Design optimization of blood shearing instrument by computational fluid dynamics

被引:26
作者
Wu, JC
Antaki, JF
Snyder, TA
Wagner, WR
Borovetz, HS
Paden, BE
机构
[1] LaunchPoint Technol LLC, Goleta, CA 93117 USA
[2] Carnegie Mellon Univ, Dept Biomed Engn & Comp Sci, Pittsburgh, PA 15213 USA
[3] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA USA
[4] Univ Pittsburgh, Dept Surg, Pittsburgh, PA USA
[5] Univ Pittsburgh, McGowan Inst Regenerat Med, Pittsburgh, PA USA
[6] Univ Calif Santa Barbara, Dept Mech & Environm Engn, Santa Barbara, CA 93106 USA
关键词
hemolysis; shear stress history; left ventricular assist device; computational fluid dynamics; optimization;
D O I
10.1111/j.1525-1594.2005.29082.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Rational design of blood-wetted devices requires a careful consideration of shear-induced trauma and activation of blood elements. Critical levels of shear exposure may be established in vitro through the use of devices specifically designed to prescribe both the magnitude and duration of shear exposure. However, it is exceptionally difficult to create a homogeneous shear-exposure history by conventional means. This study was undertaken to develop a Blood Shearing Instrument (BSI) with an optimized flow path which localized shear exposure within a rotating outer ring and a stationary conical spindle. By adjustment of the rotational speed and the gap dimension, the BSI is designed to generate shear stress magnitudes up to 1500 Pa for exposure time between 0.0015 and 0.20 s with a pressure drop of 100 mm Hg. Computational fluid dynamics (CFD) revealed that a flow path designed by first-order analysis and intuition exhibited unfavorable pressure gradient, vortices, and undesirable regions of reverse flow. An optimized design was evolved utilizing a parameterized geometric model and automatic mesh generation to eliminate vortices and reversal flow and to avoid unfavorable pressure gradients. Analysis of the flow and shear fields for the extreme limits of the shear gap demonstrated an improvement in homogeneity due to shape optimization and the limitations of an annular shear device for achieving completely uniform shear exposure.
引用
收藏
页码:482 / 489
页数:8
相关论文
共 31 条
[1]   EFFECT OF HIRUDIN ON PLATELET DEPOSITION TO AN ARTIFICIAL SURFACE DURING LOW-STRESS SHEAR-FLOW OF WHOLE-BLOOD [J].
ALKHAMIS, TM ;
BEISSINGER, RL ;
CHEDIAK, JR .
BIOMATERIALS, 1993, 14 (11) :865-870
[2]   COMPUTATIONAL FLOW OPTIMIZATION OF ROTARY BLOOD PUMP COMPONENTS [J].
ANTAKI, JF ;
GHATTAS, O ;
BURGREEN, GW ;
HE, BC .
ARTIFICIAL ORGANS, 1995, 19 (07) :608-615
[3]  
BLACKSHEAR PL, 1966, T AM SOC ART INT ORG, V12, P113
[4]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[5]   MODEL FOR A GENERAL MECHANICAL BLOOD DAMAGE PREDICTION [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :583-589
[6]   Computational fluid dynamics as a development tool for rotary blood pumps [J].
Burgreen, GW ;
Antaki, JF ;
Wu, ZJ ;
Holmes, AJ .
ARTIFICIAL ORGANS, 2001, 25 (05) :336-340
[7]   A design improvement strategy for axial blood pumps using computational fluid dynamics [J].
Burgreen, GW ;
Antaki, JF ;
Griffith, BP .
ASAIO JOURNAL, 1996, 42 (05) :M354-M360
[8]   Numerical calculation of hemolysis levels in peripheral hemodialysis cannulas [J].
De Wachter, D ;
Verdonck, P .
ARTIFICIAL ORGANS, 2002, 26 (07) :576-582
[9]  
Forstrom RJ, 1969, NEW MEASURE ERYTHROC
[10]   ESTIMATION OF SHEAR STRESS-RELATED BLOOD DAMAGE IN HEART-VALVE PROSTHESES - INVITRO COMPARISON OF 25 AORTIC VALVES [J].
GIERSIEPEN, M ;
WURZINGER, LJ ;
OPITZ, R ;
REUL, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (05) :300-306