Refractive outcomes and accuracy of IOL power calculation with the SRK/T formula for sutured, scleral-fixated Akreos AO60 intraocular lenses

被引:12
作者
Ohr, Matthew P. [1 ]
Wisely, C. Ellis [2 ]
机构
[1] Ohio State Univ, Dept Ophthalmol, 915 Olentangy River Rd Suite 5000, Columbus, OH 43212 USA
[2] Duke Univ, Dept Ophthalmol, Durham, NC USA
关键词
Sutured lens; Akreos; Gore-Tex; IOL power calculation; SRK; T formula; BIOMETRY; ABSENCE;
D O I
10.1007/s00417-020-04721-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Scleral fixation of intraocular lenses has become a popular procedure for treating aphakia in the absence of capsular support. However, the lens formulas used to predict refractive outcomes were designed for in-the-bag lens placement. This study evaluates the accuracy of the SRK/T formula in predicting a target postoperative refraction when suturing a scleral-fixated intraocular lens (IOL) implant 3 mm posterior to the limbus. Methods This is a retrospective, case series including 20 eyes of 20 patients who underwent scleral fixation of Akreos AO60 IOLs (Bausch & Lomb, Rochester, NY) by a single surgeon at the OSU Wexner Medical Center. Preoperative measurements were performed with optical biometry, and IOL power was calculated with the SRK/T formula. Following surgery, the actual refractive spherical equivalent (SE) was performed and compared with the preoperative prediction. Prediction error (PE), defined as the deviation of actual postoperative SE refraction in diopters (D) from preoperative predicted SE refraction, was the primary outcome measure. Results The mean attempted (predicted) SE was - 1.12 D (+/- 0.87). Mean achieved SE was - 0.96 D (+/- 1.04). Mean PE (actual postoperative SE versus predicted preoperative SE) was 0.16 D (+/- 0.69). A total of 9 eyes (45%) were within +/- 0.5 D of the predicted SE, 16 eyes (80%) were within +/- 1.0 D, and all 20 eyes (100%) were within +/- 1.5 D. Conclusion IOL power calculation using the SRK/T formula with optical biometry demonstrates reliable postoperative refractive outcomes in patients undergoing scleral fixation of an IOL (Akreos AO60). Further studies are needed to refine the predictive value of the SRK/T and other formulas for application in scleral fixation of IOLs.
引用
收藏
页码:2125 / 2129
页数:5
相关论文
共 19 条
[11]   Prediction of the true IOL position [J].
Norrby, Sverker ;
Bergman, Rolf ;
Hirnschall, Nino ;
Nishi, Yutaro ;
Findl, Oliver .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2017, 101 (10) :1440-+
[12]   Clinical outcomes of 4-point scleral fixated 1-piece hydrophobic acrylic equiconvex intraocular lens using polytetrafluoroethylene suture [J].
Patel, Nimesh A. ;
Shah, Parth ;
Yannuzzi, Nicolas A. ;
Ansari, Zubair ;
Zaveri, Jill S. ;
Relhan, Nidhi ;
Williams, Basil K., Jr. ;
Kuriyan, Ajay E. ;
Henry, Christopher R. ;
Sridhar, Jayanth ;
Haddock, Luis ;
Fortun, Jorge A. ;
Albini, Thomas A. ;
Davis, Janet L. ;
Flynn, Harry W., Jr. .
CLINICAL OPHTHALMOLOGY, 2018, 12 :2145-2148
[13]   DEVELOPMENT OF THE SRK/T INTRAOCULAR-LENS IMPLANT POWER CALCULATION FORMULA [J].
RETZLAFF, JA ;
SANDERS, DR ;
KRAFF, MC .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1990, 16 (03) :333-340
[14]  
Roh Young Rae, 2011, Korean J Ophthalmol, V25, P151, DOI 10.3341/kjo.2011.25.3.151
[15]   Refractive Outcomes after Pars Plana Vitrectomy and Scleral Fixated Intraocular Lens with Gore-Tex Suture [J].
Su, Daniel ;
Stephens, John D. ;
Obeid, Anthony ;
Borkar, Durga ;
Storey, Philip P. ;
Khan, M. Ali ;
Hsu, Jason ;
Garg, Sunir J. ;
Gupta, Omesh .
OPHTHALMOLOGY RETINA, 2019, 3 (07) :548-552
[16]   Small-incision 4-point scleral suture fixation of a foldable hydrophilic acrylic intraocular lens in the absence of capsule support [J].
Terveen, Daniel C. ;
Fram, Nicole R. ;
Ayres, Brandon ;
Berdahl, John P. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (02) :211-216
[17]   Minimum amount of astigmatism that should be corrected [J].
Villegas, Eloy A. ;
Alcon, Encarna ;
Artal, Pablo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (01) :13-19
[18]   Intraocular lens implantation in the absence of capsular support - A report by the American Academy of Ophthalmology [J].
Wagoner, MD ;
Cox, TA ;
Ariyasu, RG ;
Jacobs, DS ;
Karp, CL .
OPHTHALMOLOGY, 2003, 110 (04) :840-859
[19]   Accuracy of Intraocular Lens Power Calculation Formulas for Highly Myopic Eyes [J].
Zhang, Yichi ;
Liang, Xiao Ying ;
Liu, Shu ;
Lee, Jacky W. Y. ;
Bhaskar, Srinivasan ;
Lam, Dennis S. C. .
JOURNAL OF OPHTHALMOLOGY, 2016, 2016