Effect of carbon fiber type on monotonic and fatigue properties of orthopedic grade PEEK

被引:44
作者
Bonnheim, Noah [1 ]
Ansari, Farzana [2 ]
Regis, Marco [3 ]
Bracco, Pierangiola [3 ]
Pruitt, Lisa [1 ]
机构
[1] Univ Calif Berkeley, Dept Mech Engn, 2121 Etcheverry Hall, Berkeley, CA 94720 USA
[2] Exponent Inc, 149 Commonwealth Dr, Menlo Pk, CA 94025 USA
[3] Univ Torino, Dept Chem, Via P Giuria 7, I-10125 Turin, Italy
关键词
PEEK composites; Fatigue crack propagation; Orthopedic biomaterials; Fractography; CRACK GROWTH; CFR-PEEK; FRACTURE-BEHAVIOR; WEAR BEHAVIOR; MORPHOLOGY; IMPLANT; FAILURE; CRYSTALLIZATION; MICROSTRUCTURE; REINFORCEMENT;
D O I
10.1016/j.jmbbm.2018.10.033
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Carbon-fiber reinforced (CFR) PEEK implants are used in orthopedic applications ranging from fracture fixation plates to spinal fusion cages. Documented implant failures and increasing volume and variety of CFR PEEK implants warrant a clearer understanding of material behavior under monotonic and cyclic loading. To address this issue, we conducted monotonic and fatigue crack propagation (FCP) experiments on orthopedic grade unfilled PEEK and two formulations of CFR PEEK (PAN- and pitch-based carbon fibers). The effect of annealing on FCP behavior was also studied. Under monotonic loading, fiber type had a statistically significant effect on elastic modulus (12.5 +/- 1.3 versus 18.5 +/- 2.3 GPa, pitch versus PAN CFR PEEK, AVG +/- SD) and on ultimate tensile strength (145 +/- 9 versus 192 +/- 17 MPa, pitch versus PAN CFR PEEK, AVG +/- SD). Fiber type did not have a significant effect on failure strain. Under cyclic loading, PAN CFR PEEK demonstrated an increased resistance to FCP compared with unfilled and pitch CFR PEEK, and this improvement was enhanced following annealing. Pitch CFR PEEK exhibited FCP behavior similar to unfilled PEEK, and neither material was appreciably affected by annealing. The improvements in monotonic and FCP behavior of PAN CFR PEEK is attributed to a compound effect of inherent fiber properties, increased fiber number for an equivalent wt% reinforcement, and fiber aspect ratio. FCP was shown to proceed via cyclic modes during stable crack growth, which transitioned to static modes (more akin to monotonic fracture) at longer crack lengths. The mechanisms of fatigue crack propagation appear similar between carbon-fiber types.
引用
收藏
页码:484 / 492
页数:9
相关论文
共 51 条
[1]   Clinical and histologic results related to a low-modulus composite total hip replacement stem [J].
Akhavan, Sam ;
Matthiesen, Mary M. ;
Schulte, Leah ;
Penoyar, Tom ;
Kraay, Matthew J. ;
Rimnac, Clare M. ;
Goldberg, Victor M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (06) :1308-1314
[2]  
Ansari F., 2015, THESIS
[3]   Fractography and oxidative analysis of gamma inert sterilized posterior-stabilized tibial insert post fractures: Report of two cases [J].
Ansari, Farzana ;
Chang, Jennifer ;
Huddleston, James, III ;
Van Citters, Douglas ;
Ries, Michael ;
Pruitt, Lisa .
KNEE, 2013, 20 (06) :609-613
[4]   THE MORPHOLOGY OF POLY(ARYL-ETHER-ETHER-KETONE) [J].
BLUNDELL, DJ ;
OSBORN, BN .
POLYMER, 1983, 24 (08) :953-958
[5]  
Bonnheim Noah, 2017, Arthroplast Today, V3, P234, DOI 10.1016/j.artd.2017.06.005
[6]   A CARBON-FIBER IMPLANT TO AID INTERBODY LUMBAR FUSION - 2-YEAR CLINICAL-RESULTS IN THE 1ST 26 PATIENTS [J].
BRANTIGAN, JW ;
STEFFEE, AD .
SPINE, 1993, 18 (14) :2106-2117
[7]   FATIGUE FRACTURE-BEHAVIOR OF PEEK .1. EFFECTS OF LOAD LEVEL [J].
BRILLHART, M ;
GREGORY, BL ;
BOTSIS, J .
POLYMER, 1991, 32 (09) :1605-1611
[8]   FATIGUE FRACTURE-BEHAVIOR OF PEEK .2. EFFECTS OF THICKNESS AND TEMPERATURE [J].
BRILLHART, M ;
BOTSIS, J .
POLYMER, 1992, 33 (24) :5225-5232
[9]   FATIGUE-CRACK GROWTH ANALYSIS IN PEEK [J].
BRILLHART, M ;
BOTSIS, J .
INTERNATIONAL JOURNAL OF FATIGUE, 1994, 16 (02) :134-140
[10]   Long-term clinical consequences of stress-shielding after total hip arthroplasty without cement [J].
Bugbee, WD ;
Culpepper, WJ ;
Engh, CA ;
Engh, CA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (07) :1007-1012