The influence of lateral beam profile modifications in scanned proton and carbon ion therapy: a Monte Carlo study

被引:39
作者
Parodi, K. [1 ,2 ]
Mairani, A. [1 ,2 ]
Brons, S. [1 ,2 ]
Naumann, J. [1 ,2 ]
Kraemer, M. [3 ]
Sommerer, F. [1 ,2 ,4 ]
Haberer, T. [1 ,2 ]
机构
[1] Univ Clin Heidelberg, Heidelberg Ion Beam Therapy Ctr HIT, D-69120 Heidelberg, Germany
[2] Univ Clin Heidelberg, Dept Radiat Oncol, D-69120 Heidelberg, Germany
[3] GSI Helmholtzzentrum Schwerionenforsch, Dept Biophys, D-64291 Darmstadt, Germany
[4] CERN, European Org Nucl Res, CH-1211 Geneva 23, Switzerland
关键词
RADIATION-THERAPY; DOSE DISTRIBUTION; SKULL-BASE; MODEL; RADIOTHERAPY; DESIGN;
D O I
10.1088/0031-9155/55/17/018
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Scanned ion beam delivery promises superior flexibility and accuracy for highly conformal tumour therapy in comparison to the usage of passive beam shaping systems. The attainable precision demands correct overlapping of the pencil-like beams which build up the entire dose distribution in the treatment field. In particular, improper dose application due to deviations of the lateral beam profiles from the nominal planning conditions must be prevented via appropriate beam monitoring in the beamline, prior to the entrance in the patient. To assess the necessary tolerance thresholds of the beam monitoring system at the Heidelberg Ion Beam Therapy Center, Germany, this study has investigated several worst-case scenarios for a sensitive treatment plan, namely scanned proton and carbon ion delivery to a small target volume at a shallow depth. Deviations from the nominal lateral beam profiles were simulated, which may occur because of misaligned elements or changes of the beam optic in the beamline. Data have been analysed with respect to the lateral penumbra, homogeneity and coverage of the dose deposition in the target volume. The results indicate that homogeneity is not seriously compromised by extremely narrow profiles for the standard planning choices of the lateral raster scan stepping and dose grid. Differently, broad beam distributions can significantly deteriorate the conformality of the dose delivery and too large increases (above approximate to 150-200% of the nominal spotsize) must be prevented.
引用
收藏
页码:5169 / 5187
页数:19
相关论文
共 28 条
[1]  
Battistoni G, 2007, AIP CONF PROC, V896, P31, DOI 10.1063/1.2720455
[2]   INSTRUMENTATION FOR TREATMENT OF CANCER USING PROTON AND LIGHT-ION BEAMS [J].
CHU, WT ;
LUDEWIGT, BA ;
RENNER, TR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (08) :2055-2122
[3]   AN IMPROVED MULTIPLE-SCATTERING MODEL FOR CHARGED-PARTICLE TRANSPORT [J].
FERRARI, A ;
SALA, PR ;
GUARALDI, R ;
PADOANI, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1992, 71 (04) :412-426
[4]  
Ferrari A., 2005, 200510 CERN
[5]   EFFECTS OF RADIATIONS OF DIFFERENT QUALITIES ON CELLS - MOLECULAR MECHANISMS OF DAMAGE AND REPAIR [J].
GOODHEAD, DT ;
THACKER, J ;
COX, R .
INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 1993, 63 (05) :543-556
[6]   The heidelberg ion therapy center [J].
Haberer, T ;
Debus, J ;
Eickhoff, H ;
Jäkel, O ;
Schulz-Ertner, D ;
Weber, U .
RADIOTHERAPY AND ONCOLOGY, 2004, 73 :S186-S190
[7]   MAGNETIC SCANNING SYSTEM FOR HEAVY-ION THERAPY [J].
HABERER, T ;
BECHER, W ;
SCHARDT, D ;
KRAFT, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) :296-305
[8]  
Iancu G, 2009, HEAV IONS THER SPAC
[9]  
KLEMM S, 2009, HEAV IONS THER SPAC
[10]   Treatment planning for heavy-ion radiotherapy:: physical beam model and dose optimization [J].
Krämer, M ;
Jäkel, O ;
Haberer, T ;
Kraft, G ;
Schardt, D ;
Weber, U .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3299-3317