Accuracy and precision of volumetric bone mineral density assessment using dual-source dual-energy versus quantitative CT: a phantom study

被引:26
作者
Koch, Vitali [1 ]
Hokamp, Nils Grosse [2 ]
Albrecht, Moritz H. [1 ]
Gruenewald, Leon D. [1 ]
Yel, Ibrahim [1 ]
Borggrefe, Jan [3 ]
Wesarg, Stefan [4 ]
Eichler, Katrin [5 ]
Burck, Iris [5 ]
Gruber-Rouh, Tatjana [5 ]
Lenga, Lukas [1 ]
Vogl, Thomas J. [1 ,5 ]
Martin, Simon S. [1 ]
Wichmann, Julian L. [1 ]
Hammerstingl, Renate M. [5 ]
Alizadeh, Leona S. [1 ]
Mader, Christoph [5 ]
Huizinga, Nicole A. [6 ]
D'Angelo, Tommaso [7 ]
Ascenti, Giorgio [7 ]
Mazziotti, Silvio [7 ]
Booz, Christian [1 ]
机构
[1] Univ Hosp Frankfurt, Dept Diagnost & Intervent Radiol, Div Expt Imaging, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
[2] Univ Hosp Cologne, Dept Diagnost & Intervent Radiol, Cologne, Germany
[3] Univ Kiel, Minden Hosp, Dept Radiol Neuroradiol & Nucl Med, Kiel, Germany
[4] Fraunhofer IGD, Cognit Comp & Med Imaging, Darmstadt, Germany
[5] Univ Hosp Frankfurt, Dept Diagnost & Intervent Radiol, Frankfurt, Germany
[6] Goethe Univ Frankfurt, Interdisciplinary Ctr Neurosci, Frankfurt, Germany
[7] Univ Hosp Messina, Dept Biomed Sci & Morphol & Funct Imaging, Messina, Italy
关键词
Bone density; Dual-energy computed tomography; Osteoporosis; Phantoms (imaging); Tomography (x-ray computed); X-RAY ABSORPTIOMETRY; COMPUTED-TOMOGRAPHY; TRABECULAR BONE; OSTEOPOROSIS; MARROW; ERRORS; VALIDATION; FRACTURES; DESIGN; DXA;
D O I
10.1186/s41747-021-00241-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Dual-source dual-energy computed tomography (DECT) offers the potential for opportunistic osteoporosis screening by enabling phantomless bone mineral density (BMD) quantification. This study sought to assess the accuracy and precision of volumetric BMD measurement using dual-source DECT in comparison to quantitative CT (QCT). Methods A validated spine phantom consisting of three lumbar vertebra equivalents with 50 (L1), 100 (L2), and 200 mg/cm(3) (L3) calcium hydroxyapatite (HA) concentrations was scanned employing third-generation dual-source DECT and QCT. While BMD assessment based on QCT required an additional standardised bone density calibration phantom, the DECT technique operated by using a dedicated postprocessing software based on material decomposition without requiring calibration phantoms. Accuracy and precision of both modalities were compared by calculating measurement errors. In addition, correlation and agreement analyses were performed using Pearson correlation, linear regression, and Bland-Altman plots. Results DECT-derived BMD values differed significantly from those obtained by QCT (p < 0.001) and were found to be closer to true HA concentrations. Relative measurement errors were significantly smaller for DECT in comparison to QCT (L1, 0.94% versus 9.68%; L2, 0.28% versus 5.74%; L3, 0.24% versus 3.67%, respectively). DECT demonstrated better BMD measurement repeatability compared to QCT (coefficient of variance < 4.29% for DECT, < 6.74% for QCT). Both methods correlated well to each other (r = 0.9993; 95% confidence interval 0.9984-0.9997; p < 0.001) and revealed substantial agreement in Bland-Altman plots. Conclusions Phantomless dual-source DECT-based BMD assessment of lumbar vertebra equivalents using material decomposition showed higher diagnostic accuracy compared to QCT.
引用
收藏
页数:10
相关论文
共 43 条
[1]   Pitfalls in the measurement of bone mineral density by dual energy X-ray absorptiometry [J].
Antonacci, D ;
Hanson, DS ;
Heggeness, MH .
SPINE, 1996, 21 (01) :87-90
[2]   Use of dual-energy computed tomography to measure skeletal-wide marrow composition and cancellous bone mineral density [J].
Arentsen, Luke ;
Hansen, Karen E. ;
Yagi, Masashi ;
Takahashi, Yutaka ;
Shanley, Ryan ;
McArthur, Angela ;
Bolan, Patrick ;
Magome, Taiki ;
Yee, Douglas ;
Froelich, Jerry ;
Hui, Susanta K. .
JOURNAL OF BONE AND MINERAL METABOLISM, 2017, 35 (04) :428-436
[3]   Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation [J].
Bamberg, Fabian ;
Dierks, Alexander ;
Nikolaou, Konstantin ;
Reiser, Maximilian F. ;
Becker, Christoph R. ;
Johnson, Thorsten R. C. .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1424-1429
[4]   DXA in vivo BMD methodology:: An erroneous and misleading research and clinical gauge of bone mineral status, bone fragility, and bone remodelling [J].
Bolotin, H. H. .
BONE, 2007, 41 (01) :138-154
[5]   Diagnostic accuracy of quantitative dual-energy CT-based bone mineral density assessment in comparison to Hounsfield unit measurements using dual x-ray absorptiometry as standard of reference [J].
Booz, Christian ;
Noeske, Jochen ;
Albrecht, Moritz H. ;
Lenga, Lukas ;
Martin, Simon S. ;
Yel, Ibrahim ;
Huizinga, Nicole A. ;
Vogl, Thomas J. ;
Wichmann, Julian L. .
EUROPEAN JOURNAL OF RADIOLOGY, 2020, 132
[6]   Virtual Noncalcium Dual-Energy CT: Detection of Lumbar Disk Herniation in Comparison with Standard Gray-Scale CT [J].
Booz, Christian ;
Noske, Jochen ;
Martin, Simon S. ;
Albrecht, Moritz H. ;
Yel, Ibrahim ;
Lenga, Lukas ;
Gruber-Rouh, Tatjana ;
Eichler, Katrin ;
D'Angelo, Tommaso ;
Vogl, Thomas J. ;
Wichmann, Julian L. .
RADIOLOGY, 2019, 290 (02) :446-455
[7]   Quantitative computed tomography and opportunistic bone density screening by dual use of computed tomography scans [J].
Brett, Alan D. ;
Brown, J. Keenan .
JOURNAL OF ORTHOPAEDIC TRANSLATION, 2015, 3 (04) :178-184
[8]   Incidence and economic burden of osteoporosis-related fractures in the United States, 2005-2025 [J].
Burge, Russel ;
Dawson-Hughes, Bess ;
Solomon, Daniel H. ;
Wong, John B. ;
King, Alison ;
Tosteson, Anna .
JOURNAL OF BONE AND MINERAL RESEARCH, 2007, 22 (03) :465-475
[9]   EPIDEMIOLOGY OF OSTEOPOROSIS AND OSTEOPOROTIC FRACTURES [J].
CUMMINGS, SR ;
KELSEY, JL ;
NEVITT, MC ;
ODOWD, KJ .
EPIDEMIOLOGIC REVIEWS, 1985, 7 :178-208
[10]   Radiation exposure in X-ray-based imaging techniques used in osteoporosis [J].
Damilakis, John ;
Adams, Judith E. ;
Guglielmi, Giuseppe ;
Link, Thomas M. .
EUROPEAN RADIOLOGY, 2010, 20 (11) :2707-2714