Outcomes of allogenic cages in anterior and posterior lumbar interbody fusion

被引:35
|
作者
Janssen, ME [1 ]
Lam, C [1 ]
Beckham, R [1 ]
机构
[1] Ctr Spinal Disorders PC, Denver, CO 80229 USA
关键词
biological cages; femoral ring allograft spacer; posterior lumbar interbody fusion spacer; interbody lumbar fusion; arthrodesis;
D O I
10.1007/s005860100292
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Interbody lumbar fusions provide a proven logical solution to diseases of the intervertebral discs by eliminating motion of the segment. Historically, there are many techniques to achieve spinal fusion in the lumbar spine. These include anterior, posterior, and foramenal approaches, often in combination with various internal fixation devices. The surgeon's choice of the approach and mechanical or biological implant is dependent on the patient's specific pathology and anatomy, in addition to the experience and training of the surgeon in similar conditions. In the past decade, new mechanical spine implants/spacers have been designed to provide restoration of disc height and improve stabilization of the spine. The ability to radiographically assess the "biology" of bone incorporation in these mechanical (metal) spacers has become a significant limitation. The femoral ring allograft (FRA) and the posterior lumbar interbody fusion (PLIF) spacers have been developed as "biological cages" that permit restoration of the anterior column with machined allograft bone biological cages. Test results demonstrate that the FRA and PLIF spacers have a compressive strength of over 25,000 N. The pyramid-shaped teeth on the surfaces and the geometry of the implant increase the resistance to expulsion at clinically relevant loads (1053 and 1236 N). The technique of anterior column reconstruction with both the FRA and the PLIF biological cases have been previously reported. Clinical outcomes and experience with the FRA spacer (137 patients) and the PLIF spacer (13 patients) were reported on and did not reveal any evidence of bone cage resorption or infectious inflammatory process. There was clinical migration with one PLIF spacer, which was later revised with an anterior approach and a FRA spacer. The radiographic outcomes demonstrated that 94% arthrodesis was achieved with the biological spacer and additional posterior instrumentation. The clinical success of every spine fusion procedure is dependent on many factors such as the extent of the instability, the pathology, type of graft used, the patient's pathology/anatomy and lifestyle.
引用
收藏
页码:S158 / S168
页数:11
相关论文
共 50 条
  • [1] Outcomes of allogenic cages in anterior and posterior lumbar interbody fusion
    M. Janssen
    C. Lam
    R. Beckham
    European Spine Journal, 2001, 10 : S158 - S168
  • [2] Biomechanical comparison of anterior versus posterior lumbar threaded interbody fusion cages
    Vishteh, AG
    Crawford, NR
    Chamberlain, RH
    Thramann, JJ
    Park, SC
    Craigo, JB
    Sonntag, VKH
    Dickman, CA
    SPINE, 2005, 30 (03) : 302 - 310
  • [3] Posterior and anterior lumbar interbody fusion
    Madhu, Tiruveedhula S.
    CURRENT ORTHOPAEDICS, 2008, 22 (06): : 406 - 413
  • [4] Evolution of Design of Interbody Cages for Anterior Lumbar Interbody Fusion
    Phan, Kevin
    Mobbs, Ralph J.
    ORTHOPAEDIC SURGERY, 2016, 8 (03) : 270 - 277
  • [5] Anterior Placement of Cages in Posterior Lumbar Interbody Fusion for Obtaining Good Lumbar Lordosis Formation
    Inoue, Daisuke
    Shigematsu, Hideki
    Matsumori, Hiroaki
    Ueda, Yurito
    Morita, Toshiya
    Kawasaki, Sachiko
    Suga, Yuma
    Ikejiri, Masaki
    Tanaka, Yasuhito
    SPINE SURGERY AND RELATED RESEARCH, 2024, 8 (01): : 51 - 57
  • [6] Stand-alone Anterior Lumbar Interbody, Transforaminal Lumbar Interbody, and Anterior/Posterior Fusion: Analysis of Fusion Outcomes and Costs
    Bozzio, Anthony E.
    Johnson, Christopher R.
    Fattor, Jill A.
    Kleck, Christopher J.
    Patel, Vikas V.
    Burger, Evalina L.
    Noshchenko, Andriy
    Cain, Christopher M. J.
    ORTHOPEDICS, 2018, 41 (05) : E655 - E662
  • [7] Biomechanical comparison of posterior lumbar interbody fusion cages
    Rapoff, AJ
    Ghanayem, AJ
    Zdeblick, TA
    SPINE, 1997, 22 (20) : 2375 - 2379
  • [8] Motion of threaded cages in posterior lumbar interbody fusion
    Pitzen, T
    Geisler, FH
    Matthis, D
    Müller-Storz, H
    Steudel, WI
    EUROPEAN SPINE JOURNAL, 2000, 9 (06) : 571 - 576
  • [9] Motion of threaded cages in posterior lumbar interbody fusion
    T. Pitzen
    Fred H. Geisler
    Dieter Matthis
    Hans Müller-Storz
    Wolf-Ingo Steudel
    European Spine Journal, 2000, 9 : 571 - 576
  • [10] Biomechanical analysis of cages for posterior lumbar interbody fusion
    Fantigrossi, Alfonso
    Galbusera, Fabio
    Raimondi, Manuela Teresa
    Sassi, Marco
    Fornari, Maurizio
    MEDICAL ENGINEERING & PHYSICS, 2007, 29 (01) : 101 - 109