Purpose: This study evaluated the failure mode, maximum load at failure, displacement at failure, and stiffness differences of quadrupled hamstring graft-tibial tunnel fixation using 28-mm, tapered 35-mm, or bicortical 17- and 20-mm bioabsorbable interference screws (Arthrex, Naples, FL). Type of Study: Biomechanical study. Methods: Nine matched pairs (18 specimens) of young cadaver tibias (mean bone mineral density [BMD]=0.847, range, 0.689 to 1.11 g/cm(2)) were divided into 3 groups of 6 specimens. Kruskal Wallis analysis of variance and Mann-Whitney U tests post hoc comparisons were used to assess group differences (P<.05). Results: Maximum load at failure for the 28-mm screw was 488.2 +/- 197 N, with a displacement of 18.4 +/- 7.5 mm and a stiffness modulus of 38.9 +/- 7.0 N/mm. Maximum load at failure for the tapered 35-mm screw was 844.8 +/- 121.4 N, with a displacement of 15.3 +/- 1.6 mm and a stiffness modulus of 70.5 +/- 21.1 N/mm. Maximum load at failure for bicortical screw fixation was 543.7 +/- 266.5 N, with a displacement of 17.7 +/- 6.6 mm and a stiffness modulus of 57.7 +/- 14.9 N/mm. Fixation using the tapered 35-mm screw displayed greater maximum load at failure than either 28-mm screw fixation (P=.015) or bicortical screw fixation (P=.037). Significant differences were not evident for displacement. Both tapered 35-mm (P=.016) and bicortical screw (P=.026) fixation displayed greater stiffness than 28-mm screw fixation. All constructs failed by pullout. Conclusions: Fixation using a single tapered 35-mm screw displayed increased maximum load at failure compared with the 28-mm screw or bicortical fixation. Both the tapered 35-mm screw and bicortical fixation showed greater stiffness than fixation using a 28-mm screw. Use of a single 35-mm screw for quadrupled hamstring graft-tibial tunnel fixation may be preferred for patients with poor tibial BMD.
引用
收藏
页码:991 / 996
页数:6
相关论文
共 20 条
[1]   Fibrocartilage in tendons and ligaments - an adaptation to compressive load [J].
Benjamin, M ;
Ralphs, JR .
JOURNAL OF ANATOMY, 1998, 193 :481-494
[2]   Graft fixation in cruciate ligament reconstruction [J].
Brand, J ;
Weiler, A ;
Caborn, DNM ;
Brown, CH ;
Johnson, DL .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (05) :761-774
[3]  
Firoozbakhsh K K, 1992, J Orthop Trauma, V6, P43
[4]   Current trends in anterior cruciate ligament reconstruction part 1: Biology and biomechanics of reconstruction [J].
Fu, FH ;
Bennett, CH ;
Lattermann, C ;
Ma, CB .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (06) :821-830
[5]  
Gao JZ, 1996, J ANAT, V188, P367
[6]   Comparative pull-out and cyclic-loading strength tests of anchorage of hamstring tendon grafts in anterior cruciate ligament reconstruction [J].
Giurea, M ;
Zorilla, P ;
Amis, AA ;
Aichroth, P .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (05) :621-625
[7]   Hamstring tendon grafts for reconstruction of the anterior cruciate ligament: Biomechanical evaluation of the use of multiple strands and tensioning techniques [J].
Hamner, DL ;
Brown, CH ;
Steiner, ME ;
Hecker, AT ;
Hayes, WC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (04) :549-557
[8]   Evaluation of the single-incision arthroscopic technique for anterior cruciate ligament replacement - A study of tibial tunnel placement, intraoperative graft tension, and stability [J].
Howell, SM ;
Wallace, MP ;
Hull, ML ;
Deutsch, ML .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (03) :284-293
[9]   EVALUATION OF EXPERIMENTAL AND FINITE-ELEMENT MODELS OF SYNTHETIC AND CADAVERIC FEMORA FOR PRECLINICAL DESIGN-ANALYSIS [J].
MCNAMARA, BP ;
CRISTOFOLINI, L ;
TONI, A ;
TAYLOR, D .
CLINICAL MATERIALS, 1994, 17 (03) :131-140
[10]  
Morrison J B, 1969, Biomed Eng, V4, P573