTHALMOLOGY, 2017, 124 (09) :1424-1425
[9]   Randomized trial comparing visual outcomes of toric intraocular lens implantation using manual and digital marker [J].
Kodavoor, Shreesha K. ;
Divya, J. ;
Dandapani, Ramamurthy ;
Ramamurthy, Chitra ;
Ramamurthy, Shreyas ;
Sachdev, Gitansha .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2020, 68 (12) :3020-3024
[10]   Comparison of visual outcomes, alignment accuracy, and surgical time between 2 methods of corneal marking for toric intraocular lens implantation [J].
Mayer, Wolfgang J. ;
Kreutzer, Thomas ;
Dirisamer, Martin ;
Kern, Christoph ;
Kortuem, Karsten ;
Vounotrypidis, Efstathios ;
Priglinger, Siegfried ;
Kook, Daniel .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (10) :1281-1286