Probability correction for the analytical simulation approach to calibrating γ-ray cylindrical detectors for environmental measurements

被引:2
作者
Nafee, S. Sherif [1 ,2 ]
机构
[1] Univ Alexandria, Fac Sci, Dept Phys, Alexandria 21121, Egypt
[2] King Abdullaziz Univ, Fac Sci, Dept Phys, Jeddah 21589, Saudi Arabia
关键词
Marinelli beaker; gamma-ray detectors; Self-attenuation; Coincidence summing corrections; COINCIDENCE CORRECTIONS; EXTENDED SOURCES; PEAK EFFICIENCY; HPGE DETECTORS; SAMPLES;
D O I
10.1016/j.apradiso.2011.03.007
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Marinelli beakers increase geometric efficiency of detection in the gamma-spectrometry of low-activity environmental samples by positioning the sample in close proximity to the detector. A probability correction has been introduced to the analytical simulation approach proposed previously (Nafee et al., 2009) for calculating detection efficiencies of gamma-ray cylindrical detectors using extended low- and high-volume Marinelli beakers. Improved expressions for the self-attenuation and the coincidence summing effects at short source-detector distances have also been included in the analytical algorithm. A remarkable agreement between the measured and calculated efficiencies was observed as a result. Deviations of results calculated in this way from experimental values are twice as small as were observed with the previously reported analytical methods (Abbas, 2001; Nafee et al., 2009). (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1019 / 1026
页数:8
相关论文
共 29 条