Optimization for fast-scanning irradiation in particle therapy

被引:35
作者
Inaniwa, Taku [1 ]
Furukawa, Takuji [1 ]
Tomitani, Takehiro [1 ]
Sato, Shinji [1 ]
Noda, Koji [1 ]
Kanai, Tatsuaki [1 ]
机构
[1] Natl Inst Radiol Sci, Ctr Charged Particle Therapy, Med Phys Res Grp, Inage Ku, Chiba 2638555, Japan
关键词
particle therapy; pencil beam scanning; optimization; moving target;
D O I
10.1118/1.2754058
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In three-dimensional irradiation with pencil beam scanning, an extra dose is inevitably delivered to the irradiated site due to the finite reaction times of the beam delivery system, and it causes a severe distortion of the dose distribution in the target region. Since the amount of the extra dose is proportional to the beam intensity, the dose uniformity deteriorates as the beam intensity is increased in order to shorten the treatment time. In order to overcome this problem and shorten the treatment time, we have developed an optimization method in which the extra dose is integrated into the optimization process of the best weighting matrix. The effectiveness and applicability of the optimization method for spot and raster scanning irradiation were confirmed with computer simulations and also confirmed using irradiation experiments for spot scanning irradiation. The treatment time could be shortened to about one sixth of the time needed without taking the extra dose into account while obtaining the same degree of dose homogeneity in the target volume. A typical treatment time with the proposed method is about 15 s for the irradiation of a spherical target with an 80 mm diameter at 3 GyE. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:3302 / 3311
页数:10
相关论文
共 20 条
[1]   Intensity control in RF-knockout extraction for scanning irradiation [J].
Furukawa, T ;
Noda, K ;
Uesugi, TH ;
Naruse, T ;
Shibuya, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 240 (1-2) :32-35
[2]   Characteristics of fast beam switching for spot scanning [J].
Furukawa, T ;
Noda, K ;
Urakabe, E ;
Muramatsu, M ;
Kanazawa, M ;
Maeda, K .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 503 (03) :485-495
[3]   Design study of a raster scanning system for moving target irradiation in heavy-ion radiotherapy [J].
Furukawa, Takuji ;
Inaniwa, Taku ;
Sato, Shinji ;
Tomitani, Takehiro ;
Minohara, Shinichi ;
Noda, Koji ;
Kanai, Tatsuaki .
MEDICAL PHYSICS, 2007, 34 (03) :1085-1097
[4]   Global spill control in RF-knockout slow-extraction [J].
Furukawaa, T ;
Noda, K ;
Muramatsu, M ;
Uesugi, T ;
Shibuya, S ;
Kawal, H ;
Takada, E ;
Yamada, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 522 (03) :196-204
[5]  
HIRAO Y, 1992, NUCL PHYS A, V538, pC541, DOI 10.1016/0375-9474(92)90803-R
[6]   3-DIMENSIONAL BEAM SCANNING FOR PROTON THERAPY [J].
KANAI, T ;
KAWACHI, K ;
MATSUZAWA, H ;
INADA, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 214 (2-3) :491-496
[7]   Present status of secondary beam courses in HIMAC [J].
Kanazawa, M ;
Kitagawa, A ;
Kouda, S ;
Nishio, T ;
Torikoshi, M ;
Noda, K ;
Murakami, T ;
Sato, S ;
Suda, M ;
Tomitani, T ;
Kanai, T ;
Futami, Y ;
Shinbo, M ;
Urakabe, E ;
Iseki, Y .
NUCLEAR PHYSICS A, 2004, 746 :393C-396C
[8]   Treatment planning for heavy-ion radiotherapy:: physical beam model and dose optimization [J].
Krämer, M ;
Jäkel, O ;
Haberer, T ;
Kraft, G ;
Schardt, D ;
Weber, U .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3299-3317
[9]   Respiratory gated irradiation system for heavy-ion radiotherapy [J].
Minohara, S ;
Kanai, T ;
Endo, M ;
Noda, K ;
Kanazawa, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (04) :1097-1103
[10]  
NOCEDAL J, 1996, SPRINGER SERIES OPER