Reducing 4D CT imaging artifacts at the source: first experimental results from the respiratory adaptive computed tomography (REACT) system

被引:8
作者
Morton, Natasha [1 ]
Sykes, Jonathan [2 ,3 ]
Barber, Jeffrey [2 ,3 ]
Hofmann, Christian [4 ]
Keall, Paul [1 ]
O'Brien, Ricky [1 ]
机构
[1] Univ Sydney, Fac Med & Hlth, ACRF Image X Inst, Sydney, NSW, Australia
[2] Blacktown Hosp, Blacktown Canc & Haematol Ctr, Sydney West Canc Network, Blacktown, NSW, Australia
[3] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW, Australia
[4] Siemens AG, Imaging & Radiotherapy, Forchheim, Germany
基金
澳大利亚国家健康与医学研究理事会;
关键词
prospective gating; motion artifacts; 4D CT; RADIATION-THERAPY; MOTION; ACQUISITION; REDUCTION; IMPACT; RADIOTHERAPY; DATASET; IMAGES; 4D-CT;
D O I
10.1088/1361-6560/ab7abe
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Breathing variations during 4D CT imaging often manifest as geometric irregularities known as respiratory-induced image artifacts and ultimately effect radiotherapy treatment efficacy. To reduce such image artifacts we developed Respiratory Adaptive Computed Tomography (REACT) to trigger CT acquisition during periods of regular breathing. For the first time, we integrate REACT with clinical hardware and hypothesize that REACT will reduce respiratory-induced image artifacts >= 4 mm compared to conventional 4D CT. 4D image sets were acquired using REACT and conventional 4D CT on a Siemens Somatom scanner. Scans were taken for 13 respiratory traces (12 patients) that were reproduced on a lung-motion phantom. Motion was observed by the Varian RPM system and sent to the REACT software where breathing irregularity was evaluated in real-time and used to trigger the imaging beam. REACT and conventional 4D CT images were compared to a ground truth static-phantom image and compared for absolute geometric differences within the region-of-interest. Breathing irregularity during imaging was retrospectively assessed using the root-mean-square error of the RPM measured respiratory signal during beam on (RMSE_Beam_on) for each phase of the respiratory cycle. REACT significantly reduced the average frequency of respiratory-induced image artifacts >= 4 mm by 70% for the tumor (p = 0.003) and 76% for the lung (p = 0.0002) compared to conventional 4D CT. Volume reductions of 10% to 6% of the tumor and 2% to 1% of the lung compared to conventional 4D CT were seen. Breathing irregularity during imaging (RMSE_Beam_on) was significantly reduced by 27% (p = 0.013) using the REACT method. For the first time, REACT was successfully integrated with clinical hardware. Our findings support the hypothesis that REACT significantly reduced respiratory-induced image artifacts compared to conventional 4D CT. These experimental results provide compelling evidence for further REACT investigation, potentially providing clearer images for clinical use.
引用
收藏
页数:12
相关论文
共 36 条
[1]   Quantifying the impact of respiratory-gated 4D CT acquisition on thoracic image quality: A digital phantom study [J].
Bernatowicz, K. ;
Keall, P. ;
Mishra, P. ;
Knopf, A. ;
Lomax, A. ;
Kipritidis, J. .
MEDICAL PHYSICS, 2015, 42 (01) :324-334
[2]   Evaluation of 4D CT acquisition methods designed to reduce artifacts [J].
Castillo, Sarah J. ;
Castillo, Richard ;
Castillo, Edward ;
Pan, Tinsu ;
Ibbott, Geoffrey ;
Balter, Peter ;
Hobbs, Brian ;
Guerrero, Thomas .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02) :23-32
[3]   Experimental evaluations of the accuracy of 3D and 4D planning in robotic tracking stereotactic body radiotherapy for lung cancers [J].
Chan, Mark K. H. ;
Kwong, Dora L. W. ;
Ng, Sherry C. Y. ;
Tong, Anthony S. M. ;
Tam, Eric K. W. .
MEDICAL PHYSICS, 2013, 40 (04)
[4]   Radiomic feature stability across 4D respiratory phases and its impact on lung tumor prognosis prediction [J].
Du, Qian ;
Baine, Michael ;
Bavitz, Kyle ;
McAllister, Josiah ;
Liang, Xiaoying ;
Yu, Hongfeng ;
Ryckman, Jeffrey ;
Yu, Lina ;
Jiang, Hengle ;
Zhou, Sumin ;
Zhang, Chi ;
Zheng, Dandan .
PLOS ONE, 2019, 14 (05)
[5]   Tumor Tracking Method Based on a Deformable 4D CT Breathing Motion Model Driven by an External Surface Surrogate [J].
Fassi, Aurora ;
Schaerer, Joel ;
Fernandes, Mathieu ;
Riboldi, Marco ;
Sarrut, David ;
Baroni, Guido .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 88 (01) :182-188
[6]   Audio-visual biofeedback for respiratory-gated radiotherapy: Impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy [J].
George, Rohini ;
Chung, Theodore D. ;
Vedam, Sastry S. ;
Ramakrishnan, Viswanathan ;
Mohan, Radhe ;
Weiss, Elisabeth ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03) :924-933
[7]   Assessment of tumor motion reproducibility with audio-visual coaching through successive 4D CT sessions [J].
Goossens, Samuel ;
Senny, Frederic ;
Lee, John Aldo ;
Janssens, Guillaume ;
Geets, Xavier .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (01) :47-56
[8]   Reduction of irregular breathing artifacts in respiration-correlated CT images using a respiratory motion model [J].
Hertanto, Agung ;
Zhang, Qinghui ;
Hu, Yu-Chi ;
Dzyubak, Oleksandr ;
Rimner, Andreas ;
Mageras, Gig S. .
MEDICAL PHYSICS, 2012, 39 (06) :3070-3079
[9]   4D CT image reconstruction with diffeomorphic motion model [J].
Hinkle, Jacob ;
Szegedi, Martin ;
Wang, Brian ;
Salter, Bill ;
Joshi, Sarang .
MEDICAL IMAGE ANALYSIS, 2012, 16 (06) :1307-1316
[10]  
Hugo G. D., 2016, Data from 4D Lung Imaging of NSCLC Patients