Dosimetric accuracy of planning and delivering small proton therapy fields

被引:23
作者
Bednarz, Bryan
Daartz, Juliane
Paganetti, Harald
机构
[1] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA 02114 USA
关键词
MONTE-CARLO SIMULATIONS; DOSE CALCULATION; OUTPUT FACTORS; GEANT4; NOZZLE; RANGE;
D O I
10.1088/0031-9155/55/24/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A detailed Monte Carlo model of a proton therapy treatment head was commissioned to simulate the delivery of small field proton treatments. Small fields are challenging in the planning and quality assurance process because of aperture scattering and dosimetric disequilibrium. Four patients with small fields used in all or parts of their treatment course were studied, including two stereotactic patients and two fractionated patients. For the two stereotactic patients the overall difference of the dose covering at least 95% of the gross tumor volume between the Monte Carlo calculations and the delivery was -0.2% and -1.6%, respectively. For the two fractionated patients the overall difference of the dose covering at least 95% of the clinical target volume was -3.0% and 1.0%, respectively. We have thus confirmed that our current planning and delivery process for small proton fields is accurate enough to treat small lesions in the patient. Furthermore, we studied the impact of field size corrections and identified limitations of the pencil beam algorithm for predicting hot and cold spots and range degradation in the target due scattering in heterogeneities. For the four cases studied in this paper, these limitations appear to impact individual field calculations, but do not have a significant impact on the prescribed dose over multiple fields.
引用
收藏
页码:7425 / 7438
页数:14
相关论文
共 18 条
[11]   RADIOLOGICAL PROPERTIES OF BEAMS OF HIGH-ENERGY PROTONS [J].
LARSSON, B .
RADIATION RESEARCH, 1967, S :304-&
[12]   Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facility [J].
Paganetti, H ;
Jiang, H ;
Lee, SY ;
Kooy, HM .
MEDICAL PHYSICS, 2004, 31 (07) :2107-2118
[13]   Clinical implementation of full Monte Carlo dose calculation in proton beam therapy [J].
Paganetti, Harald ;
Jiang, Hongyu ;
Parodi, Katia ;
Slopsema, Roelf ;
Engelsman, Martijn .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (17) :4825-4853
[14]   Dose to water versus dose to medium in proton beam therapy [J].
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (14) :4399-4421
[15]   Evaluation of a pencil-beam dose calculation technique for charged particle radiotherapy [J].
Petti, PL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (05) :1049-1057
[16]  
PRESTON WM, 1968, EFFECTS SCATTE UNPUB
[17]   Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions [J].
Schneider, W ;
Bortfeld, T ;
Schlegel, W .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) :459-478
[18]   Monte Carlo investigation of collimator scatter of proton-therapy beams produced using the passive scattering method [J].
Titt, Uwe ;
Zheng, Yuanshui ;
Vassiliev, Oleg N. ;
Newhauser, Wayne D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (02) :487-504