TRACHEAL INTUBATION IN THE EMERGENCY DEPARTMENT: A COMPARISON OF GLIDESCOPE® VIDEO LARYNGOSCOPY TO DIRECT LARYNGOSCOPY IN 822 INTUBATIONS

被引:96
作者
Sakles, John C. [1 ]
Mosier, Jarrod M. [1 ]
Chiu, Stephen [1 ]
Keim, Samuel M. [1 ]
机构
[1] Univ Arizona, Coll Med, Dept Emergency Med, Tucson, AZ 85718 USA
关键词
intubation; video laryngoscopy; direct laryngoscopy; GlideScope (R); MACINTOSH LARYNGOSCOPE; VIDEOLARYNGOSCOPE GLIDESCOPE(R); TRUVIEW EVO2(R); DIFFICULT;
D O I
10.1016/j.jemermed.2011.05.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Video laryngoscopy has, in recent years, become more available to emergency physicians. However, little research has been conducted to compare their success to conventional direct laryngoscopy. Objectives: To compare the success rates of GlideScope (R) (Verathon Inc., Bothell, WA) videolaryngoscopy (GVL) with direct laryngoscopy (DL) for emergency department (ED) intubations. Methods: This was a 24-month retrospective observational study of all patients intubated in a single academic ED with a level I trauma center. Structured data forms were completed after each intubation and entered into a continuous quality improvement database. All patients intubated in the ED with either the GlideScope standard, Cobalt, Ranger, or traditional Macintosh or Miller laryngoscopes were included. All patients intubated before arrival were excluded. Primary analysis evaluated overall and first-attempt success rates, operator experience level, performance characteristics of GVL, complications, and reasons for failure. Results: There were 943 patients intubated during the study period; 120 were excluded due to alternative management strategies. DL was used in 583 (62%) patients, and GVL in 360 (38%). GVL had higher first-attempt success (75%, p = 0.03); DL had a higher success rate when more than one attempt was required (57%, p = 0.003). The devices had statistically equivalent overall success rates. GVL had fewer esophageal intubations (n = 1) than DL (n = 18); p = 0.005. Conclusion: The two techniques performed equivalently overall, however, GVL had a higher overall success rate, and lower number of esophageal complications. In the setting of ED intubations, GVL of-fers an excellent option to maximize first-attempt success for airway management. Published by Elsevier Inc.
引用
收藏
页码:400 / 405
页数:6
相关论文
共 14 条
[1]  
Cooper RM, 2005, CAN J ANAESTH, V52, P191, DOI 10.1007/BF03027728
[2]  
Cooper RM, 2003, CAN J ANAESTH, V50, P611, DOI 10.1007/BF03018651
[3]   Comparison of the GlideScope video laryngoscope and Macintosh laryngoscope in simulated tracheal intubation scenarios [J].
Kim, H. J. ;
Chung, S. P. ;
Park, I. C. ;
Cho, J. ;
Lee, H. S. ;
Park, Y. S. .
EMERGENCY MEDICINE JOURNAL, 2008, 25 (05) :279-282
[4]   GlideScope® video laryngoscope:: a randomized clinical trial in 203 paediatric patients [J].
Kim, J. -T. ;
Na, H. -S. ;
Bae, J. -Y. ;
Kim, D. -W. ;
Kim, H. -S. ;
Kim, C. S. ;
Kim, S. D. .
BRITISH JOURNAL OF ANAESTHESIA, 2008, 101 (04) :531-534
[5]   Comparison of the GlideScope® with the Macintosh laryngoscope for tracheal intubation in patients with simulated difficult airway [J].
Lim, Y ;
Yeo, SW .
ANAESTHESIA AND INTENSIVE CARE, 2005, 33 (02) :243-247
[6]   Comparison of Macintosh, Truview EVO2®, Glidescope®, and Airwayscope® laryngoscope use in patients with cervical spine immobilization [J].
Malik, M. A. ;
Maharaj, C. H. ;
Harte, B. H. ;
Laffey, J. G. .
BRITISH JOURNAL OF ANAESTHESIA, 2008, 101 (05) :723-730
[7]   Comparison of the Glidescope®, the Pentax AWS®, and the Truview EVO2® with the Macintosh laryngoscope in experienced anaesthetists: a manikin study [J].
Malik, M. A. ;
O'Donoghue, C. ;
Carney, J. ;
Maharaj, C. H. ;
Harte, B. H. ;
Laffey, J. G. .
BRITISH JOURNAL OF ANAESTHESIA, 2009, 102 (01) :128-134
[8]  
Malik MA, 2009, CAN J ANAESTH, V56, P802, DOI 10.1007/s12630-009-9165-z
[9]   Comparison of Intubation Success of Video Laryngoscopy Versus Direct Laryngoscopy in the Difficult Airway Using High-Fidelity Simulation [J].
Narang, Aneesh T. ;
Oldeg, Paula F. ;
Medzon, Ron ;
Mahmood, Ahmed R. ;
Spector, Jordan A. ;
Robinett, Derek A. .
SIMULATION IN HEALTHCARE, 2009, 4 (03) :160-165
[10]   A Comparison of GlideScope Video Laryngoscopy Versus Direct Laryngoscopy Intubation in the Emergency Department [J].
Platts-Mills, Timothy F. ;
Campagne, Danielle ;
Chinnock, Brian ;
Snowden, Brandy ;
Glickman, Larry T. ;
Hendey, Gregory W. .
ACADEMIC EMERGENCY MEDICINE, 2009, 16 (09) :866-871