Unusual behavior of the height equivalent to a theoretical plate of a new poroshell stationary phase at high temperatures

被引:49
作者
Gritti, Fabrice
Guiochon, Georges [1 ]
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Oak Ridge Natl Lab, Dept Chem Sci, Oak Ridge, TN 37831 USA
基金
美国国家科学基金会;
关键词
column performance; mass transfer mechanism; RP-HPLC; HETP equation; axial diffusion; eddy dispersion; film mass transfer; transparticle mass transfer; Halo-C-18; naphtho[2,3-a]pyrene; bovine serum albumin;
D O I
10.1016/j.chroma.2007.08.078
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The reduced HETPs of naphtho[2,3-a]pyrene (a polycyclic aromatic hydrocarbon with six condensed phenyl rings) was measured on three different RP-C,8columns packed with 5 and 3 mu m totally porous silica-B particles, and with 2.7 mu m Halo shell particles. The measurements were made at temperatures between ambient and 323 K, the retention factor being kept constant by modifying the eluent composition (water/acetonitrile mixtures). This compound was chosen because (1) it is strongly retained on RP-C-18 stationary phases and is eluted only by acetonitrile-rich mobile phases, which have a low viscosity (similar or equal to 0.35 cP). As a result, the C-term of the HETP plots was measured very precisely; (2) this compound does not interact strongly with the bare silica surface. Its adsorption isotherm remains linear up to high concentrations, its UV-absorbance (lambda = 294 nm) is high, and its HETP can be measured accurately and precisely (relative error of 2%). The experimental data were fitted to a general HETP model derived recently. The B term was measured independently, using the parking method to validate the fit in the low velocity range. The kinetics parameters derived from this model allow a qualitative comparison of the performances of the three columns. The column packed with the C-18-Halo particles exhibit an unusual rate of HETP increase at high linear velocities, particularly at elevated temperatures (310 and 323 K). A comparison of the HETP data measured for naphtho[2,3-a]pyrene and for bovine serum albumin (BSA), a large protein that is excluded from the mesopore network of all three columns, demonstrates that this behavior is related not to the mass transfer kinetics in the stationary phase but to some unexpected variation of the eddy dispersion with the linear velocity at high temperatures. The coupling theory of eddy dispersion proposed by Giddings fails to describe the data obtained for the Halo column while they predict well those measured on the totally porous materials. The explanation does not reside in the width of the particle size distribution (5% for Halo and similar or equal to 15% for the silica-B particles) but more likely in the roughness of the external surface of the Halo particles. (c) 200,7, Elsevier B.V. All rights reserved.
引用
收藏
页码:125 / 138
页数:14
相关论文
共 25 条
[1]   Mass-transfer kinetics in a shell packing material for chromatography [J].
Cavazzini, Alberto ;
Gritti, Fabrice ;
Kaczmarski, Krzysztof ;
Marchetti, Nicola ;
Guiochon, Georges .
ANALYTICAL CHEMISTRY, 2007, 79 (15) :5972-5979
[2]   Effect of mobile phase additives on peptide retention in reversed-phase chromatography with pellicular and totally porous sorbents [J].
Corradini, D ;
Kalghatgi, K ;
Horvath, C .
JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) :225-233
[3]  
Giddings J.C., 1965, DYNAMICS CHROMATOGRA
[4]  
Giddings J.C., 1991, Unified separation science
[5]   Comparison between the efficiencies of columns packed with fully and partially porous C18-bonded silica materials [J].
Gritti, Fabrice ;
Cavazzini, Alberto ;
Marchetti, Nicola ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1157 (1-2) :289-303
[6]   Measurement of the axial and radial temperature profiles of a chromatographic column - Influence of thermal insulation on column efficiency [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1138 (1-2) :141-157
[7]   Influence of the errors made in the measurement of the extra-column volume on the accuracies of estimates of the column efficiency and the mass transfer kinetics parameters [J].
Gritti, Fabrice ;
Felinger, Attila ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1136 (01) :57-72
[8]   Effect of the surface coverage of C18-bonded silica particles on the obstructive factor and intraparticle diffusion mechanism [J].
Gritti, Fabrice ;
Guiochon, Georges .
CHEMICAL ENGINEERING SCIENCE, 2006, 61 (23) :7636-7650
[9]   General HETP equation for the study of mass-transfer mechanisms in RPLC [J].
Gritti, Fabrice ;
Guiochon, Georges .
ANALYTICAL CHEMISTRY, 2006, 78 (15) :5329-5347
[10]  
Guiochon G, 2006, Fundamentals of Preparative and Nonlinear Chromatography, V2nd