Titanium clip piston versus platinum-ribbon Teflon piston: Piston and fenestra size affect air-bone gap

被引:22
作者
Mangham, Charles A., Jr. [1 ]
机构
[1] Seattle Ear Clin, Seattle, WA 98121 USA
关键词
otosclerosis; piston; stapedotomy; Teflon; titanium;
D O I
10.1097/mao.0b013e31815c2575
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To analyze if titanium material and a clip attachment to the incus offer a hearing result advantage over the traditional Teflon piston in stapes surgery. Study Design: Retrospective chart review. Setting: Subspecialty private practice. Patients: One hundred seven ears in 97 consecutive patients who had primary stapes surgery during 2003 to 2005. Intervention: The first 74 ears received either a Teflon 0.5- or 0.6-mm piston, and the last 33 received a titanium 0.6-mm piston. Main Outcome Measures: American Academy of Otolaryngology-Head and Neck Surgery guidelines, including 4-frequency pure-tone average air-bone gap and success rate (gap, 10 dB). Results: Mean pure-tone average air-bone gap for the Teflon 0.6-mm piston (5.1 dB) was significantly smaller than for the titanium 0.6-mm piston (8.1 dB) and the Teflon 0.5-mm piston (7.5 dB). Success rate did not differ. Although labeled as 0.6-mm pistons, the Teflon piston diameter was 10% greater than the titanium piston. Results were comparable among devices when adjusted for true piston diameter. Results for the titanium piston were significantly better when the fenestra was no more than 0.05 mm larger than the piston diameter. Conclusion: Comparing 2 pistons designated 0.6 mm in diameter, the Teflon piston produced better hearing results than the titanium device. However, actual piston diameter differed between devices that contributed to the superior results with the larger Teflon piston. In addition, the titanium piston performed better with a small stapes fenestra diameter that suggests an advantage for titanium over Teflon in certain conditions. The clip design was problematic for a few patients.
引用
收藏
页码:8 / 12
页数:5
相关论文
共 10 条
[1]   A dynamic and harmonic damped finite element analysis model of stapedotomy [J].
Blayney, AW ;
Williams, KR ;
Rice, HJ .
ACTA OTO-LARYNGOLOGICA, 1997, 117 (02) :269-273
[2]   Prosthesis size in stapedectomy [J].
Fucci, MJ ;
Lippy, WH ;
Schuring, AG ;
Rizer, FM .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1998, 118 (01) :1-5
[3]   First experience with a new stapes clip piston in stapedotomy [J].
Grolman, W ;
Tange, RA .
OTOLOGY & NEUROTOLOGY, 2005, 26 (04) :595-598
[4]  
Hüttenbrink KB, 2003, OTOL NEUROTOL, V24, P548
[5]   Comparative electron microscopic study of the surface structure of gold, Teflon, and titanium stapes prostheses [J].
Kwok, P ;
Fisch, U ;
Strutz, J ;
Jacob, P .
OTOLOGY & NEUROTOLOGY, 2001, 22 (05) :608-613
[6]   Platinum ribbon-Teflon piston reduces device failure after stapes surgery [J].
Mangham, CA .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2000, 123 (01) :108-113
[7]   COMMITTEE ON HEARING AND EQUILIBRIUM GUIDELINES FOR THE EVALUATION OF RESULTS OF TREATMENT OF CONDUCTIVE HEARING-LOSS [J].
MONSELL, EM ;
BALKANY, TA ;
GATES, GA ;
GOLDENBERG, RA ;
MEYERHOFF, WL ;
HOUSE, JW .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1995, 113 (03) :186-187
[8]   LASER STAPEDOTOMY - A COMPARATIVE-STUDY OF PROSTHESES AND SEALS [J].
PERKINS, R ;
CURTO, FS .
LARYNGOSCOPE, 1992, 102 (12) :1321-1327
[9]  
ROSOWSKI JJ, 1995, AM J OTOL, V16, P486
[10]  
Shabana YK, 1999, CLIN OTOLARYNGOL, V24, P91