Diamond coated total hip replacements

被引:0
|
作者
Lappalainen, R [1 ]
Anttila, A [1 ]
Heinonen, H [1 ]
机构
[1] Univ Helsinki, Dept Phys, FIN-00014 Helsinki, Finland
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Diamond has many superior, desired characteristics of implant materials such as low friction, high wear and corrosion resistance, and well bonding surface to bone. The potential of diamond for total hip replacement implants was studied in the form of amorphous diamond coatings on conventional metal implant materials. Amorphous diamond coatings (sp(3) bonding fraction 80%, thickness 0.2 to 10 mu m) were deposited on stainless steel AISI316L, Ti6A14V, and CoCrMo alloys using filtered pulsed plasma are discharge method. Superior attachment of coatings to the implant materials was achieved by using high energy plasma beams to deposit amorphous diamond and proper intermediate layers. Previously it was shown that these coatings are biocompatible causing no local tissue reactions. Tribologic studies using a pin on disk apparatus with coated or uncoated implant materials in 1 wt.% NaCl distilled water were performed, A simplified hip joint simulator was used for preliminary testing of metal on polyethylene and metal on metal artificial hip joints modified with amorphous diamond coating,. The average coefficients of friction were typically in the range of 0.03 to 0.11 for amorphous diamond coated materials. In the case of metal on metal hip implants, the average friction during initial running in period was improved (coefficient of friction = 0.07) compared with the same metal on metal pair (coefficient of friction = 0.22) and sliding was significantly smoother. In pin on disk wear tests, the average wear factors obtained were 140.10(6), 5.0.10(-6), and much less than 0.1.10(-6) mm(3)/Nm for the pairs of AISI316L, CoCrMo, and the same materials with amorphous diamond coating. The corrosion rates of these implant materials in 10 wt,% HCl solution were decreased by a factor of 10,000 to 15,000 and any damage of the coatings was not observed in 6 months. The results of the tests show that in all the combinations studied, amorphous diamond coating improved definitely the wear and corrosion resistance compared with the uncoated materials.
引用
收藏
页码:118 / 127
页数:10
相关论文
共 50 条
  • [41] The type of cement and failure of total hip replacements
    Espehaug, B
    Furnes, O
    Havelin, LI
    Engesæter, LB
    Vollset, SE
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (06): : 832 - 838
  • [42] A frictional study of total hip joint replacements
    Scholes, SC
    Unsworth, A
    Goldsmith, AAJ
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12): : 3721 - 3735
  • [43] COST PROFIT ANALYSIS OF TOTAL HIP REPLACEMENTS
    KEET, JGM
    ACTA ORTHOPAEDICA SCANDINAVICA, 1986, 57 (04): : 391 - 391
  • [44] Acetabular fixation options in total hip replacements
    Ritter, Merrill A.
    Galley, Matthew R.
    ORTHOPEDICS, 2008, 31 (09) : 887 - 888
  • [45] Infected total hip replacements: assessment and management
    Matar, Hosam E.
    Stritch, Paula
    Emms, Nicholas
    BRITISH JOURNAL OF HOSPITAL MEDICINE, 2018, 79 (05) : 265 - 269
  • [46] Hip disease and the prognosis of total hip replacements - A review of 53 698 primary total hip replacements reported to the Norwegian Arthroplasty Register 1987-99
    Furnes, O
    Lie, SA
    Espehaug, B
    Vollset, SE
    Engesaeter, LB
    Havelin, LI
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (04): : 579 - 586
  • [47] THE BENEFITS OF METAL-ON-METAL TOTAL HIP REPLACEMENTS
    MULLER, ME
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (311) : 54 - 59
  • [48] WEAR AND CORROSION OF MODULAR INTERFACES IN TOTAL HIP REPLACEMENTS
    COOK, SD
    BARRACK, RL
    BAFFES, GC
    CLEMOW, AJT
    SEREKIAN, P
    DONG, N
    KESTER, MA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1994, (298) : 80 - 88
  • [49] Influence of surgical approach on dislocation of total hip replacements
    Kohn, D
    Ruhmann, O
    Wirth, CJ
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1997, 135 (01): : 40 - 44
  • [50] Oxide ceramics for articulating components of total hip replacements
    Willmann, G