Unusual Case of a Trunnion Fracture Following a Revision Hip Arthroplasty Surgery: A Brief Review of the Literature and Discussion About Causes of Failure

被引:0
作者
Vick, Hamilton P. [1 ]
Domingue, Grayson A. [1 ,2 ]
Richburg, Kacy [1 ]
Dean, Ryan [1 ]
Khan, Fahmida [1 ]
Thakral, Rishi [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Oklahoma City, OK USA
[2] Univ Oklahoma, Hlth Sci Ctr, 13221 Englewood Ln, Edmond, OK 73013 USA
来源
ARTHROPLASTY TODAY | 2022年 / 18卷
关键词
Arthroplasty; Trunnion; Corrosion; Implant failure; Biomechanics; FEMORAL PROSTHESES; NECK FRACTURE; CORROSION; COMPONENT; STEM;
D O I
10.1016/j.artd.2022.09.012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Trunnion fracture is an incredibly rare complication of total hip arthroplasty. Of the few reported cases, all involve implants with faulty designs, a small neck taper, or an extended neck length or offset. Most also report corrosion and an adverse soft-tissue reaction. We present a review of the literature and report on the first case, to our knowledge, of trunnion fracture in a well-fixed, cemented cobalt-chromium femoral component with a standard neck length and offset with no evidence of corrosion. This failure was likely related to scratching of the metal during previous procedures which led to crack propagation and catastrophic failure. The patient was treated with revision hip arthroplasty to an uncemented, distalfit femoral component and insertion of new bearing surfaces.0 2022 The Authors. Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/ licenses/by-nc-nd/4.0/).
引用
收藏
页码:103 / 106
页数:4
相关论文
共 20 条
  • [1] Gross Trunnion Failure After Primary Total Hip Arthroplasty
    Banerjee, Samik
    Cherian, Jeffrey J.
    Bono, James V.
    Kurtz, Steven M.
    Geesink, Rudolph
    Meneghini, R. Michael
    Delanois, Ronald E.
    Mont, Michael A.
    [J]. JOURNAL OF ARTHROPLASTY, 2015, 30 (04) : 641 - 648
  • [2] Trunnion corrosion WHAT SURGEONS NEED TO KNOW IN 2018
    Berstock, J. R.
    Whitehouse, M. R.
    Duncan, C. P.
    [J]. BONE & JOINT JOURNAL, 2018, 100B (01) : 44 - 49
  • [3] Bolarinwa Surajudeen A, 2019, Arthroplast Today, V5, P5, DOI 10.1016/j.artd.2019.02.001
  • [4] Trunion fracture of a fully porous-coated femoral stem - Case report
    Botti, TP
    Gent, J
    Martell, JM
    Manning, DW
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (07) : 943 - 945
  • [5] NECK FRACTURES OF FEMORAL PROSTHESES A REPORT OF 2 CASES
    BURSTEIN, AH
    WRIGHT, TM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (03) : 497 - 499
  • [6] CHARNLEY J, 1975, CLIN ORTHOP RELAT R, P105
  • [7] Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty
    Cooper, H. John
    Della Valle, Craig J.
    Berger, Richard A.
    Tetreault, Matthew
    Paprosky, Wayne G.
    Sporer, Scott M.
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18) : 1655 - 1661
  • [8] Trunnion fracture of femoral prosthesis following a large metal-on-metal uncemented total hip arthroplasty: a case report
    Eachempati, Krishna Kiran
    Dannana, Chandra Sekhar
    Apsingi, Sunil
    Ponnala, Vinay Kishore
    Boyapati, Giridhar
    Parameswaran, Apurve
    [J]. ARTHROPLASTY, 2020, 2 (01)
  • [9] Early fatigue failures of cemented, forged, cobalt-chromium femoral stems at the neck-shoulder junction
    Lee, EW
    Kim, HT
    [J]. JOURNAL OF ARTHROPLASTY, 2001, 16 (02) : 236 - 238
  • [10] FACTORS IN MECHANICAL FAILURE OF FEMORAL COMPONENT IN TOTAL HIP PROSTHESIS - REPORT OF SIX FATIGUE FRACTURES OF FEMORAL STEM AND RESULTS OF EXPERIMENTAL LOADING TESTS
    MARTENS, M
    AERNOUDT, E
    MEESTER, PD
    DUCHEYNE, P
    MULIER, JC
    LANGH, RD
    KESTELIJN, P
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1974, 45 (05): : 693 - 710