A consideration of the signal-to-noise ratio in phase contrast mammography

被引:0
|
作者
Kato, Yuri [1 ]
Fujita, Naotoshi [2 ]
Kodera, Yoshie [3 ]
机构
[1] Nagoya Univ, Grad Sch Med Sci, Higashi Ku, 1-1-20 Daiko Minami, Nagoya, Aichi 4618673, Japan
[2] Nagoya Univ Hosp, Dept Radiol Technol, Shouwa Ku, Nagoya, Aichi 4668560, Japan
[3] Nagoya Univ, Sch Hlth Sci, Higashi Ku, Nagoya, Aichi 4618673, Japan
关键词
phase contrast mammography; scatter fraction ratio; signal-to-noise ratio; noise equivalent quanta; detective quantum efficiency;
D O I
10.1117/12.843771
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Recently, with developments in medicine, digital systems such as computed radiography (CR) and flat-panel detector (FPD) systems are being employed for mammography instead of analog systems such as the screen-film system. Phase-contrast mammography (PCM) is a commercially available digital system that uses images with a magnification of 1.75x. To study the effect of the air gap in PCM, we measured the scatter fraction ratio (SFR) and calculated the signal-to-noise ratio (SNR) in PCM, and compared it to that in conventional mammography (CM). Then, to extend the SNR to the spatial frequency domain, we calculated the noise equivalent quanta (NEQ) and detective quantum efficiency (DQE) used by the modulation transfer function (MTF), noise power spectrum of the pixel value (NPS Delta PV), gradient of the digital characteristic curve, and number of X-ray photons. The obtained results indicated that the SFR of the PCM was as low as that of the CM with a grid. When the exposure dose was constant, the SNR of the PCM was the highest in all systems. Moreover, the NEQ and DQE for the PCM were higher than those for the CM (G-) in the spatial frequency domain over 2.5 cycles/mm. These results showed that the number of scattered X-rays was reduced sufficiently by the air gap in the PCM and the NEQ and DQE for PCM were influenced by the presampled MTF in the high-spatial-frequency domain.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Low-dose, phase-contrast mammography with high signal-to-noise ratio
    Gromann, Lukas B.
    Beque, Dirk
    Scherer, Kai
    Willer, Konstantin
    Birnbacher, Lorenz
    Willner, Marian
    Herzen, Julia
    Grandl, Susanne
    Hellerhoff, Karin
    Sperl, Jonathan I.
    Pfeiffer, Franz
    Cozzini, Cristina
    BIOMEDICAL OPTICS EXPRESS, 2016, 7 (02): : 381 - 391
  • [2] CONTRAST AND SIGNAL-TO-NOISE RATIO IN RADIOGRAPHIC IMAGES
    ROBSON, KJ
    KOTRE, CJ
    FAULKNER, K
    RADIOLOGY, 1993, 188 (03) : 878 - 879
  • [3] Improved Signal-to-Noise Ratio in Laboratory-Based Phase Contrast Tomography
    Boone, Matthieu N.
    De Witte, Yoni
    Dierick, Manuel
    Almeida, Ana
    Van Hoorebeke, Luc
    MICROSCOPY AND MICROANALYSIS, 2012, 18 (02) : 399 - 405
  • [4] On the Definition of Signal-To-Noise Ratio and Contrast-To-Noise Ratio for fMRI Data
    Welvaert, Marijke
    Rosseel, Yves
    PLOS ONE, 2013, 8 (11):
  • [5] CONTRAST AND SIGNAL-TO-NOISE RATIO IN RADIOGRAPHIC IMAGES - REPLY
    ABILDGAARD, A
    NOTTHELLEN, JA
    RADIOLOGY, 1993, 188 (03) : 879 - 880
  • [6] Contrast signal-to-noise ratio for image quality assessment
    Yao, S
    Lin, W
    Ong, E
    Lu, Z
    2005 INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), VOLS 1-5, 2005, : 441 - 444
  • [7] Signal-to-noise ratio and contrast-to-noise ratio measurements for next generation tomosynthesis
    Martin, David A.
    Vent, Trevor L.
    Choi, Chloe J.
    Barufaldi, Bruno
    Acciavatti, Raymond J.
    Maidment, Andrew D. A.
    Progress in Biomedical Optics and Imaging - Proceedings of SPIE, 2021, 11595
  • [8] 'Signal-to-Noise Ratio'
    McCooey, D
    POETRY REVIEW, 1999, 89 (01): : 73 - 73
  • [9] On Wiener Phase Noise Channels at High Signal-to-Noise Ratio
    Ghozlan, Hassan
    Kramer, Gerhard
    2013 IEEE INTERNATIONAL SYMPOSIUM ON INFORMATION THEORY PROCEEDINGS (ISIT), 2013, : 2279 - +
  • [10] SIGNAL-TO-NOISE RATIO AND CONTRAST IN IMAGE DURING HOLOGRAM STORAGE
    GALPERN, AD
    OPTIKA I SPEKTROSKOPIYA, 1974, 37 (05): : 965 - 972