Scleral topography analysed by optical coherence tomography

被引:37
作者
Bandlitz, Stefan [1 ]
Baeumer, Joachim [1 ]
Conrad, Uwe [1 ]
Wolffsohn, James [2 ]
机构
[1] Cologne Sch Optometry, Hohere Fachschule Augenopt Koln, Bayenthalgurtel 6-8, D-50968 Cologne, Germany
[2] Aston Univ, Ophthalm Res Grp Life & Hlth Sci, Birmingham, W Midlands, England
关键词
Scleral radius; Optical coherence tomography; Cornea radius; Corneal eccentricity; Corneo-scleral topography; CORNEAL TOPOGRAPHY; LENSES; DESCRIPTORS; CURVATURE; ACCURACY; RADIUS; SHAPE;
D O I
10.1016/j.clae.2017.04.006
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: A detailed evaluation of the corneo-scleral-profile (CSP) is of particular relevance in soft and scleral lenses fitting. The aim of this study was to use optical coherence tomography (OCT) to analyse the profile of the limbal sclera and to evaluate the relationship between central corneal radii, corneal eccentricity and scleral radii. Methods: Using OCT (Optos OCT/SLO; Dunfermline, Scotland, UK) the limbal scleral radii (SR) of 30 subjects (11 M, 19F; mean age 23.8 +/- 2.0SD years) were measured in eight meridians 45 degrees apart. Central corneal radii (CR) and corneal eccentricity (CE) were evaluated using the Oculus Keratograph 4 (Oculus, Wetzlar, Germany). Differences between SR in the meridians and the associations between SR and corneal topography were assessed. Results: Median SR measured along 45 degrees (58.0; interquartile range, 46.8-84.8 mm) was significantly (p < 0.001) flatter than along 0 degrees (30.7; 24.5-44.3 mm), 135 degrees (28.4; 24.9-30.9 mm), 180 degrees (23.40; 21.3-25.4 mm), 225 degrees (25.8; 22.4-32.4 mm), 270 degrees (28.8; 25.3-33.1 mm), 315 degrees (30.0; 25.0-36.9 mm), and 90 degrees (37.1; 29.1-43.4 mm). In addition, the nasal SR along 0 degrees were significant flatter than the temporal SR along 180 degrees (p < 0.001). Central corneal radius in the flat meridian (7.83 +/- 0.26 mm) and in the steep meridian (7.65 +/- 0.26 mm) did not correlate with SR (p = 0.186 to 0.998). There was no statistically significant correlation between corneal eccentricity and scleral radii in each meridian (p = 0.422). Conclusions: With the OCT device used in this study it was possible to measure scleral radii in eight different meridians. Scleral radii are independent of corneal topography and may provide additional data useful in fitting soft and scleral contact lenses.
引用
收藏
页码:242 / 247
页数:6
相关论文
共 38 条
[1]   WHAT RADIUS DOES THE CONVENTIONAL KERATOMETER MEASURE [J].
BENNETT, AG ;
RABBETTS, RB .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1991, 11 (03) :239-247
[2]  
Bokern S, 2007, HEALTH PSYCHOL, V40, P26
[3]   ACCURACY OF VIDEOKERATOGRAPHY FOR INSTANTANEOUS RADIUS IN KERATOCONUS [J].
CHAN, JS ;
MANDELL, RB ;
BURGER, DS ;
FUSARO, RE .
OPTOMETRY AND VISION SCIENCE, 1995, 72 (11) :793-799
[4]   Measurement of Anterior Scleral Curvature Using Anterior Segment OCT [J].
Choi, Hyuk Jin ;
Lee, Sang-Mok ;
Lee, Ja Young ;
Lee, Seung Yong ;
Kim, Mee Kum ;
Wee, Won Ryang .
OPTOMETRY AND VISION SCIENCE, 2014, 91 (07) :793-802
[5]   Orbscan corneal topography system in evaluating the anterior structures of the human eye [J].
Cosar, CB ;
Sener, AB .
CORNEA, 2003, 22 (02) :118-121
[6]   Metrics of the normal anterior sclera: imaging with optical coherence tomography [J].
Ebneter, Andreas ;
Haener, Nathanael U. ;
Zinkernagel, Martin S. .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2015, 253 (09) :1575-1580
[7]  
Gaggioni M, 1987, NOJ, V1, P66
[8]   A novel method of fitting scleral high resolution optical coherence lenses using tomography [J].
Gemoules, Greg .
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2008, 34 (02) :80-83
[9]  
Gemoules Greg, 2006, Eye Contact Lens, V32, P158, DOI 10.1097/01.icl.0000219361.00199.d9
[10]   Factors Affecting Corneoscleral Topography [J].
Hall, Lee A. ;
Hunt, Chris ;
Young, Graeme ;
Wolffsohn, James .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (05) :3691-3701