Femoral stem incorporating a diamond cubic lattice structure: Design, manufacture and testing

被引:96
作者
Jette, Bruno [1 ]
Brailovski, Vladimir [1 ]
Dumas, Mathieu [1 ]
Simoneau, Charles [1 ]
Terriault, Patrick [1 ]
机构
[1] Ecole Technol Super, Dept Mech Engn, 1100 Notre Dame St West, Montreal, PQ H3C 1K3, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Biomimetic femoral stem; Metallic lattice structure; Additive manufacturing; Digital image correlation; Finite element analysis; Hip prosthesis; BONE APPARENT DENSITY; MECHANICAL-PROPERTIES; POROUS BIOMATERIALS; STRAIN-MEASUREMENT; TITANIUM IMPLANTS; PORE-SIZE; FIXATION; SCAFFOLDS; BEHAVIOR; INGROWTH;
D O I
10.1016/j.jmbbm.2017.08.034
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The current total hip prostheses with dense femoral stems are considerably stiffer than the host bones, which leads to such long-term complications as aseptic loosening, and eventually, the need for a revision. Consequently, the lifetime of the implantation does not match the lifetime expectation of young patients. Method: A femoral stem design featuring a porous structure is proposed to lower its stiffness and allow bone tissue ingrowth. The porous structure is based on a diamond cubic lattice in which the pore size and the strut thickness are selected to meet the biomechanical requirements of the strength and the bone ingrowth. A porous stem and its fully dense counterpart are produced by laser powder-bed fusion using Ti-6Al-4V alloy. To evaluate the stiffness reduction, static testing based on the ISO standard 7206-4 is performed. The experimental results recorded by digital image correlation are analyzed and compared to the numerical model. Results & conclusions: The numerical and experimental force-displacement characteristics of the porous stem show a 31% lower stiffness as compared to that of its dense counterpart. Moreover, the correlation analysis of the total displacement and equivalent strain fields allows the preliminary validation of the numerical model of the porous stem. Finally, the analysis of the surface-to-volume and the strength-to-stiffness ratios of diamond lattice structures allow the assessment of their potential as biomimetic constructs for load-bearing orthopaedic implants.
引用
收藏
页码:58 / 72
页数:15
相关论文
共 60 条
[1]   Mechanical behavior of regular open-cell porous biomaterials made of diamond lattice unit cells [J].
Ahmadi, S. M. ;
Campoli, G. ;
Yavari, S. Amin ;
Sajadi, B. ;
Wauthle, R. ;
Schrooten, J. ;
Weinans, H. ;
Zadpoor, A. A. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 34 :106-115
[2]  
Aitkenhead AdamH., 2013, Mesh voxelisation-Voxelise a triangular-polygon mesh
[3]  
[Anonymous], 2007, Med. Sci, DOI [10.3923/jms.2007.460.467, DOI 10.3923/JMS.2007.460.467]
[4]  
[Anonymous], 2010, 72064 ISO
[5]  
[Anonymous], 2015, D69515 ASTM INT
[6]   High-strength porous biomaterials for bone replacement: A strategy to assess the interplay between cell morphology, mechanical properties, bone ingrowth and manufacturing constraints [J].
Arabnejad, Sajad ;
Johnston, R. Burnett ;
Pura, Jenny Ann ;
Singh, Baljinder ;
Tanzer, Michael ;
Pasini, Damiano .
ACTA BIOMATERIALIA, 2016, 30 :345-356
[7]  
Bahraminasab M., 2017, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
[8]   Application of Laser Engineered Net Shaping (LENS) to manufacture porous and functionally graded structures for load bearing implants [J].
Bandyopadhyay, Amit ;
Krishna, B. V. ;
Xue, Weichang ;
Bose, Susmita .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2009, 20 :29-34
[9]   AN APPROACH FOR TIME-DEPENDENT BONE MODELING AND REMODELING - THEORETICAL DEVELOPMENT [J].
BEAUPRE, GS ;
ORR, TE ;
CARTER, DR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1990, 8 (05) :651-661
[10]  
BOBYN JD, 1980, CLIN ORTHOP RELAT R, P263