A power-voltage scaling law for liquid anode X-ray tubes

被引:1
|
作者
Harding, G [1 ]
机构
[1] Yxlon Int Secur, D-22419 Hamburg, Germany
关键词
X-ray source; liquid metal; cooling; scatter technique;
D O I
10.1016/j.radphyschem.2004.08.001
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
An analysis is presented of the relative magnitudes of factors affecting the specific loadability (W mm(-2) K-1) of liquid metal anode X-ray sources (LIMAX). These factors are to a certain extent analogous to those, such as focus width, focus-anode relative sweep velocity and the anode thermal conductivity, of importance in solid-state (e.g. rotating anode) X-ray tubes. It is shown that in general, convection cooling of the electron window by the liquid metal stream represents the least efficient thermal transport process in LIMAX. As the energy deposited in the electron window through inelastic electron scattering is approximately inversely proportional to electron energy, the power loadability of a LIMAX source operated under constant conditions scales to first approximation with the square of the tube potential. This power-voltage scaling law of liquid metal anode X-ray sources is without analogue in solid-state X-ray sources. A summary is presented of potential applications of high continuous power X-ray sources operating in the high-voltage region of several 100 kV. (c) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 50 条
  • [21] X-ray tubes
    Zink, F
    RADIOLOGY, 1996, 201 : 16 - 16
  • [22] Monitoring of the shape of the anode voltage curve in X-ray diagnostic apparatuses installed in X-ray rooms
    Shengeliya N.A.
    Biomedical Engineering, 2001, 35 (5) : 279 - 281
  • [23] TRANSMISSION ANODE X-RAY TUBES - ALTERNATE SOURCE OF MONOENERGETIC X-RAYS
    CATE, JL
    CAMPBELL, GW
    HEALTH PHYSICS, 1976, 31 (06): : 578 - 578
  • [24] Modeling and experimental investigation of x-ray spectra from a liquid metal anode x-ray tube
    David, B
    Thran, A
    Eckart, R
    X-RAY SOURCES AND OPTICS, 2004, 5537 : 45 - 56
  • [25] Paraffin liquid as a substitute for water in x-ray dianostic tubes
    Klason, T
    ACTA RADIOLOGICA, 1923, 2 (02) : 197 - 198
  • [26] X-ray diffraction imaging of the diamond anvils based on the microfocus x-ray source with a liquid anode
    Barannikov, Aleksandr
    Troyan, Ivan
    Snigireva, Irina
    Snigirev, Anatoly
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2022, 93 (08):
  • [27] Liquid metal anode X-ray sources and their potential applications
    Harding, G
    David, B
    Schlomka, JP
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 213 : 189 - 196
  • [28] The liquid-metal-jet anode x-ray source
    Hemberg, O
    Otendal, M
    Hertz, HM
    LASER-GENERATED AND OTHER LABORATORY X-RAY AND EUV SOURCES, OPTICS, AND APPLICATIONS, 2003, 5196 : 421 - 431
  • [29] Liquid-metal-jet anode X-ray tube
    Hemberg, O
    Otendal, M
    Hertz, HM
    OPTICAL ENGINEERING, 2004, 43 (07) : 1682 - 1688
  • [30] Low-power X-ray tubes (the current state)
    Bugaev, A. S.
    Eroshkin, P. A.
    Romanko, V. A.
    Sheshin, E. P.
    PHYSICS-USPEKHI, 2013, 56 (07) : 691 - 703