Exact 3D cone-beam reconstruction from two short-scans using a C-arm imaging system

被引:0
作者
Ramamurthi, K [1 ]
Strobel, N [1 ]
Prince, JL [1 ]
机构
[1] Johns Hopkins Univ, Baltimore, MD 21218 USA
来源
COMPUTATIONAL IMAGING III | 2005年 / 5674卷
关键词
cone-beam; CT; C-arms; X-ray;
D O I
10.1117/12.597672
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
In this paper we present a source path for the purpose of exact cone-beam reconstruction using a C-arm X-ray imaging-system. The proposed path consists of two intersecting segments, each of which is a short-scan. Any C-arm capable of a short-scan sweep can thus be used to obtain data on our proposed source path as well., since it only requires an additional sweep on a, tilted plane. This tilt, can be achieved by either using the propeller axis of mobile C-arms, or the vertical axis of ceiling mounted C-arms. While the individual segments are only capable of exact reconstruction in their mid-plane, we show that the combined path is capable of exact reconstruction within an entire volumetric region. In fact, we show that the largest sphere that can be captured in the field of view of the C-arm can be exactly reconstructed if the tilt between the planes is at least equal to the cone-angle of the system. For the purpose of cone-beam inversion we use a generalized cone-beam filtered backprojection algorithm (CB-FBP). The exactness of this method relies on the design of a set of redundancy weights, which we explicitly evaluate for the proposed dual short-scan source path.
引用
收藏
页码:87 / 98
页数:12
相关论文
共 17 条
[1]   A THEORETICAL FRAMEWORK OF REGIONAL CONE-BEAM TOMOGRAPHY [J].
CHEN, JX .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (03) :342-350
[2]   A CONE-BEAM RECONSTRUCTION ALGORITHM USING SHIFT-VARIANT FILTERING AND CONE-BEAM BACKPROJECTION [J].
DEFRISE, M ;
CLACK, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :186-195
[3]  
FELDKAMP L, 1984, OPTICAL SOC AM, V1
[4]  
GRANGEAT P, 1991, LECT NOTES MATH, V1497, P66
[5]   Image reconstruction for the circle and line trajectory [J].
Katsevich, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (22) :5059-5072
[6]   Analysis of an exact inversion algorithm for spiral cone-beam CT [J].
Katsevich, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (15) :2583-2597
[7]  
Katsevich A, 2003, INT J MATH MATH SCI, V2003, DOI DOI 10.1155/S0161171203209315
[8]   Fast and stable cone-beam filtered backprojection method for non-planar orbits [J].
Kudo, H ;
Saito, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04) :747-760
[9]  
MULLERMERBACH J, 1996, MANUAL SIMULATION XR
[10]   Flat panel detector-based cone beam computed tomography with a circle-plus-two-arcs data acquisition orbit: Preliminary phantom study [J].
Ning, RL ;
Tang, XY ;
Conover, D ;
Yu, RF .
MEDICAL PHYSICS, 2003, 30 (07) :1694-1705