Accuracy of Platelet Counting by Automated Hematologic Analyzers in Acute Leukemia and Disseminated lntravascular Coagulation Potential Effects of Platelet Activation

被引:22
作者
Kim, Seon Young [1 ]
Kim, Ji-Eun [1 ,2 ]
Kim, Hyun Kyung [1 ,2 ]
Han, Kyou-Sup [1 ]
Toh, Cheng Hock [3 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Lab Med, Seoul 110744, South Korea
[2] Seoul Natl Univ, Coll Med, Canc Res Inst, Seoul 110744, South Korea
[3] Univ Liverpool, Dept Haematol, Liverpool L69 3BX, Merseyside, England
关键词
Platelet count; Acute leukemia; Disseminated intravascular coagulation; Platelet activation; FRAGMENTS; IMPACT;
D O I
10.1309/AJCP88JYLRCSRXPP
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Platelet counting in patients with acute leukemia or disseminated intravascular coagulation (DIC) may have a risk for erroneous counts owing to the presence of nonplatelet particles or platelet activation. We evaluated automated platelet counting methods using the Abbott Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA), Sysmex XE-2100 (Sysmex, Kobe, Japan), AD VIA 2120 (Siemens Diagnostics, Tarrytown, NY), and Beckman Coulter LH 750 (Beckman Coulter, Miami, FL) compared with the international reference method (IRM). Automated platelet counting methods were inaccurate compared with the IRM, without evidence of interfering nonplatelet particles. It is interesting that platelet activation markers were associated with DIC severity and erroneous platelet counting, suggesting that platelet activation is a potential source of inaccuracy. Furthermore, the artifactual in vitro platelet activation induced a high degree of intermethod variation in platelet counts. The inaccuracy of automated platelet counts increased the risk for misdiagnosis of DIC. More attention needs to be given to the accuracy of platelet counts, especially in clinical conditions with florid platelet activation.
引用
收藏
页码:634 / 647
页数:14
相关论文
共 32 条
[1]  
[Anonymous], FED REG
[2]   In vitro evaluation of platelet concentrates during storage: Platelet counts and markers of platelet destruction [J].
Apelseth, Torunn O. ;
Hervig, Tor .
TRANSFUSION AND APHERESIS SCIENCE, 2007, 37 (03) :261-268
[3]   Evaluation of a CD61 MoAb method for enumeration of platelets in thrombocytopenic patients and its impact on the transfusion decision-making process [J].
Arroyo, JL ;
García-Marcos, MA ;
López, A ;
Menéndez, P ;
Tabernero, MD ;
Sánchez-Abarca, LI ;
Avila-Zarza, C ;
San Miguel, JF ;
Orfao, A .
TRANSFUSION, 2001, 41 (10) :1212-1216
[4]   CYTOPLASMIC FRAGMENTS CAUSING SPURIOUS PLATELET COUNTS IN HAIRY-CELL LEUKEMIA - ULTRASTRUCTURAL CHARACTERIZATION [J].
BALLARD, HS ;
SIDHU, G .
ARCHIVES OF INTERNAL MEDICINE, 1981, 141 (07) :942-944
[5]   Applying the right statistics: analyses of measurement studies [J].
Bland, JM ;
Altman, DG .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 22 (01) :85-93
[6]   Effects of yeast on automated cell counting [J].
Branda, John A. ;
Kratz, Alexander .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2006, 126 (02) :248-254
[7]   New quantitative parameters on a recently introduced automated blood cell counter -: the XE 2100™ [J].
Briggs, C ;
Harrison, P ;
Grant, D ;
Staves, J ;
Machin, SJ .
CLINICAL AND LABORATORY HAEMATOLOGY, 2000, 22 (06) :345-350
[8]  
Chapman DH, 2001, AM J CLIN PATHOL, V115, P894
[9]  
HAMMERSTROM J, 1992, CLIN LAB HAEMATOL, V14, P239
[10]  
Harrison P, 2000, BRIT J HAEMATOL, V108, P228