High-resolution imaging of coronary calcifications by intense low-energy fluoroscopic X-ray obtained from synchrotron radiation

被引:0
作者
Ohtsuka, S
Sugishita, Y
Takeda, T
Itai, Y
Tada, J
Hyodo, K
Ando, M
机构
[1] Univ Tsukuba, Inst Clin Med, Dept Cardiol, Tsukuba, Ibaraki 3058575, Japan
[2] Univ Tsukuba, Inst Clin Med, Dept Radiol, Tsukuba, Ibaraki 3058575, Japan
[3] Univ Tsukuba, Inst Basic Med Sci, Tsukuba, Ibaraki 3058575, Japan
[4] Natl Lab High Energy Accelerator Res Org, Tsukuba, Ibaraki, Japan
关键词
coronary arteries; calcifications; radiation; technology;
D O I
10.1034/j.1600-0455.2000.041001064.x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose. To obtain an intense monochromatic low-energy X-ray from synchrotron radiation (SR) and apply it to detect coronary calcifications. Methods and Results: The SR beam was reflected with a silicon crystal to be expanded (150 mm in height and 80 mm in width) and to be monochromatized at an energy level of 37 keV. The X-ray was intermittently irradiated to obtain dynamic imaging of 30 images/s. Images were recorded by a digital fluorography system. The low-energy X-ray from SR sharply visualized calcification of coronary arteries, while conventional X-ray could not visualize coronary calcification. Conclusion: The intense monochromatic low-energy X-ray from SR is sensitive, has high-resolution for imaging coronary calcification and may serve as a screening method for coronary artery disease.
引用
收藏
页码:64 / 66
页数:3
相关论文
共 7 条
[1]   CORONARY-ARTERY CALCIFICATION DETECTED WITH ULTRAFAST CT AS AN INDICATION OF CORONARY-ARTERY DISEASE - WORK IN PROGRESS [J].
BREEN, JF ;
SHEEDY, PF ;
SCHWARTZ, RS ;
STANSON, AW ;
KAUFMANN, RB ;
MOLL, PP ;
RUMBERGER, JA .
RADIOLOGY, 1992, 185 (02) :435-439
[2]   Development of a two-dimensional imaging system for clinical applications of intravenous coronary angiography using intense synchrotron radiation produced by a multipole wiggler [J].
Hyodo, K ;
Ando, M ;
Oku, Y ;
Yamamoto, S ;
Takeda, T ;
Itai, Y ;
Ohtsuka, S ;
Sugishita, Y .
JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 :1123-1126
[3]  
HYODO K, 1991, HDB SYNCHROTRON RAD, V4, P55
[4]   DETECTION OF HEART CALCIFICATION WITH ELECTRON-BEAM CT - INTEROBSERVER AND INTRAOBSERVER RELIABILITY FOR SCORING QUANTIFICATION [J].
KAUFMANN, RB ;
SHEEDY, PF ;
BREEN, JF ;
KELZENBERG, JR ;
KRUGER, BL ;
SCHWARTZ, RS ;
MOLL, PP .
RADIOLOGY, 1994, 190 (02) :347-352
[5]   Dynamic intravenous coronary angiography using 2D monochromatic synchrotron radiation [J].
Ohtsuka, S ;
Sugishita, Y ;
Takeda, T ;
Itai, Y ;
Tada, J ;
Hyodo, K ;
Ando, M .
BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (853) :24-28
[6]   Electron beam computed tomography and coronary artery disease: Scanning for coronary artery calcification [J].
Rumberger, JA ;
Sheedy, PF ;
Breen, JF ;
Fitzpatrick, LA ;
Schwartz, RS .
MAYO CLINIC PROCEEDINGS, 1996, 71 (04) :369-377
[7]   Coronary artery calcification: Pathophysiology, epidemiology, imaging methods, and clinical implications - A statement for health professionals from the American Heart Association [J].
Wexler, L ;
Brundage, B ;
Crouse, J ;
Detrano, R ;
Fuster, V ;
Maddahi, J ;
Rumberger, J ;
Stanford, W ;
White, R ;
Taubert, K .
CIRCULATION, 1996, 94 (05) :1175-1192